Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Instrumental limitations

The lowest frequency typically used is 10 Hz. This limit is connected with the possible changes in the state of the electrode during long-period measurements. At this frequency, measurements averaged over five wave periods take 1 h 23 min. Measurement at all frequencies takes a much longer time.  [Pg.240]

It is relatively easy to get measurements of good precision for impedances between 1 and 10 Q at frequencies below 5 x 10 Hz. However, for lower and higher impedances, distortions may be observed. Very high impedances are found, for example, in measurements of protective coatings on metallic surfaces, and very low impedances are found in molten salts. The errors for high-impedance measurements originate from the finite potentiostat input impedance. Such resistance should be at least 100 times larger than the measured impedance if not, a calibration procedure is necessary. [Pg.240]

Another distortion is observed at very low impedances, corresponding to an inductance in series with the electrode impedance. It is observed at high frequencies and leads to large positive imaginary impedances. This inductance arises from that of the leads and the current [Pg.240]

Parsons, in Modern Aspects of Electrochemistry, Vol. 1, Plenum, New York, 1954, p. 103. [Pg.242]

Delahay, Double Layer and Electrode Kinetics, Wiley-Interscience, New York, 1965. D. M. Mohilner, in Electroanalytical Chemistry, A. J. Bard, ed., Vol. 1, Marcel Dekker, New York, 1966, p. 241. [Pg.243]


One instrumental limitation to Beer s law is the use of polychromatic radiation instead of monochromatic radiation. Consider a radiation source that emits two wavelengths of... [Pg.450]

A second instrumental limitation to Beer s law is stray radiation. The following data were obtained using a cell with a pathlength of 1.00 cm, when stray light is insignificant... [Pg.450]

This is the ratio of instrument limit primary current to the rated primary current. Consequently a high SF will mean a high transformation of the primary current and can damage instruments connected to its secondary. For measuring instruments therefore it is kept low, as it is required to measure only the normal current and not the fault current. [Pg.475]

Rated instrument limit primary current Rated primary current... [Pg.492]

Instrument Limitations Every instrument has certain limitations which must be recognized. A number of these have already been discussed, and it is very important that they be kept in mind at all times. Otherwise, serious errors can result, rendering the tests meaningless or misleading. Following is a brief summary of limitations common to many instruments, but it must be emphasized that any given instrument may not be susceptible to any panicular error ... [Pg.278]

Whilst nothing can improve upon the disadvantage of low molar absorption coefficients, instrumental designs and improvements with ratio recording and FT-IR instruments have virtually overcome the accuracy and instrumental limitations referred to in (b) and (c) above. As a result, quantitative infrared procedures are now much more widely used and are frequently applied in quality control and materials investigations. Applications fall into several distinct groups ... [Pg.752]

Raman scattering is essentially undelayed with respect to the arrival of the incident light, in this technique the detector is activated only during each laser pulse and deactivated at all other times. This allows only Raman signals to be recorded but fluorescence signals and detector noise are gated out (Fig. 19). Improvement in Raman signal to fluorescence ratio has been achieved as illustrated in Fig. 20. The technique, however, at present seems to be restricted by several instrumental limitations [37). [Pg.327]

Another factor of interest in defining fhe instrumental limitations is the instrumental quantification limit (IQL), which may be defined as the smallest amount of an analyte that can be reliably quantified by the instrument. [Pg.63]

The acquisition throughput of a microscope is often determined by photon statistics, but depends also on many parameters including instrumental limitations, for example, the read-out and dead-time of the detector and electronics [40],... [Pg.134]

Comparison with the modes of vibration of molecules given in Table I or with the LVM observed in proton implanted materials given in Table V and discussed in the next section clearly indicates that the LVMs observed in bulk material correspond to the stretching vibrations of P—H bonds in GaP and InP and As—H bonds in GaAs. It has to be noted that these lines are extremely sharp the FWHPs are in the 0.01-0.5 cm"1 range. For instance the line at 2202.4 cm"1 at 5 K in InP, which is shown in Fig. 19, has a FWHP of 0.015 cm"1, which could be instrument limited since the unapodized resolution limit of the interferometer used is 0.013 cm"1. [Pg.503]

Expensive and complex instrumentation. Moderate to poor sensitivity with continuous wave (scanning) instruments, but greatly enhanced by Fourier transform instruments. Limited range of solvents for studying proton spectra unless they are deuterated. [Pg.396]

In this case, sodium emission is monitored at a wavelength of 589.6 nm and potassium at a wavelength of 769.9 nm. The intensity of emission is calibrated with appropriate standards for the samples to be analyzed. In this way it is possible to automatically determine 100 values of sodium and potassium for 100 samples/h using modern clinical instruments. Limits of detection are sub-ppm and for serum values 140 mg/m the range of reproducibility is on the order of 2-3%. [Pg.256]

Perhaps the simplest mass analyzer of all, the TOF mass spectrometer [46] has experienced a reemergence in the past several years. Like the 3D quadrupole ion trap, the TOF analyzer has come to commercial prominence several decades after its initial introduction. The limitations of electronic components in the 1960s constrained the capabilities of the instrument, limiting its mass range and resolving power. The TOF analyzer operates in a pulsed mode, requiring either a pulsed ion... [Pg.361]

From both theory and experimental evidence, raising the temperature by 10°C decreases the retention time by about 20% in isocratic chromatography and decreases the backpressure by 10% to 20% because of a reduction in the viscosity of the mobile phase. This can help to overcome the instrument limitations associated with running shorter columns packed with smaller particles, i.e., the pressure limitations of current HPLC systems. However, since the majority of reversed-phase columns available are silica-based, operating at temperatures above... [Pg.620]

Most process Raman instruments have few, if any, moving parts and thus are quite stable and robust. This reduces the number of parts likely to need replacement. The laser and detector shutter are the most common service items. Special utilities and consumables are not required. Proper enclosure selection allows instruments to survive harsh environments, such as installation outdoors or around vibrations. An instrument built to accurate specifications and well installed is unlikely to require much care or daily attention. However, it is important to remember that there is no such thing as a maintenance-free instrument. Limited additional maintenance will ensure robust performance for years. [Pg.198]

BDE 17, 28, 47, 66, 85, 99, 100, 153, 154 183 Adipose tissues from marine mammals, chicken and trout Adipose Tissues, chicken and trout MSPD with silica gel/anhydrous sodium sulfate powder, purification thorug GPC extraction with 400 mL of 1 1 (v/v) acetone/hexane mixture Gas Chromatography (VF-5MS Eactor Eour, Varian) IT-MS 0.07-1.3 pg (instrumental limit of detection) [42]... [Pg.10]

Unfortunately, this technique is not selective and all components are affected in the same way so that if exact co-elution of sample components occurs, this may not be detected. Using a single quadrupole instrument limits the user to using in-source fragmentation, but in many cases this can provide enough information to identify unknowns. [Pg.171]

Anacon (Instruments) Limited, Bourne End, Buckinghamshire, England... [Pg.544]


See other pages where Instrumental limitations is mentioned: [Pg.387]    [Pg.398]    [Pg.420]    [Pg.1487]    [Pg.155]    [Pg.233]    [Pg.13]    [Pg.407]    [Pg.68]    [Pg.561]    [Pg.73]    [Pg.481]    [Pg.117]    [Pg.306]    [Pg.129]    [Pg.276]    [Pg.148]    [Pg.95]    [Pg.260]    [Pg.209]    [Pg.488]    [Pg.9]    [Pg.247]    [Pg.181]    [Pg.370]    [Pg.693]    [Pg.888]    [Pg.166]    [Pg.170]    [Pg.240]    [Pg.278]    [Pg.194]   
See also in sourсe #XX -- [ Pg.239 ]

See also in sourсe #XX -- [ Pg.333 , Pg.334 , Pg.335 , Pg.336 , Pg.337 , Pg.338 ]




SEARCH



Fourier transform instruments, limitations

Fundamental Limitations on Instrumental Precision

Impedance measurements instrumental limitations

Instrument Detection Limit (IDL

Instrumental detection limit

Instrumental detection limit (IDL

Instrumental detection limits definition

Instrumental limitations high frequencies

Instrumental limitations high impedances

Instrumental limitations reference electrode

Instrumental neutron activation detection limits

Instrumental quantification limit

Instrumentation limitations

Instrumentation limitations

Limits instrument detection

Pressure, measurement instrumentation limitation

© 2024 chempedia.info