Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Linearity of calibration

TSK-GEL PW type columns are commonly used for the separation of synthetic water-soluble polymers because they exhibit a much larger separation range, better linearity of calibration curves, and much lower adsorption effects than TSK-GEL SW columns (10). While TSK-GEL SW columns are suitable for separating monodisperse biopolymers, such as proteins, TSK-GEL PW columns are recommended for separating polydisperse compounds, such as polysaccharides and synthetic polymers. [Pg.106]

The blend of PVA with PEG- modified glucose oxidase could be used as glucose sensor characterized by the linearity of calibration curve in the range of concentration by 5 x 10"5 - 5 x 10"3 mol glucose L 1 [194],... [Pg.155]

Linearity of calibration Main source of uncertainty arising from the balance itself take manufacturers figure ( a mg) and treat as rectangular distribution, = / / = 0.15 mg... [Pg.177]

The most critical part of the method selection is determining the relative merits of seemingly diverse HPLC separations. Unfortunately, there is no standard against which all HPLC methods for vitamins are presently compared. However, there are indicators of assay reliability for which the analyst should look. These include measures of precision, accuracy, and reproducibility recoveries from spiked food samples linearity of calibration limits of detection measures of peak purity comparisons with existing recognized methods and results of collaborative or interlaboratory trials (9,10). [Pg.405]

As a supplemental approach to determine the linearity of calibration curves that span a wide concentration range, many analytical chemists replot the data on a log-log scale. Although that approach makes the data points at the lower end of the concentration scale easier to see, the data are displayed in a nonlinear system which distorts the patterns that existed in a linear scale.24 The tendency, however, is to interpret the curve using a linear scale, so log-log plots are easily misinterpreted. [Pg.238]

The temptation to allow data systems to determine the linearity of calibration curves on the basis of the value of the correlation coefficient alone has the potential of introducing determinate errors into chromatographic quantitation. Those errors can be avoided by always studying the results carefully before drawing any conclusions. [Pg.239]

Figure 4.14. Linearity of calibration curves for (a) Ag and (b) In using different side pixels for calibration, compared to that obtained with LS (HKL) (from Heitmann et al. [18]). Figure 4.14. Linearity of calibration curves for (a) Ag and (b) In using different side pixels for calibration, compared to that obtained with LS (HKL) (from Heitmann et al. [18]).
J. Van Loco, M. Elskens, C. Croux, H. Beemaert, Linearity of calibration curves use and misuse of the correlation coefficient, Accredit. Qual. Assur., 1 (2002), 281-285. [Pg.160]

A further advantage of combining IPC and MS is that isotope-labeled internal standards may be used for accurate quantification of analytes. No recovery check is needed for analytes because the labeled analogues of the analytes that serve as internal standards are added to the samples before the extraction step. This allows sample mixtures to be analyzed without pretreatment because the labeled internal standards share the fate of the un labeled analyte during sample processing without differences in extraction efficiency between the labeled standards and unlabeled analytes. This method achieved excellent precision and improved linearity of calibration lines despite interference from sample matrices in the quantitative analysis of important secondary intracellular metabolites in a complex biological sample solution from cultured cells. The technique is a valuable strategy for metabolomics [75],... [Pg.151]

Some molecules tend to show multimer formation [M +Adduct]" in ESI, where the adduct can be H, Na, NH4 or otherwise. Some examples are given by Kamel et al. [100-101]. In most cases, only adduct-bound dimers are observed, perhaps of the limited scan range chosen. These multimers can give problems with the unambiguous determination of the molecular mass of an unknown and with the linearity of calibration curve in quantitative analysis. Stefansson et al. [112] reported that the multimer formation of artemisinin, lasalocid, and deoxynivalenol could be reduced by the addition of primary amines. [Pg.166]

Figure 27, Linearity of calibration drip for manganese (10-40 ppb) in seawater with graphite tube use... Figure 27, Linearity of calibration drip for manganese (10-40 ppb) in seawater with graphite tube use...
Abstract Validation of analytical methods of well-characterised systems, such as are found in the pharmaceutical industry, is based on a series of experimental procedures to establish selectivity, sensitivity, repeatability, reproducibility, linearity of calibration, detection limit and limit of determination, and robustness. It is argued that these headings become more difficult to apply as the complexity of the analysis increases. Analysis of environmental samples is given as an example. Modern methods of analysis that use arrays of sensors challenge validation. The output may be a classification rather than a concentration of analyte, it may have been established by imprecise methods such as the responses of human taste panels, and the state space of possible responses is too large to cover in any experimental-design procedure. Moreover the process of data analysis may be done by non-linear methods such as neural networks. Validation of systems that rely on computer software is well established. [Pg.134]

The linearity of calibration should be established. However in practice laboratories using ICPOES or ICPMS tend to calibrate over a very wide range, often using single-point calibration in the vicinity of the concentration of that of the sample. The problem arises with a batch of samples that show a wide variation in concentrations of analyte. Proper attention to the calibration protocols will require a number of ranges to be vahdated. Again, in practice, this may not be adhered to because of pressure of the number of samples to be processed. [Pg.136]

LOD 0.04 to 1 /ng/1 RSD 8% linearity of calibration curves ranges over three orders of magnitude... [Pg.115]

Hguro 3 Cause and effect diagram of a titration in which the standard solution is made up from pure solid. (A) Initial diagram with elements from the measurement model (see eqn [1]). (B) Diagram with expanded influence factors. 7=temperature, s =repeatability, cal = manufacturer s calibration (see also eqn [2]). The mass measurements done by difference will contribute only linearity of calibration to the uncertainty labeled cal . [Pg.3984]

Linearity of calibration graph A simple experiment involving analysis of a series of standard solutions... [Pg.4084]

Fluorescent derivatives of prostaglandins were recently separated and determined by OPLC on HPTLC plates using ethyl acetate-diethyl ether-benzene-dioxane-hexane (45 12 5 8 30, v/v) mixture as eluent (88). The linearity of calibration graphs were good in the range 1-100 ng. This range was satisfactory for the determination because the detection limit of analytes were 10-40 pg. [Pg.198]

In one example, Masukawa and colleagues studied the quantification of Cer species in human stratum corneum by LC-MS [52], The authors employed a normal-phase column (Inertsil SIL lOOA-3 150 x 1.5 mm GL Science, Tokyo, Japan) with a nonlinear gradient from mobile phase A (hexane-isopropanol-formic acid, 95 5 0.1) to B (hexane-isopropanol-ammonium formate (50 mM) aqueous solution, 25 65 10) for 80 min at a flow rate of 0.1 mL/min. The researchers prepared numerous Cer analogs used as standards. They determined the limit of detection (S/N > 3), limit of quantification (S/N > 10), and linearity of calibration curves using these compounds. Moreover, they employed a Cer species (dl8 l-N17 0) as an internal standard for normalization of all analyzed samples. By using these controls, they achieved a maximum interday reproducibility of 14.3 RSD% and a systematic error of 21.4%. By using SIM, as many as 182 molecular-related ions derived from the diverse Cer species in the stratum corneum were measured. [Pg.323]

D. B. Hibbert, Further comments on the (miss-)use of r for testing the linearity of calibration functions, Accred. Qual. Assur., 2005, 10, 300-301. [Pg.138]

However, some problems of quantification accuracy persist, such as a lack of linearity of calibration lines (LECO, 2004) and last, but not least, the cost and complexity of such instruments are not suitable for quality control. [Pg.268]

Linearity of calibration plots and minimum detection limits (MDLs) were determined by injecting progressively lower concentrations of each species until the detection limit was reached. This was defined as a signal-to-noise ratio of about 2 1, using the absolute back-ground/baseline noise level as reference. In HPLC-Uy both selenite and selenate showed linear responses at concentrations up to about 50 ppm (50 /tl injections). Detection limits were on the order of about 0.1 ppm (100 ppb) for selenate and 10 ppb for selenite. Calibration plots were... [Pg.271]


See other pages where Linearity of calibration is mentioned: [Pg.584]    [Pg.103]    [Pg.83]    [Pg.359]    [Pg.168]    [Pg.100]    [Pg.371]    [Pg.103]    [Pg.146]    [Pg.141]    [Pg.45]    [Pg.100]    [Pg.138]    [Pg.285]    [Pg.1325]    [Pg.356]    [Pg.547]    [Pg.735]   
See also in sourсe #XX -- [ Pg.313 ]

See also in sourсe #XX -- [ Pg.277 ]




SEARCH



Calibration curve and linear response slope of pH microelectrodes

Calibration linear

Linear regression of straight-line calibration curves

Method Characteristic Parameters of a Linear Calibration Function

© 2024 chempedia.info