Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

In molecular mass determination

Mass spectrometry is more than 100 times more accurate than gel electrophoresis in molecular mass determination, and obtaining the data for sequence analysis requires only a fraction of the time needed for Edman degradation. Moreover, mass spectrometry is well suited for analysis of posttranslational modifications, which cannot be determined by Edman degradation. [Pg.309]

The chains of the produced polymer are lengthened by combination, but not by disproportionation. This affects the molecular mass distribution but the differences are not very large, differing by a factor of 2 at most. Due to the inaccuracies in molecular mass determinations, it is almost impossible to make estimates of the relation between termination and disproportionation from the distributions. Even labelling of the initiator and determination of the average number of its fragments in a macromolecule (one for disproportionation and two for combination) is usually unsafe because of transfer. [Pg.389]

When 0.412 g of naphthalene (CioHg) was dissolved in 10.0 g of camphor, the freezing point was found to be 13.0° below that of pure camphor. What is the mass-molar freezing-point constant for camphor, calculated from this observation Can you explain why camphor is frequently used in molecular mass determinations (Answer 40.4°.)... [Pg.314]

Besides affinity chromatography, capillary zone electrophoresis can easily be coupled to an ESI high-performance RTOF mass spectrometer. The high accuracy achievable in molecular mass determination by newly developed reflectron mass spectrometers makes the assignment of isomeric glycoproteins feasible [93] and can furthermore be a powerful tool to discriminate between different serologically relevant molecules. [Pg.478]

The techniques described thus far cope well with samples up to 10 kDa. Molecular mass determinations on peptides can be used to identify modifications occurring after the protein has been assembled according to its DNA code (post-translation), to map a protein structure, or simply to confirm the composition of a peptide. For samples with molecular masses in excess of 10 kDa, the sensitivity of FAB is quite low, and such analyses are far from routine. Two new developments have extended the scope of mass spectrometry even further to the analysis of peptides and proteins of high mass. [Pg.290]

In molecular weight determinations it is conventional to dissolve a measured mass of polymer m2 into a volumetric flask and dilute to the mark with an appropriate solvent. We shall use the symbol Cj to designate concentrations in mass per volume units. In practice, 100-ml volumetric flasks are often used, in which case C2 is expressed in grams per 100 ml or grams per deciliter. Even though these are not SI units, they are encountered often enough in the literature to be regarded as conventional solution units in polymer chemistry. [Pg.550]

The resolution required in any analytical SEC procedure, e.g., to detect sample impurities, is primarily based on the nature of the sample components with respect to their shape, the relative size differences of species contained in the sample, and the minimal size difference to be resolved. These sample attributes, in addition to the range of sizes to be examined, determine the required selectivity. Earlier work has shown that the limit of resolvability in SEC of molecules [i.e., the ability to completely resolve solutes of different sizes as a function of (1) plate number, (2) different solute shapes, and (3) media pore volumes] ranges from close to 20% for the molecular mass difference required to resolve spherical solutes down to near a 10% difference in molecular mass required for the separation of rod-shaped molecules (Hagel, 1993). To approach these limits, a SEC medium and a system with appropriate selectivity and efficiency must be employed. [Pg.30]

The molecular mass determined osmometrically corresponds to the formula S5O. The SO stretching vibration was observed in the infrared spectrum at 1119 cm (at -65 °G) indicating an exocyclic sulfoxide group similar to the one in SsO (see below). At -50 °G the solution of S5O may be kept for several days without decomposition which usually results in a Tyndall effect caused by a colloidal polymeric sulfuroxide which is the expected decomposition product. At 25 °G some decomposition already occurs within... [Pg.213]

The third described enzyme form with pH optimum about 4.7 [11, 4], we found in Fraction C - the fraction from carrot roots pulp (Fig. 2). We supposed that this form of exopolygalacturonase is relatively strongly bound on carrot cell walls and so it can be released only by higher salt concentrations. The approximative molecular mass determination on Superose 12 (Fig. 3c) showed the molecular mass about 50 000 for this form and the second, with more acidic pH optimum, form present in the fraction. The further characterization of these enzymes showed the exopolygalacturonase with pH optimum 4.7 to be identical with enzyme described sooner by Pressey and Avants [4] and exopolygalacturonase with pH optimum 3.8 to be identical with the enzyme from Fraction A. In conclusion, the exopolygalacturonase form with pH optimum 3.8 can be considered to be the main enzyme form present in carrot roots. [Pg.813]

The variability in number of kringles is genetically determined. Marcovina has described 34 different apo(a) isoforms differing in molecular mass by approximately 12.54 kDa, which closely corresponds to the predicted molecular mass of one kringle unit (M12, M30) (Fig. 5). [Pg.78]

Masses computed with Equation 22-8 appear in the last column of Table 22-3. A limitation on the accuracy of molecular mass determination is the accuracy of the mlz scale of the mass spectrum. Proteins of known mass can be used to calibrate the instrument. [Pg.495]

Molecular exclusion chromatography is based on the inability of large molecules to enter small pores in the stationary phase. Small molecules enter these pores and therefore exhibit longer elution times than large molecules. Molecular exclusion is used for separations based on size and for molecular mass determinations of macromolecules. In affinity chromatography, the stationary phase retains one particular solute in a complex mixture. After all other components have been eluted, the desired species is liberated by a change in conditions. [Pg.623]

Several newer methods of molecular mass determination were developed in the 1960s-1980s. One is gel filtration. A column of gel beads such as Sephadex is prepared carefully and is calibrated by passing a series of protein solutions through it. The volume V, at which a protein peak emerges from the column can be expressed as the sum of two terms (Eq. 3-19) in... [Pg.112]

Raman spectra. In a collision between a photon and a molecule, the photon may undergo elastic collision in which the photon loses no energy but changes its direction of travel. Such scattering is known as Rayleigh scattering and forms the basis for a method of molecular mass determination. Sometimes inelastic collisions occur in which both the... [Pg.1278]

Aluminum, gallium, indium, and thallium have been successfully incorporated in cyclic systems, although little structural information has been obtained (75). Hoberg (34) has prepared many such ring compounds, including, for example, 8a and 8b, and has demonstrated the former to be dimers from molecular mass determinations and from mass spectra. [Pg.273]

We said in the previous section that elemental analysis gives only an empirical formula. To determine a compound s molecular formula, it s also necessary to know the compound s molecular mass. How is molecular mass determined ... [Pg.102]

The molecular mass of ethylene glycol (Problem 3.109) is 62.0689 amu when calculated using the atomic masses found in a standard periodic table, yet the molecular mass determined experimentally by high-resolution mass spectrometry is 62.0368 amu. Explain the discrepancy. [Pg.113]

Modern mass spectrometers make it possible for substances in the femtomole range (10-15M) to be analyzed, and for high molecular masses of complex molecules well over 100 kDa to be obtained. Mass spectrometers allow molecular mass determination, and the generation of fragmentation data to allow sequencing and structure elucidation, generally in tandem mass-spectrometric experiments. [Pg.61]

In view of experimental simplicity and accuracy, viscosity measurements are extremely useful for routine relative molecular mass determinations on a particular polymer-solvent system. K and a for the system are determined by measuring the intrinsic viscosities of polymer fractions for which the relative molecular masses have been determined independently - e.g. by osmotic pressure, sedimentation or light scattering. [Pg.252]

The absorbance of la remained almost constant even after 850 cycles, while the absorbance of 3a gradually declined. At the same time, a photostable, violet product with an absorption maximum at 547 nm was formed. The photostable by-product could be isolated by HPLC, and was found by elemental analysis and molecular mass determination to be isomeric with compound 3a. Its molecular structure was determined by X-ray crystallographic analysis to be a six-membered condensed ring structure 9, as shown in Figure 3. The by-product was produced from the closed-ring form more efficiently by UV irradiation. [Pg.43]

For molecular mass determination, dissolve/dilute the purified peptide in acidified water (0.5% TFA) and deposit 0.5 xL of the acidified peptide solution. [Pg.23]


See other pages where In molecular mass determination is mentioned: [Pg.866]    [Pg.61]    [Pg.388]    [Pg.205]    [Pg.339]    [Pg.402]    [Pg.383]    [Pg.33]    [Pg.611]    [Pg.538]    [Pg.33]    [Pg.149]    [Pg.350]    [Pg.213]    [Pg.83]    [Pg.21]    [Pg.127]    [Pg.165]    [Pg.202]    [Pg.4]    [Pg.5]    [Pg.6]    [Pg.486]    [Pg.488]    [Pg.54]    [Pg.300]    [Pg.365]    [Pg.418]   
See also in sourсe #XX -- [ Pg.190 ]




SEARCH



Mass Determination

Molecular determinant

Molecular determination

Molecular mass

© 2024 chempedia.info