Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Fluorescence measurements flow cell

Radiation from a xenon or deuterium source is focussed on the flow cell. An interchangeable filter allows different excitation wavelengths to be used. The fluorescent radiation is emitted by the sample in all directions, but is usually measured at 90° to the incident beam. In some types, to increase sensitivity, the fluorescent radiation is reflected and focussed by a parabolic mirror. The second filter isolates a suitable wavelength from the fluorescence spectrum and prevents any scattered light from the source from reaching the photomultiplier detector. The 90° optics allow monitoring of the incident beam as well, so that dual uv absorption and fluorescence... [Pg.63]

A PMMA-Phe film was seated in the flow cell and was continuously exposed to 290 nm radiation throughout the experiment (Figure 1). The Phe fluorescence was monitored at 365 nm maximum which did not shift appreciably with a change in MEK concentration in PMMA-Phe. Therefore, the decay in the Phe fluorescence intensity provides an accurate measure of MEK diffusion and its SCP. [Pg.387]

Radiation from a Xenon-radiation or a Deuterium-source is focussed on the flow cell through a filter. The fluorescent radiation emitted by the sample is usually measured at 90° to the incident beam. The second filter picks up a suitable wavelength and avoids all scattered light to reach ultimately the photomultiplier detector. [Pg.463]

For a fluorescence detector, quinine sulfate is used as the standard compound. The flow cell is filled with a standard solution and the fluorescence intensity is measured. The value is compared with that measured by a fluorescence spectrophotometer. This standard solution is also used for fixing the wavelength and position of the flow cell. The Raman spectrum of water can also be used for this purpose. [Pg.23]

Laser microbeams offer several advantages over other fluorescence excitation techniques. In spectrofluorometry, observations are often made on a population of cells in a cuvette, resulting in a combined signal that lacks information about individual cellular responses. In flow cytometry, many individual cells are measured, but there is no temporal resolution since each cell is observed only once, and there is no spatial resolution since the entire cell is illuminated as it passes through the laser beam (see Chapter 30). In conventional fluorescence microscopy, individual cells can be monitored over time, and information about the two-dimensional spatial distribution of fluorescence can be obtained. However, some samples may be more susceptible to photobleaching by the arc lamps used for excitation, and the temporal resolution is limited to video-rate data acquisition (30 frames/s) (see Chapter 14). [Pg.161]

It was shown by Creech and Jones (1) in 1940 that proteins, including antibodies, could be labeled with a fluorescent dye (phenylisocyanate) without biological or immunological effects to the intended target. In theory, fluorescent reporters (tracers, probes, antibodies, stains, and so on) can be used to detect or measure any cell constituent, provided that the tag reacts specifically and stoichiometrically with the cellular constituent in question (2). Today, the repertoire of fluorescent probes is expanding almost daily see Chapter 14). One area that has benefited from the ever-increasing number of fluorescent probes is flow cytometry. [Pg.249]

Flow cytometry (FCM) is a high-precision technique for rapid analysis and sorting of cells and particles. In theory, it can be used to measure any cell component, provided that a fluorescent tracer is available that reacts specifically and stoichiometrically with that constituent. The technique provides statistical accuracy, reproducibility, and sensitivity. [Pg.271]

In some manufacturing process analysis applications the analyte requires sample preparation (dilution, derivatization, etc.) to afford a suitable analytical method. Derivatization, emission enhancement, and other extrinsic fluorescent approaches described previously are examples of such methods. On-line methods, in particular those requiring chemical reaction, are often reserved for unique cases where other PAT techniques (e.g., UV-vis, NIR, etc.) are insufficient (e.g., very low concentrations) and real-time process control is imperative. That is, there are several complexities to address with these types of on-line solutions to realize a robust process analysis method such as post reaction cleanup, filtering of reaction byproducts, etc. Nevertheless, real-time sample preparation is achieved via an on-line sample conditioning system. These systems can also address harsh process stream conditions (flow, pressure, temperature, etc.) that are either not appropriate for the desired measurement accuracy or precision or the mechanical limitations of the inline insertion probe or flow cell. This section summarizes some of the common LIF monitoring applications across various sectors. [Pg.349]

The fluorescence of pyridoxal is dramatically increased by complexation with beryllium ion, which provides a means for accurate measurements of these Bg vitamers by using a flow-cell packed with C,g bonded silica beads. The complex is eluted from the sensor by the aqueous stream itself, into which the sample and beryllium are simultaneously injected [68]. [Pg.218]

Drug residues in foods that strongly fluoresce can be more efficiently detected by fluorescence detectors. Typically, fluorescence sensitivity is 10-1000 times higher than that offered by a UV detector for strong UV-absorbing materials (125). Using a fluorescence detector, it has been possible to detect the presence of even a single analyte molecule in an LC flow cell. This type of detection is very versatile because of its ability to measure the intensity of the fluorescent radiation emitted from analytes excited by UV. [Pg.697]

HPLC analysis of polycyclic aromatic hydrocarbons (PAH) in drinking water is one of the current and classical applications of fluorescence. In this case, the detector contains a fluorescence flow cell placed after the chromatographic column. This mode of detection is specifically adapted to obtain threshold measurements imposed by legislation. The same process allows the measurement of aflatoxins (Fig. 12.11) and many other organic compounds (such as adrenaline, quinine, steroids and vitamins). [Pg.230]

Figure 6. Instrumental schematic for vacuum UV photofragmentation-laser induced fluorescence measurement of ammonia SHGC, second harmonic generation crystal SFMC, sum frequency mixing crystal BS, beam splitter BD, beam dump TP, turning prism CL, cylindrical lens R, reflector TD, trigger diode OSC, oscillator cell AMP, amplifier cell BE, beam expander G, grating OC, output coupler M, mirror BC, beam combiner L, lens A, aperture PD, photodiode SC, sample cell RC, reference cell FP, filter pack SAM.PMT, sample cell photomultiplier REF.PMT, reference cell photomultiplier PP, additional photomultiplier port EX, exhaust and CGI, calibration gas inlet to flow line. (Reproduced with permission from reference 15. Copyright 1990 Optical Society of America.)... Figure 6. Instrumental schematic for vacuum UV photofragmentation-laser induced fluorescence measurement of ammonia SHGC, second harmonic generation crystal SFMC, sum frequency mixing crystal BS, beam splitter BD, beam dump TP, turning prism CL, cylindrical lens R, reflector TD, trigger diode OSC, oscillator cell AMP, amplifier cell BE, beam expander G, grating OC, output coupler M, mirror BC, beam combiner L, lens A, aperture PD, photodiode SC, sample cell RC, reference cell FP, filter pack SAM.PMT, sample cell photomultiplier REF.PMT, reference cell photomultiplier PP, additional photomultiplier port EX, exhaust and CGI, calibration gas inlet to flow line. (Reproduced with permission from reference 15. Copyright 1990 Optical Society of America.)...
The complexity of the data generated by multiparameter flow cytometric analysis means that the scatter and fluorescence measurements from individual cells are stored on the computer in the form of a list (list mode data). This means that the relationships between measurements on individual cells is preserved, and these relationships can be explored by effectively rerunning the analysis on the computer. [Pg.340]

Three types of inline HPLC detector have been used to measure fat-soluble vitamin concentrations in food sample extracts absorbance, fluorescence, and electrochemical detectors. Each of these detectors provides a continuous electrical output that is a function of the concentration of solute in the column effluent passing through the flow cell. [Pg.352]

Occasionally, a laboratory will need an in-line detector of radio-labeled molecules. These detectors take the flow from the column or from an initial detector, mix it with fluorescing compound, and measure the fluorescence due to radioactive breakdown. A different system uses beads in the flow cell with an immobilized fluorescing compound, but these systems suffer from ghosting and cannot be used with very hot labeled compounds because of secondary radiation problems. These systems are very useful with tritiated samples and less so with carbon14 labeled compounds. Some success has been reported with sulfur32 label detection. [Pg.123]

By far most of the work discussed in this review has been based on LIF detection, usually with an 488 nm Ar-ion laser as the excitation source. Only very few other examples exist in the literature where other detection principles were investigated. One of these exceptions is an integrated detection cell for chip CE that has been described by Liang et al. [78]. In combination with the U-shaped separation channel, two additional well aligned channels to take up the excitation and collection fibers where micromachined in a glass plate. The U-cell provides a longitudinal path of 120 -140 pm in length parallel to the flow direction and can be used both for absorption and fluorescence measurements. The absorption detection limit was 0.003 AU ( 6 pM of a fluorescein dye) in the fluorescence mode a detection limit of 3 nM fluorescein (20 000 molecules) was achieved. [Pg.75]


See other pages where Fluorescence measurements flow cell is mentioned: [Pg.123]    [Pg.143]    [Pg.301]    [Pg.102]    [Pg.584]    [Pg.16]    [Pg.52]    [Pg.297]    [Pg.62]    [Pg.160]    [Pg.98]    [Pg.225]    [Pg.147]    [Pg.12]    [Pg.456]    [Pg.35]    [Pg.76]    [Pg.223]    [Pg.59]    [Pg.216]    [Pg.291]    [Pg.135]    [Pg.101]    [Pg.108]    [Pg.179]    [Pg.355]    [Pg.458]    [Pg.542]    [Pg.74]    [Pg.74]    [Pg.139]    [Pg.95]    [Pg.106]    [Pg.227]    [Pg.8]   
See also in sourсe #XX -- [ Pg.134 ]




SEARCH



Flow measurement

Flow measuring

Fluorescence cells

Fluorescence flow cells

Fluorescence measurements

Fluorescent cells

© 2024 chempedia.info