Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Feedback sensor

A = Input Element. b = Feedback Variable. d = Disturbance or Load Variable. Gh = Feedback Sensor Transfer Function. Gp = Process Transfer Function. ... [Pg.200]

Considering a widely used ball screw feed drive design, the system consists of a current amplifier, servomotor, lead screw coupling mechanism, ball screw with preloaded nut, table carrying workpiece, guide friction, and feedback sensors. Figure 1 shows the overall structure. [Pg.282]

Tat, T. L. K. (2006), Optimization of thermoset composites pultrusion with feedback sensors , M.Sc. Dissertation, School of Mechanical and Aerospace Engineering, Nanyang Technological University, Singapore. [Pg.413]

The main difference compared to the non-feedback case is that the adjusting procedure gives a much more accurate response of a seismometer within the passband than for the non-feedback sensor. Generally the high- and low-cutoff frequencies are extended by 2-3 octaves. In parallel with the feedback adjustment, a fine-timing of the output filter is done. In case the closed-loop response is actually flat over the passband, the device characteristics coincide with the output filter characteristics. This decreases the complexity of further device response presentation in terms of poles and zeroes. [Pg.954]

Some of these calibration steps have been automated in the laboratory to speed-up data collection time and reduce the required operator attention to the process. The automation of data collection requires a PC, some type of feedback sensor (i.e..three beam infra-red position sensor), and the necessary sofware to make the PC act as a typing emulator to direct the motion of the robot. [Pg.457]

Jarvis S P, Durig U, Lantz M A, Yamada H and Tokumoto H 1998 Feedback stabilized force-sensors a gateway to the direct measurement of interaction potentials Appl. Phys. A 66 S211... [Pg.1724]

Two modes of operation are available. In the simpler mode, a fixed current is appHed to the wire. In the other mode, the sensor temperature (hence is maintained at a fixed value above that of the gas by means of a feedback amplifier. In both cases, velocity is measured by the voltage drop. [Pg.110]

Technology has been introduced for on-line estimation of the kappa number based on absorption of ultraviolet light (35). This breakthrough ia optical sensor technology permits closed-loop feedback control of digesters from on-line measurement of the kappa number. [Pg.266]

Electrochemical Microsensors. The most successful chemical microsensor in use as of the mid-1990s is the oxygen sensor found in the exhaust system of almost all modem automobiles (see Exhaust control, automotive). It is an electrochemical sensor that uses a soHd electrolyte, often doped Zr02, as an oxygen ion conductor. The sensor exemplifies many of the properties considered desirable for all chemical microsensors. It works in a process-control situation and has very fast (- 100 ms) response time for feedback control. It is relatively inexpensive because it is designed specifically for one task and is mass-produced. It is relatively immune to other chemical species found in exhaust that could act as interferants. It performs in a very hostile environment and is reHable over a long period of time (36). [Pg.392]

Sensors. One growth area for electronic ceramics is in sensor appHcations. Sensors (qv) are devices that transform nonelectrical inputs into electrical outputs, thus providing environmental feedback. Smart, or intelligent, sensors also allow for mechanisms such as self-diagnosis, recovery, and adjustment for process monitoring and control (see Process control). [Pg.345]

The function of the oxygen sensor and the closed loop fuel metering system is to maintain the air and fuel mixture at the stoichiometric condition as it passes into the engine for combustion ie, there should be no excess air or excess fuel. The main purpose is to permit the TWC catalyst to operate effectively to control HC, CO, and NO emissions. The oxygen sensor is located in the exhaust system ahead of the catalyst so that it is exposed to the exhaust of aU cylinders (see Fig. 4). The sensor analyzes the combustion event after it happens. Therefore, the system is sometimes caUed a closed loop feedback system. There is an inherent time delay in such a system and thus the system is constandy correcting the air/fuel mixture cycles around the stoichiometric control point rather than maintaining a desired air/fuel mixture. [Pg.490]

Feedforward Control If the process exhibits slow dynamic response and disturbances are frequent, then the apphcation of feedforward control may be advantageous. Feedforward (FF) control differs from feedback (FB) control in that the primary disturbance or load (L) is measured via a sensor and the manipulated variable (m) is adjusted so that deviations in the controlled variable from the set point are minimized or eliminated (see Fig. 8-29). By taking control action based on measured disturbances rather than controlled variable error, the controller can reject disturbances before they affec t the controlled variable c. In order to determine the appropriate settings for the manipulated variable, one must develop mathematical models that relate ... [Pg.730]

Figure 8-41 includes two conventional feedback controllers G i controls Cl by manipulating Mi, and G o controls C9 by manipidating Mo. The output sign s from the feedback controllers serve as input signals to the two decouplers D o and D91. The block diagram is in a simplified form because the load variables and transfer functions for the final control elements and sensors have been omitted. [Pg.737]

Process-variable feedback for the controller is achieved by one of two methods. The process variable can (I) be measured and transmitted to the controller by using a separate measurement transmitter with a 0.2-I.0-bar (3-15-psi pneumatic output, or (2) be sensed directly by the controller, which contains the measurement sensor within its enclosure. Controllers with integral sensing elements are available that sense pressure, differential pressure, temperature, and level. Some controller designs have the set point adjustment knob in the controller, making set point adjustment a local and manual operation. Other types receive a set point from a remotely located pneumatic source, such as a manual air set regulator or another controller, to achieve set point adjustment. There are versions of the pneumatic controller that support the useful one-, two-, and three-mode combinations of proportional, integral, and derivative actions. Other options include auto/manual transfer stations, antireset windup circuitry, on/off control, and process-variable and set point indicators. [Pg.776]

Joyce, S.A. and Houston, J.E., A new force sensor incorporating force-feedback control for interfacial force microscopy. Rev. Sci. Instrum., 62(3), 710-715 (1991). [Pg.219]

Design alternative 1 is inherently safer because the pressure sensor provides continuous feedback to the operator. The operator has some confidence that the pressure sensor is working (although not complete... [Pg.14]

Fig. 9. Principle of single catalytic bed for simultaneous reduction and oxidation with oxygen sensor and feedback control on air-to-fuel ratio. Fig. 9. Principle of single catalytic bed for simultaneous reduction and oxidation with oxygen sensor and feedback control on air-to-fuel ratio.
Figure 23.2 The SmartPen system is a pen with an optical sensor that records each keystroke on a special form. The pen is docked at a computer or data can be wirelessly transmitted, and data from anywhere in the world are immediately sent for validation. Queries are generated within minutes, closing the feedback loop and markedly rednc-ing query rates as compared to conventional systems. Figure 23.2 The SmartPen system is a pen with an optical sensor that records each keystroke on a special form. The pen is docked at a computer or data can be wirelessly transmitted, and data from anywhere in the world are immediately sent for validation. Queries are generated within minutes, closing the feedback loop and markedly rednc-ing query rates as compared to conventional systems.
Figure 7.5. Feedback diagram for skeletal mechanical regulation. When forces are applied to a whole bone, the stimulus that results is sensed by the bone cells in the tissue. The sensor cells then signal bone-forming and -removing cells to change the geometry and material properties of the bone. Figure 7.5. Feedback diagram for skeletal mechanical regulation. When forces are applied to a whole bone, the stimulus that results is sensed by the bone cells in the tissue. The sensor cells then signal bone-forming and -removing cells to change the geometry and material properties of the bone.
Instrumentation is also fitted to provide a continuous display of important variables such as temperature and pH, the power used hy the electric motor, airflow, dissolved oxygen and exhaust gas analysis. Manual or computer feedback control can be based either directly on the signals provided hy the prohes and sensors or on derived data calculated from those signals, such as the respiratory coefficient or the rate of change of pH. Mass spectronomical analysis of exhaust gases can provide valuable physiological information. [Pg.154]

Other thermal zones which should be thermostated separately from the column oven include the Injector and detector ovens. These are generally insulted metal blocks heated by cartridge heaters controlled by sensors located in a feedback loop with the power supply. Detector blocks are usually maintained at a temperature selected to minimize detector contamination from condensation of column bleed or sample components and to optimize the response of the detector to the sample. The requirements for i injectors may be different depending on the injector design and may include provision for temperature program operation. [Pg.123]


See other pages where Feedback sensor is mentioned: [Pg.804]    [Pg.490]    [Pg.557]    [Pg.796]    [Pg.17]    [Pg.1123]    [Pg.6]    [Pg.441]    [Pg.804]    [Pg.490]    [Pg.557]    [Pg.796]    [Pg.17]    [Pg.1123]    [Pg.6]    [Pg.441]    [Pg.43]    [Pg.63]    [Pg.121]    [Pg.349]    [Pg.1689]    [Pg.2535]    [Pg.92]    [Pg.80]    [Pg.464]    [Pg.738]    [Pg.910]    [Pg.1081]    [Pg.71]    [Pg.75]    [Pg.114]    [Pg.123]    [Pg.33]    [Pg.533]    [Pg.253]    [Pg.738]    [Pg.323]   
See also in sourсe #XX -- [ Pg.10 ]




SEARCH



Sensor/actuator feedback systems

© 2024 chempedia.info