Fast Ion Bombardment FAB

The standard MAT 90 ion source is used for optimized FD/FI mode by means of the newly designed FD/FI probe. Conversion from electron impact (El), chemical ionization (Cl) or fast ion bombardment (FAB) to FD/FI operation does not require the exchange of the ion source. The FD/FI probe accommodates both the field emitter and the extraction electrodes, mounted at the probe tip. Both are introduced as a unit into the ion source through the ionization volume exchange lock without breaking vacuum. The fast and simple changeover illustrates the versatility of the Finnigan MAT 90 with no compromise on the performance. [Pg.74]

The earliest applications of MS/MS that utilized LSI methods (FAB or FIB) required placing a few microliters of a viscous glycerol/methanol matrix on a stainless steel probe, insertion into the fast-atom bombardment/ fast-ion bombardment (FAB/FIB) source housing, removal of air from the source, then further insertion of the probe to expose the sample and matrix to the atom or ion beam. Although this procedure was tedious, a... [Pg.281]

Attempts at direct MS characterisation of additives in bulk polymer samples have centred on direct thermal adsorption of additives for the bulk polymer, followed by EI-MS, chemical ionisation (CI-MS) or field ionisation (FI-MS). However, this approach is linked to polymer additives that are stable or can provide meaningful fragment ions at elevated temperatures. Desorption/ionisation methods such as fast ion bombardment (FAB) [41], laser desorption [42, 43] and secondary ion MS (SIMS) have also been applied to the analysis of additives in bulk polymer samples. However, these single step techniques suffer to varying degrees from matrix interferences in the resulting mass spectra. [Pg.19]

A connnon feature of all mass spectrometers is the need to generate ions. Over the years a variety of ion sources have been developed. The physical chemistry and chemical physics communities have generally worked on gaseous and/or relatively volatile samples and thus have relied extensively on the two traditional ionization methods, electron ionization (El) and photoionization (PI). Other ionization sources, developed principally for analytical work, have recently started to be used in physical chemistry research. These include fast-atom bombardment (FAB), matrix-assisted laser desorption ionization (MALDI) and electrospray ionization (ES). [Pg.1329]

Fast-Atom Bombardment (FAB) and Liquid-Phase Secondary Ion Mass Spectrometry (LSIMS) Ionization... [Pg.17]

The basic principles of fast-atom bombardment (FAB) and liquid-phase secondary ion mass spectrometry (LSIMS) are discussed only briefly here because a fuller description appears in Chapter 4. This chapter focuses on the use of FAB/LSIMS as part of an interface between a liquid chromatograph (LC) and a mass spectrometer (MS), although some theory is presented. [Pg.81]

The LC/TOF instmment was designed specifically for use with the effluent flowing from LC columns, but it can be used also with static solutions. The initial problem with either of these inlets revolves around how to remove the solvent without affecting the substrate (solute) dissolved in it. Without this step, upon ionization, the large excess of ionized solvent molecules would make it difficult if not impossible to observe ions due only to the substrate. Combined inlet/ionization systems are ideal for this purpose. For example, dynamic fast-atom bombardment (FAB), plas-maspray, thermospray, atmospheric-pressure chemical ionization (APCI), and electrospray (ES)... [Pg.163]

Some mild methods of ionization (e.g., chemical ionization. Cl fast-atom bombardment, FAB electrospray, ES) provide molecular or quasi-molecular ions with so little excess of energy that little or no fragmentation takes place. Thus, there are few, if any, normal fragment ions, and metastable ions are virtually nonexistent. Although these mild ionization techniques are ideal for yielding molecular mass information, they are almost useless for providing details of molecular structure, a decided disadvantage. [Pg.228]

To achieve sufficient vapor pressure for El and Cl, a nonvolatile liquid will have to be heated strongly, but this heating may lead to its thermal degradation. If thermal instability is a problem, then inlet/ionization systems need to be considered, since these do not require prevolatilization of the sample before mass spectrometric analysis. This problem has led to the development of inlet/ionization systems that can operate at atmospheric pressure and ambient temperatures. Successive developments have led to the introduction of techniques such as fast-atom bombardment (FAB), fast-ion bombardment (FIB), dynamic FAB, thermospray, plasmaspray, electrospray, and APCI. Only the last two techniques are in common use. Further aspects of liquids in their role as solvents for samples are considered below. [Pg.279]

Fast-atom bombardment (FAB) is an ionization technique that produces a protonated or deprotonated molecular ion, hence a molecular mass for the sample. It can be used for analysis of peptides up to m/z about 5000. [Pg.417]

Mass spectral analysis of quaternary ammonium compounds can be achieved by fast-atom bombardment (fab) ms (189,190). This technique rehes on bombarding a solution of the molecule, usually in glycerol [56-81-5] or y -nitroben2yl alcohol [619-25-0], with argon and detecting the parent cation plus a proton (MH ). A more recent technique has been reported (191), in which information on the stmcture of the quaternary compounds is obtained indirectly through cluster-ion formation detected via Hquid secondary ion mass spectrometry (Isims) experiments. [Pg.378]

An abundant molecular ion may indicate that an aromatic compound or highly unsaturated ring compound is present. If no molecular ion is observed and one cannot be deduced, the use of chemical ionization (ci), negative chemical ionization (nci), fast atom bombardment (FAB), or electrospray ionization (ESI) should provide a molecular ion. [Pg.20]

The mass spectrometry of diazo compounds was reviewed by Zeller (1983) and by Lebedev (1991). It is difficult to record mass spectra of diazonium salts using conventional techniques. With the water thermospray method, however, Schmelzeisen-Redeker et al. (1985) observed the diazonium ion and various fragments such as [Ar+ - N2 + 2H]+ and [Ar + N2 + H20]+. Ambroz et al. (1988) applied the fast atom bombardment (FAB) technique using a 3-nitrobenzylalcohol matrix. Peaks for ArNJ, Ar+, and [M + ArN2]+ and further peaks due to solvated ions were found. [Pg.82]

Laali and Lattimer (1989 see also Laali, 1990) observed arenediazonium ion/crown ether complexes in the gas phase by field desorption (FD) and by fast atom bombardment (FAB) mass spectrometry. The FAB-MS spectrum of benzenediazonium ion/18-crown-6 shows a 1 1 complex. In the FD spectrum, apart from the 1 1 complex, a one-cation/two-crown complex is also detected. Dicyclo-hexano-24-crown-6 appears to complex readily in the gas phase, whereas in solution this crown ether is rather poor for complexation (see earlier in this section) the presence of one or three methyl groups in the 2- or 2,4,6-positions respectively has little effect on the gas-phase complexation. With 4-nitrobenzenediazonium ion, 18-crown-6 even forms a 1 3 complex. The authors assume charge-transfer complexes such as 11.13 for all these species. There is also evidence for hydride ion transfer from the crown host within the 1 1 complex, and for either the arenediazonium ion or the aryl cation formed from it under the reaction conditions in the gas phase in tandem mass spectrometry (Laali, 1990). [Pg.301]

Fast-atom bombardment (FAB) is one of a number of ionization techniques which utilize a matrix material, in which the analyte is dissolved, to transfer sufficient energy to the analyte to facilitate ionization. In FAB, the matrix material is a liquid, such as glycerol, and the energy for ionization is provided by a high-energy atom (usually xenon) or, more recently, an ion (Cs+) beam. In conventional FAB, the solution of analyte in the matrix material is applied to the end of a probe which is placed in the source of the mass spectrometer where it is bombarded with the atom/ion beam. [Pg.54]

Mass Spectrometry. Mass spectrometry holds great promise for low-level toxin detection. Previous studies employed electron impact (El), desorption chemical ionization (DCI), fast atom bombardment (FAB), and cesium ion liquid secondary ion mass spectrometry (LSIMS) to generate positive or negative ion mass spectra (15-17, 21-23). Firm detection limits have yet to be reported for the brevetoxins. Preliminary results from our laboratory demonstrated that levels as low as 500 ng PbTx-2 or PbTx-3 were detected by using ammonia DCI and scans of 500-1000 amu (unpublished data). We expect significant improvement by manipulation of the DCI conditions and selected monitoring of the molecular ion or the ammonia adduction. [Pg.177]

Due to the high mass, low volatility, and thermal instability of chlorophylls and derivatives, molecular weight determination by electron impact (El) MS is not recommended. Desorption-ionization MS techniques such as chemical ionization, secondary ion MS, fast-atom bombardment (FAB), field, plasma- and matrix-assisted laser desorption have been very effective for molecular ion detection in the characterization of tetrapyrroles. These techniques do not require sample vaporization prior to ionization and they are effective tools for allomerization studies. [Pg.438]

Recent attention has focused on MS for the direct analysis of polymer extracts, using soft ionisation sources to provide enhanced molecular ion signals and less fragment ions, thereby facilitating spectral interpretation. The direct MS analysis of polymer extracts has been accomplished using fast atom bombardment (FAB) [97,98], laser desorption (LD) [97,99], field desorption (FD) [100] and chemical ionisation (Cl) [100]. [Pg.46]

Fast atom bombardment (FAB) Plasma desorption (PD) Liquid secondary-ion mass spectrometry (LSIMS) Thermospray (TSP)/plasmaspray (PSP) Electrohydrodynamic ionisation (EHI) Multiphoton ionisation (MPI) Atmospheric pressure chemical ionisation (APCI) Electrospray ionisation (ESI) Ion spray (ISP) Matrix-assisted laser desorption/ionisation (MALDI) Atmospheric pressure photoionisation (APPI) Triple quadrupole (QQQ) Four sector (EBEB) Hybrid (EBQQ) Hybrid (EB-ToF, Q-ToF) Tandem ToF-ToF Photomultiplier... [Pg.352]

Fast atom/ion bombardment (FAB/FtB/LSIMS) MH+, (M + H)+,M + (M - H) , M-Intact molecular ions M, Q Polar organics, proteins, organometalhcs, nonvolatiles <20000 ... [Pg.358]

Whereas the use of conventional fast atom bombardment (FAB) in the analysis of polymer/additive extracts has been reported (see Section 6.2.4), the need for a glycerol (or other polar) matrix might render FAB-MS analysis of a dissolved polymer/additive system rather unattractive (high chemical background, high level of matrix-, solvent- and polymer-related ions, complicated spectra). Yet, in selected cases the method has proved quite successful. Lay and Miller [53] have developed an alternative method to the use of sample extraction, cleanup, followed by GC in the quantitative analysis of PVC/DEHP with plasticiser levels as typically found in consumer products (ca. 30 %). The method relied on addition of the internal standard didecylphthalate (DDP) to a THF solution of the PVC sample with FAB-MS quantitation based on the relative signal levels of the [MH]+ ions of DEHP and DDP obtained from full-scan spectra, and on the use of a calibration curve (intensity ratio m/z 391/447 vs. mg DEHP/mg DDP). No FAB-matrix was added. No ions associated with the bulk of the PVC polymer were observed. It was... [Pg.702]

Mass spectrometry (MS) in its various forms, and with various procedures for vaporization and ionization, contributes to the identification and characterization of complex species by their isotopomer pattern of the intact ions (usually cation) and by their fragmentation pattern. Upon ionization by the rough electron impact (El) the molecular peak often does not appear, in contrast to the more gentle field desorption (FD) or fast-atom bombardment (FAB) techniques. An even more gentle way is provided by the electrospray (ES) method, which allows all ionic species (optionally cationic or anionic) present in solution to be detected. Descriptions of ESMS and its application to selected problems are published 45-47 also a representative application of this method in a study of phosphine-mercury complexes in solution is reported.48... [Pg.1256]

Fast atom bombardment FAB Particle induced desorption/ ionization Nonvolatile molecular ions Soft method Large molecules... [Pg.18]


See other pages where Fast Ion Bombardment FAB is mentioned: [Pg.4379]    [Pg.184]    [Pg.4379]    [Pg.184]    [Pg.1331]    [Pg.177]    [Pg.287]    [Pg.545]    [Pg.143]    [Pg.578]    [Pg.468]    [Pg.493]    [Pg.372]    [Pg.237]    [Pg.260]    [Pg.47]    [Pg.427]    [Pg.144]   


SEARCH



Bombardment

FAB

Fast Atom Bombardment (FAB) and Liquid-matrix Secondary Ion Mass Spectrometry (LSIMS)

Fast atom bombardment (FAB) and liquid secondary ion mass spectrometry (LSIMS)

Fast-ion bombardment

Ion bombardement

Ion bombardment

© 2024 chempedia.info