Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

External References

In this equation and t are observable. The pressure of the flavor must be obtained from an external reference, and D and S must be measured or estimated. If steady state is achieved, the quantity sorbed is given by equation 11. In either case the quantity sorbed can be important. [Pg.493]

The characterization internal and external, refers more to the analytical procedure than to the plant configuration. [Pg.236]

Transport numbers are intended to measure the fraction of the total ionic current carried by an ion in an electrolyte as it migrates under the influence of an applied electric field. In essence, transport numbers are an indication of the relative ability of an ion to carry charge. The classical way to measure transport numbers is to pass a current between two electrodes contained in separate compartments of a two-compartment cell These two compartments are separated by a barrier that only allows the passage of ions. After a known amount of charge has passed, the composition and/or mass of the electrolytes in the two compartments are analyzed. Erom these data the fraction of the charge transported by the cation and the anion can be calculated. Transport numbers obtained by this method are measured with respect to an external reference point (i.e., the separator), and, therefore, are often referred to as external transport numbers. Two variations of the above method, the Moving Boundary method [66] and the Eiittorff method [66-69], have been used to measure cation (tR+) and anion (tx ) transport numbers in ionic liquids, and these data are listed in Table 3.6-7. [Pg.121]

Because an increase in resolution causes a decrease in sensitivity, it is best to operate at the lowest resolution commensurate with good results. Some instrument data systems will allow calibration with an external reference material such as perfluorokerosene and then use of a secondary reference material for the internal mass reference. Tetraiodothiophene, vaporized using the solids probe inlet, is recommended as the secondary reference. The accurate masses are 79.9721, 127.9045, 162.9045, 206.8765, 253.8090, 293.7950, 333.7810, 460.6855, and 587.5900. For a higher mass standard, use hexaiodobenzene. Because the mass defect for these internal reference ions are so large, a resolution of 2000 is ample to separate these ions from almost any sample ions encountered in GC/MS. [Pg.375]

The potential of die electrode is registered with respect to the external reference electrode. Hence, die cell potential (at 25°C and after introducing the definition of pH) follows the relation... [Pg.149]

A representative ISE is shown schematically in Fig. 1. The electrode consists of a membrane, an internal reference electrolyte of fixed activity, (ai)i , ai and an internal reference electrode. The ISE is immersed in sample solution that contains analyte of some activity, (ajXampie and into which an external reference electrode is also immersed. The potential measured by the pH/mV meter (Eoe,) is equal to the difference in potential between the internal (Eraf.int) and external (Eref.ext) reference electrodes, plus the membrane potential (E emb), plus the liquid junction potential... [Pg.4]

E,j) that exists at the junction between the external reference electrode and the sample solution. [Pg.5]

The present state of the art in blood pH measurements allows for rapid (1 minute) determination of pH between 6.4 and 8.0 to within at least 0.005 units for whole blood sample volumes < 100 microliters. The temperature of the electrodes and sample is generally controlled to within 0.1 °C for this level of precision and frequent calibration is carried out (in some cases a one point calibration for each sample). The electrodes require (both the glass and external reference) some maintenance due to protein fouling, however this procedure is largely automated. The useful life of an electrode is one year or less and the cost is well over 100 (U.S.) each. New technologies, both electrochemical and non-electrochemical, must compete with this attractive performance and provide for lower operating costs in order to be successful. [Pg.52]

Without adequate silanization, an aqueous film can short-circuit the d-pipette by linking its two halves. In the absence of a surface aqueous layer the currents flow between each of two pipettes and the external reference electrode. In contrast, when the liquid film connects two orifices, the current flows directly from one barrel to the other. Two experiments allowing the distinction between these two cases were carried out using a well-characterized reaction of potassium transfer from water into DCE facilitated by dibenzo-18-crown-6 (DB18C6) [11]... [Pg.390]

It has become fairly common to adopt the manufacture of combinations of internal reference electrode and its inner electrolyte such that the (inner) potential at the glass electrode lead matches the (outer) potential at the external reference electrode if the glass electrode has been placed in an aqueous solution of pH 7. In fact, each pH glass electrode (single or combined) has its own iso-pH value or isotherm intersection point ideally it equals 0 mV at pH 7 0.5 according to a DIN standard, as is shown in Fig. 2.11 the asymmetry potential can be easily eliminated by calibration with a pH 7.00 0.02 (at 25° C) buffer solution. [Pg.77]

Whether the assumption about this linear relationship can be used for the zero shift as such is doubtful the situation becomes more reliable if the internal and external reference electrodes are equal so that E°mer and °uter cancel, hence eqn. 2.95 becomes En = (- 2.3026RT/F) pHinncr. Therefore, the zero shift can be eliminated instrumentally by setting the mechanical zero of the pH meter to pHjnncr (if previously known). With a non-combined glass electrode the external... [Pg.91]

Two different and possibly complementary approaches have been explored. One utilizes a panel of quantifiable internal reference standards (QIRS), which are common proteins present widely in tissues in relatively consistent amounts.11,22 In this instance because the reference proteins are intrinsic to the tissue they are necessarily subjected to identical fixation and processing, and incur no additional handling or cost, other than synchronous performance of a second IHC assay (stain), such that the intensity of reaction for the QIRS and the test analyte can be compared by IA, allowing calculation of the amount of test analyte (protein) present on a formulaic standard curve basis. The other approach seeks to identify external reference materials and to introduce these into each step of tissue preparation for cases where IHC studies are anticipated in this instance the logistical issues of production, distribution, and inclusion of the reference standard into all phases of tissue processing also must be considered, along with attendant costs. [Pg.81]

The integrated planar silver chloride electrode uses a thin layer of 150 pm polymer that consists of a heat curing epoxy resin poly-hydroxy-ethylmethacrylate (PHEMA) to immobilize the KC1 electrolyte. The potential drift of the reference electrode reduced to 59 pV/h after a conditioning phase of several hours. However, this reference electrode was only used for P02 measurement, while an external reference electrode was used for pH measurement. [Pg.305]

After the pioneering work of Tao [111], a number of studies reported how to control the current flow via an external reference electrode in electrochemical STMJs andLAJs [93, 112-143]. [Pg.96]

The bulb is usually fabricated with common soda glass, i.e. glass containing a high concentration of sodium ions. Finally, a small reference electrode, such as an SCE, is positioned beside the bulb. For this reason, the pH electrode ought properly to be called a pH combination electrode, because it is combined with a reference electrode. If the pH electrode does not have an SCE, it is termed a glass electrode (GE). The operation of a glass electrode is identical to that of a combination pH electrode, except that an external reference electrode is required. [Pg.336]

It is interesting to note that when using two fluorophores (whose fluorescence decays are known to be single exponentials), one as a sample and the other as a reference, it is possible to determine the lifetimes of both the fluorophores without external reference this can be achieved in data analysis by varying the reference lifetime until a minimum value of x is reached. [Pg.189]

Chemical shift values upheld (from HCONH2 as external reference) in 1.7 M Me2CO-d solutions. [Pg.308]


See other pages where External References is mentioned: [Pg.67]    [Pg.70]    [Pg.195]    [Pg.421]    [Pg.462]    [Pg.566]    [Pg.595]    [Pg.617]    [Pg.148]    [Pg.165]    [Pg.130]    [Pg.12]    [Pg.52]    [Pg.209]    [Pg.224]    [Pg.283]    [Pg.390]    [Pg.781]    [Pg.782]    [Pg.668]    [Pg.753]    [Pg.66]    [Pg.223]    [Pg.153]    [Pg.78]    [Pg.79]    [Pg.100]    [Pg.96]    [Pg.658]    [Pg.125]    [Pg.105]    [Pg.632]   
See also in sourсe #XX -- [ Pg.72 , Pg.108 ]

See also in sourсe #XX -- [ Pg.89 , Pg.91 ]

See also in sourсe #XX -- [ Pg.62 ]

See also in sourсe #XX -- [ Pg.185 ]




SEARCH



Chemical shift external reference

External Reference by Selecting

External Reference by Using Paste Link

External reference compounds

External reference electrode

External reference pricing

No external references

Reference electrode external pressure-balanced

Using External References

© 2024 chempedia.info