Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Reference Electrodes potential

Figure 7. SNIFTIRS difference spectra for 1.3 x 10" mol-dm" isoquinoline solution at a mercury electrode. Reference potential was 0.0 V vs. SCE, sample potential (a) -0.60V,... Figure 7. SNIFTIRS difference spectra for 1.3 x 10" mol-dm" isoquinoline solution at a mercury electrode. Reference potential was 0.0 V vs. SCE, sample potential (a) -0.60V,...
Working electrode reference potential scan rate. [Pg.625]

Finding the End Point Potentiometrically Another method for locating the end point of a redox titration is to use an appropriate electrode to monitor the change in electrochemical potential as titrant is added to a solution of analyte. The end point can then be found from a visual inspection of the titration curve. The simplest experimental design (Figure 9.38) consists of a Pt indicator electrode whose potential is governed by the analyte s or titrant s redox half-reaction, and a reference electrode that has a fixed potential. A further discussion of potentiometry is found in Chapter 11. [Pg.339]

Potentiometric measurements are made using a potentiometer to determine the difference in potential between a working or, indicator, electrode and a counter electrode (see Figure 11.2). Since no significant current flows in potentiometry, the role of the counter electrode is reduced to that of supplying a reference potential thus, the counter electrode is usually called the reference electrode. In this section we introduce the conventions used in describing potentiometric electrochemical cells and the relationship between the measured potential and concentration. [Pg.466]

Potentiometric electrochemical cells are constructed such that one of the half-cells provides a known reference potential, and the potential of the other half-cell indicates the analyte s concentration. By convention, the reference electrode is taken to be the anode thus, the shorthand notation for a potentiometric electrochemical cell is... [Pg.471]

The reference potential of the Ag-AgCl electrode in brackish water must be eorreeted for ehloride ion eontent (i.e., a ehange in ehloride ion eoneentration by a faetor of 10 shifts the referenee potential by about 50 mV in the positive direetion... [Pg.368]

Measuring electrodes for impressed current protection are robust reference electrodes (see Section 3.2 and Table 3-1) which are permanently exposed to seawater and remain unpolarized when a small control current is taken. The otherwise usual silver-silver chloride and calomel reference electrodes are used only for checking (see Section 16.7). All reference electrodes with electrolytes and diaphragms are unsuitable as long-term electrodes for potential-controlled rectifiers. Only metal-medium electrodes which have a sufficiently constant potential can be considered as measuring electrodes. The silver-silver chloride electrode has a potential that depends on the chloride content of the water [see Eq. (2-29)]. This potential deviation can usually be tolerated [3]. The most reliable electrodes are those of pure zinc [3]. They have a constant rest potential, are slightly polarizable and in case of film formation can be regenerated by an anodic current pulse. They last at least 5 years. [Pg.408]

The electrolysis is carried out at a reference potential of -2.4 volts vs a standard calomel electrode. An initial current density of 0.0403 amp/cm is obtained which drops to 0.0195 amp/cm at the end of the reduction, which is carried on over a period of 1,682 minutes at 15° to 20°C. The catholyte is filtered, the solid material is washed with water and dried. 430 g of the 2,3-bis-(3-pyridyl)-butane-2,3-diol is recrystallized from water, MP 244° to 245°C. [Pg.1013]

However, in the case of stress-corrosion cracking of mild steel in some solutions, the potential band within which cracking occurs can be very narrow and an accurately known reference potential is required. A reference half cell of the calomel or mercury/mercurous sulphate type is therefore used with a liquid/liquid junction to separate the half-cell support electrolyte from the process fluid. The connections from the plant equipment and reference electrode are made to an impedance converter which ensures that only tiny currents flow in the circuit, thus causing the minimum polarisation of the reference electrode. The signal is then amplifled and displayed on a digital voltmeter or recorder. [Pg.33]

Finally, calomel electrodes (and more especially hydrogen electrodes) are not suitable for field measurements because they are not sufficiently robust. The calomel electrodes are however essential for calibrating the field reference electrodes. Saturated KCI calomel electrodes are the most suitable because there is then no doubt about the reference potential of the calibrating electrode. Lack of adequate calibration is a common cause of cathodic protection system mismanagement. [Pg.124]

The corrosion process is observed as a series of events which all contribute to the overall corrosion rate. Measurement of rest potential fluctuations between two identical electrodes of potential fluctuations with respect to a fixed reference can be carried out. The electrochemical noise output spectrum is analysed using digitised data. The interpretation requires electrochemical expertise, and the method is therefore usually provided as a specialised service. [Pg.1140]

This procedure of using a single measurement of electrode potential to determine the concentration of an ionic species in solution is referred to as direct potentiometry. The electrode whose potential is dependent upon the concentration of the ion to be determined is termed the indicator electrode, and when, as in the case above, the ion to be determined is directly involved in the electrode reaction, we are said to be dealing with an electrode of the first kind . [Pg.548]

Logarithmic scale for expressing acidity or alkalinity of water (7.0 to 0 indicates increasing acidity 7.0 to 14 indicates increasing alkalinity). Measured by means of a glass electrode/reference electrode pair immersed in the water sample under test. The potential difference depends upon the pH which is then displayed on a pH meter (high input impedance, millivoltmeter). [Pg.749]

Changes in the reference electrode junction potential result from differences in the composition of die sample and standard solutions (e.g., upon switching from whole blood samples to aqueous calibrants). One approach to alleviate this problem is to use an intermediate salt bridge, with a solution (in the bridge) of ions of nearly equal mobility (e.g., concentrated KC1). Standard solutions with an electrolyte composition similar to that of the sample are also desirable. These precautions, however, will not eliminate the problem completely. Other approaches to address this and other changes in the cell constant have been reviewed (13). [Pg.147]

When the potential of the electrode is E with respect to a reference electrode, the potential of the surface of the working electrode will be ... [Pg.158]

In addition to defined standard conditions and a reference potential, tabulated half-reactions have a defined reference direction. As the double arrow in the previous equation indicates, E ° values for half-reactions refer to electrode equilibria. Just as the value of an equilibrium constant depends on the direction in which the equilibrium reaction is written, the values of S ° depend on whether electrons are reactants or products. For half-reactions, the conventional reference direction is reduction, with electrons always appearing as reactants. Thus, each tabulated E ° value for a half-reaction is a standard reduction potential. [Pg.1383]


See other pages where Reference Electrodes potential is mentioned: [Pg.74]    [Pg.77]    [Pg.74]    [Pg.77]    [Pg.1941]    [Pg.2720]    [Pg.462]    [Pg.465]    [Pg.777]    [Pg.466]    [Pg.395]    [Pg.51]    [Pg.51]    [Pg.55]    [Pg.63]    [Pg.765]    [Pg.385]    [Pg.33]    [Pg.124]    [Pg.124]    [Pg.246]    [Pg.1190]    [Pg.1250]    [Pg.63]    [Pg.510]    [Pg.573]    [Pg.595]    [Pg.105]    [Pg.147]    [Pg.118]    [Pg.179]    [Pg.410]    [Pg.412]    [Pg.81]    [Pg.280]   
See also in sourсe #XX -- [ Pg.8 , Pg.113 ]

See also in sourсe #XX -- [ Pg.8 , Pg.113 ]




SEARCH



Calomel electrodes, reference electrode standard potential

Electrochemical potential reference electrodes

Ferrocenium/ferrocene redox potential reference electrode

Organic solvent-water mixtures, reference electrode potentials

Potential Measurement Reference Electrodes and Electrometers (Ref

Potential measurement reference electrodes

Potentials of reference electrodes

Reference electrode (half cell) potential measurements

Reference electrode oxidizing potential

Reference electrode potential conversion scale

Reference electrodes

Reference electrodes electrode potential

Reference electrodes electrode potential

Reference electrodes reduction/oxidation potential evaluation

Reference electrodes reversible potential

Reference potential

Reference-electrode potentials, temperature

Reference-electrode potentials, temperature coefficients

Thermodynamics electrode potential references

© 2024 chempedia.info