Indicator electrode — In -> potentiometry the electrode for which the -> potential is recorded, in -> amperometry the electrode for which the -> current is recorded. [Pg.351]

The functional grouping is as follows. The electrode which is under study is called working electrode in voltammetry or indicator electrode in potentiometry. The electrode the potential of which is practically constant and used to make comparison of electrode potentials, i.e., to define the value of the potential of the electrode on the scale based on standard hydrogen electrode, is called a reference electrode. The electrode that serves to maintain the current in the circuit formed with the working electrode in voltammetric experiments in three-compartment cells is the auxiliary (or counter) electrode. [Pg.4]

When first developed, potentiometry was restricted to redox equilibria at metallic electrodes, limiting its application to a few ions. In 1906, Cremer discovered that a potential difference exists between the two sides of a thin glass membrane when opposite sides of the membrane are in contact with solutions containing different concentrations of H3O+. This discovery led to the development of the glass pH electrode in 1909. Other types of membranes also yield useful potentials. Kolthoff and Sanders, for example, showed in 1937 that pellets made from AgCl could be used to determine the concentration of Ag+. Electrodes based on membrane potentials are called ion-selective electrodes, and their continued development has extended potentiometry to a diverse array of analytes. [Pg.465]

Potentiometric measurements are made using a potentiometer to determine the difference in potential between a working or, indicator, electrode and a counter electrode (see Figure 11.2). Since no significant current flows in potentiometry, the role of the counter electrode is reduced to that of supplying a reference potential thus, the counter electrode is usually called the reference electrode. In this section we introduce the conventions used in describing potentiometric electrochemical cells and the relationship between the measured potential and concentration. [Pg.466]

The potential of the indicator electrode in a potentiometric electrochemical cell is proportional to the concentration of analyte. Two classes of indicator electrodes are used in potentiometry metallic electrodes, which are the subject of this section, and ion-selective electrodes, which are covered in the next section. [Pg.473]

The potential of a metallic electrode is determined by the position of a redox reaction at the electrode-solution interface. Three types of metallic electrodes are commonly used in potentiometry, each of which is considered in the following discussion. [Pg.473]

Buck, R. P. Potentiometry pH Measurements and Ion Selective Electrodes. In Weissberger, A., ed.. Physical Methods of Organic Chemistry, Vol. 1, Part IIA. Wiley New York, 1971, pp. 61-162. Cammann, K. Working with Ion-Selective Electrodes. Springer-Verlag Berlin, 1977. [Pg.541]

Buck, R. P. Potentiometry, pH measurements and ion-selective electrodes, in Physical Methods of Chemistry, part Ha (eds.) Weissberger, A., Rossiter, B. W., New York, Interscience 1971... [Pg.257]

Electrical methods of analysis (apart from electrogravimetry referred to above) involve the measurement of current, voltage or resistance in relation to the concentration of a certain species in solution. Techniques which can be included under this general heading are (i) voltammetry (measurement of current at a micro-electrode at a specified voltage) (ii) coulometry (measurement of current and time needed to complete an electrochemical reaction or to generate sufficient material to react completely with a specified reagent) (iii) potentiometry (measurement of the potential of an electrode in equilibrium with an ion to be determined) (iv) conductimetry (measurement of the electrical conductivity of a solution). [Pg.7]

Enzyme sensors are based primarily on the immobilization of an enzyme onto an electrode, either a metallic electrode used in amperometry (e.g., detection of the enzyme-catalyzed oxidation of glucose) or an ISE employed in potentiometry (e.g., detection of the enzyme-catalyzed liberation of hydronium or ammonium ions). The first potentiometric enzyme electrode, which appeared in 1969 due to Guilbault and Montalvo [140], was a probe for urea with immobilized urease on a glass electrode. Hill and co-workers [141] described in 1986 the second-generation biosensor using ferrocene as a mediator. This device was later marketed as the glucose pen . The development of enzyme-based sensors for the detection of glucose in blood represents a major area of biosensor research. [Pg.340]

In measurements of conductivity, no electrochemical reactions occur. Differences in conductivity are due to differences in the ionic strengths of solutions. An alternating potential is applied to the solution at a known potential. The current is measured and the conductivity in Siemens/cm calculated.16 In potentiometry, the analyte is presumed to undergo no electrochemical reaction. The potential at the electrode changes due to changes in potential across the surface of the membrane in a membrane electrode or at the electrode surface of a solid electrode. The most familiar example of a potentiometric electrode is the pH electrode. In amperometry, current does flow, due to reduction or oxidation of the substance being analyzed. [Pg.219]

This method is primarily concerned with the phenomena that occur at electrode surfaces (electrodics) in a solution from which, as an absolute method, through previous calibration a component concentration can be derived. If desirable the technique can be used to follow the progress of a chemical reaction, e.g., in kinetic analysis. Mostly, however, potentiometry is applied to reactions that go to completion (e.g. a titration) merely in order to indicate the end-point (a potentiometric titration in this instance) and so do not need calibration. The overwhelming importance of potentiometry in general and of potentiometric titration in particular is due to the selectivity of its indication, the simplicity of the technique and the ample choice of electrodes. [Pg.42]

As in normal potentiometry one uses and indicator electrode versus a reference electrode, the electrodes should, especially in pH measurements, be those recommended by the supplier of the pH meter in order to obtain a direct reading of the pH value displayed. In redox or other potential measurements any suitable reference electrode of known potential can be applied. However, a reference electrode is only suitable if a junction potential is excluded, e.g., an Ag-AgCl electrode in a solution of fixed Ag+ concentration or a calomel electrode in a saturated KC1 solution as a junction in many instances a direct contact of Cl" with the solution under test (possibly causing precipitation therein) is not allowed, so that an extra or so-called double junction with KN03 solution is required. Sometimes micro-electrodes or other adaptations of the surface are required. [Pg.86]

The electrochemical detection of pH can be carried out by voltammetry (amper-ometry) or potentiometry. Voltammetry is the measurement of the current potential relationship in an electrochemical cell. In voltammetry, the potential is applied to the electrochemical cell to force electrochemical reactions at the electrode-electrolyte interface. In potentiometry, the potential is measured between a pH electrode and a reference electrode of an electrochemical cell in response to the activity of an electrolyte in a solution under the condition of zero current. Since no current passes through the cell while the potential is measured, potentiometry is an equilibrium method. [Pg.287]

A schematic diagram of a typical pH electrode system is shown in Fig. 10.1. The cell potential, i.e. the electromotive force, is measured between a pH electrode and a reference electrode in a test solution. The pH electrode responds to the activity or concentration of hydrogen ions in the solution. The reference electrode has a very stable half-cell potential. The most commonly used reference electrodes for potentiometry are the silver/silver chloride electrodes (Ag/AgCl) and the saturated calomel electrodes (SCE). [Pg.288]

In potentiometry, the variation of the potential of a Pt electrode relative to a calomel reference electrode represents the time-dependent bromine concentration. Available [Br2] is about 2 x 10-5-10-4 m pseudo-first-order conditions ([Ol] [Br2]) have to be used. Rate constants up to 104 5 m- 1 s 1 can thus be obtained (Atkinson and Bell, 1963 Dubois et ai, 1968). [Pg.216]

Ion-selective electrode (ISE) In potentiometry, an electrode having a nemstian response to one ion, ideally to the exclusion of others. [Pg.340]

In potentiometry with ion-selective electrodes (ISE), as a rule, it is the electromotive force of a cell such as the following that is measured. [Pg.33]

As stated on p. 28, an analytically ideal sensor would determine the deter-minand both specifically and quantitatively. In potentiometry, this would require an electrode sensitive to one single substance among all the components of the system. [Pg.62]

The most common analytical methods used were gas chromatography, HPLC, AA spectrophotometry, polarography, colorimetry, and potentiometry with ion-selective electrodes. In this study GC/MS and other more expensive instrumentation were avoided. If sorbent tubes could not be used for gaseous substances, then the less desirable miniature bubblers or impingers were considered. Although these devices are inconvenient they were often used because no better alternatives were available. Bags were used in a few cases where the analyte could not be retained on a sorbent because of volatility and a small tendency to sorb. Filters were used for particulates. Combinations of collection devices were used if we felt that both particulates and vapor might be present in the analyte. [Pg.11]

S. L. Truman, Potentiometry pH and ion-selective electrode, in W. E. Galen, Analytical Instrument Handbook, Marcel Dekker, New York, 1997. [Pg.242]

In potentiometry, we measure the emf of a cell consisting of an indicator electrode and a reference electrode. For emf measurements, we generally use a pH/ mV meter of high input impedance. The potential of the reference electrode must be stable and reproducible. If there is a liquid junction between the indicator electrode and the reference electrode, we should take the liquid junction potential into account. [Pg.167]

When three-electrode devices are used, reference electrodes similar to those in potentiometry (Section 6.1.2) are applicable, because no appreciable current flows through them. The reference electrodes used in non-aqueous solutions can be classified into two groups [1, 2, 5, 10]. Reference electrodes of the first group are prepared by using the solvent under study and those of the second group are... [Pg.224]

Hg2 ion-selective electrode calibration curve from J. A. Shatkin, H. S. Brown, and S. licht, Composite Graphite Ion Selective Electrode Array Potentiometry for the Detection of Mercury and Other Relevant Ions in Aquatic Systems, Anal. Chem. 1995, 67,1147. It was not stated in the paper, but we presume that all solutions had the same ionic strength. [Pg.326]

Acetonitrile interacts with the d10 metal ions Cu1 and Ag1 to form solvated species of marked stability. This stability has been used in potentiometry where the Ag, 0.01 M AgN03 couple in acetonitrile has been recommended as a reversible reference electrode.154... [Pg.797]

Table 5), and several are now being used, or are potentially useful, for measuring key ocean elements. The most common use of direct potentiometry (as compared with potentiometric titrations) is for measurement of pH (Culberson, 1981). Most other cation electrodes are subject to some degree of interference from other major ions. Electrodes for sodium, potassium, calcium, and magnesium have been used successfully. Copper, cadmium, and lead electrodes in seawater have been tested, with variable success. Anion-selective electrodes for chloride, bromide, fluoride, sulfate, sulfide, and silver ions have also been tested but have not yet found wide application. [Pg.50]