Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Current to voltage converter

The electrochemical circuitry required for SECM is relatively straightforward. Since the interface is not generally externally polarized in SECM measurements of liquid-liquid interfaces, a simple two-electrode system suffices (Fig. 3). A potential is applied to the tip, with respect to a suitable reference electrode, to drive the process of interest at the tip and the corresponding current that flows is typically amplified by a current-to-voltage converter. [Pg.295]

Any device that can monitor radiation can be used as detector. In most cases, end-on and side-on photomultiplier tubes (PMT) are used. The end-on PMT has the benefit of circular photocathode, which views more effectively the radiation emitted from the circular flow cells. Nonetheless, the side-on PMTs are more commonly used due to their lower cost. The signal from the PMT is current and, hence, in most applications a current-to-voltage converter is required to convert the current to voltage, which is then monitored. [Pg.339]

Figure 11 Schematic diagram of a batch chemiluminometer. H.V., high-voltage power supply I/V, current-to-voltage converter. Figure 11 Schematic diagram of a batch chemiluminometer. H.V., high-voltage power supply I/V, current-to-voltage converter.
The electrode potential was controlled with an EG G Princeton Applied Research (PAR) model 173 potentiostat/galvanostat and is referenced to a saturated calomel electrode (SCE). A PAR model 276 current-to-voltage converter allowed monitoring of current during the ORC and SERS experiments and it also provided for positive feedback iR compensation for accurate potential control. [Pg.397]

The current noise signal was monitored by using a sensitive, low noise zero resistance ammeter (ZRA) to couple pairs of identical electrodes the ZRA acting as a current to voltage converter. This derived potential signal was then fed into a potential noise monitor. [Pg.39]

Figure 6.7 Evolution of a potentiostat (continued), (a) Equivalent circuit of a three-electrode cell, (b) Addition of current-to-voltage converter. Figure 6.7 Evolution of a potentiostat (continued), (a) Equivalent circuit of a three-electrode cell, (b) Addition of current-to-voltage converter.
Figure 6.8 Two-amplifier adder-type potentiostat with current-to-voltage converter. Figure 6.8 Two-amplifier adder-type potentiostat with current-to-voltage converter.
To carry out amperometric or voltammetric experiments simultaneously at different electrodes in the same solution is not difficult. In principle, any number of working electrodes could be studied however, it is unlikely that more than two or three would ever be widely used in practice. The bulk of the solution can have only one controlled potential at a time (if there are significant iR drops, there will be severe control problems with multiple-electrode devices). It is necessary to use a single reference electrode to monitor the difference between this inner solution potential and the inner potential of W1 at the summing point of an operational amplifier current-to-voltage converter (this is the potential of the circuit common see OA-2 in Fig. 6.17). The potential difference between... [Pg.185]

If the external potential of W2 were at the potential of the circuit common, it would necessarily have the same interfacial potential as Wl, since the bulk solution potential would also be the same for both electrodes. The remedy is to float the circuit common for the second (third, etc.) current-to-voltage converter (OA-3) to a potential that is the difference between the desired electrode potentials for Wl and W2 (Wl and W3, etc.). [Pg.186]

The arrival of large-scale integrated circuits in the last 20 years has revolutionized chemical instrumentation just as it has kitchens, automobiles, and television sets. With respect to electrochemistry, the microprocessor has been incorporated in signal generation and data processing, while the basic instrumentation (e.g., potentiostat and current-to-voltage converter) remains as described in earlier sections of this chapter. Microprocessor instruments provide flexibility... [Pg.189]

Figure 7.1 (A) Typical controlled-potential circuit and cell OA1, the control amplifier OA2, the voltage follower (Vr = Er) OA3, the current-to-voltage converter. (B) Equivalent circuit of cell Rc, solution resistance between auxiliary and working electrodes Ru, solution resistance between reference and working electrodes, Rs = Rc + Ru and Cdl, capacitance of interface between solution and working electrode. (C) Equivalent circuit with the addition of faradaic impedance Zf due to charge transfer. Potentials are relative to circuit common, and working electrode is effectively held at circuit common (Ew = 0) by OA3. Figure 7.1 (A) Typical controlled-potential circuit and cell OA1, the control amplifier OA2, the voltage follower (Vr = Er) OA3, the current-to-voltage converter. (B) Equivalent circuit of cell Rc, solution resistance between auxiliary and working electrodes Ru, solution resistance between reference and working electrodes, Rs = Rc + Ru and Cdl, capacitance of interface between solution and working electrode. (C) Equivalent circuit with the addition of faradaic impedance Zf due to charge transfer. Potentials are relative to circuit common, and working electrode is effectively held at circuit common (Ew = 0) by OA3.
The tunneling current is measured with a current-to-voltage converter and is used to measure the distance between the tip and the sample. The current noise sources in the current-to-voltage converter have to be small enough such that the corresponding vertical noise SzpT is considerably smaller than the atomic corrugation of the sample. [Pg.73]

A potentiostat as a laboratory device is normally equipped with further features such as monitoring of the -> reference electrode, current detection by current-to-voltage-converters, or differential amplifiers, pulse form generators, displays of current and potential, and computer interfaces. See also - IRU potential drop and - IR drop compensation, - Randles. [Pg.545]

Figure 3.5. Engineering drawing of a solid-state photodetector with integrated amplifier, current-to-voltage converter, and temperature controller, (courtesy PerkinElmer Optoelectronics)... Figure 3.5. Engineering drawing of a solid-state photodetector with integrated amplifier, current-to-voltage converter, and temperature controller, (courtesy PerkinElmer Optoelectronics)...
In other words, the current in the electrochemical cell is proportional to the voltage output of the op amp. The value of the current can then be calculated from the measured values of and the resistance R. The circuit is called a current-to-voltage converter. [Pg.669]

Current-to-voltage converter A device for converting an electric current into a voltage that is proportional to the circuit. Cuvette The container that holds the analyte in the Light path in absorption spectroscopy. [Pg.1106]

Electrochemical experiments involved the use of a potentiostat/galvanostat, a current-to-voltage converter, and a universal programmer (PAR Models 173, 176 or 179, and 175). Results were displayed using either an X-Y recorder (Hewlett-Packard 7047A) or a digital oscilloscope (Nicolet 4094 A). The Pd-H versus SCE potentials were measured with both the PAR potentiostat and a digital multimeter (Keithley 175). [Pg.172]

ACIS measurements were performed at frequencies between 1 mHz and 1 kHz using a Solartron Model 1250 Frequency Response Analyzer. Output from the comb specimens was amplified with a Keithley Model 427 Current-to-Voltage Converter before waveform analysis. Reference electrodes were not used owing to the geometry of the encapsulated test specimens. The data reported herein were obtained with a 0.1 V rms amplitude sinusoidal excitation waveform. In one experiment, DC bias was superimposed on this waveform. [Pg.319]

As an example, for a UV-Vis spectrophotometer, the signal generator is the radiation source, the input transducer is the photodetector, the electronic signal modifier is a current-to-voltage converter and the output transducer is the digital display. [Pg.4]

Schematic diagrams of the instrumentation for the ANL capacitive flowmeter are given in Fig. 6.20. A 100-kHz sine-wave oscillator, with stable frequency and amplitude controls, was used to pulse the drive electrode. Each sensing electrode was connected to a current-to-voltage converter preamplifier. The preamplifier outputs were bandpass filtered at 100 kHz 5 Hz and amplitude-demodulated. The demodulated signals were amplified and DC-coupled to a first-order low-pass filter to give density signals. Schematic diagrams of the instrumentation for the ANL capacitive flowmeter are given in Fig. 6.20. A 100-kHz sine-wave oscillator, with stable frequency and amplitude controls, was used to pulse the drive electrode. Each sensing electrode was connected to a current-to-voltage converter preamplifier. The preamplifier outputs were bandpass filtered at 100 kHz 5 Hz and amplitude-demodulated. The demodulated signals were amplified and DC-coupled to a first-order low-pass filter to give density signals.

See other pages where Current to voltage converter is mentioned: [Pg.669]    [Pg.11]    [Pg.107]    [Pg.159]    [Pg.161]    [Pg.339]    [Pg.175]    [Pg.176]    [Pg.177]    [Pg.178]    [Pg.83]    [Pg.57]    [Pg.329]    [Pg.121]    [Pg.172]    [Pg.538]    [Pg.136]    [Pg.324]    [Pg.669]    [Pg.2972]    [Pg.443]    [Pg.155]    [Pg.403]    [Pg.405]    [Pg.242]   
See also in sourсe #XX -- [ Pg.145 ]

See also in sourсe #XX -- [ Pg.668 ]

See also in sourсe #XX -- [ Pg.229 ]




SEARCH



Current converters

Current-voltage

© 2024 chempedia.info