Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Constant-current coulometry titrations

Figure 4.19 A constant current coulometry titration cell. The reagent is produced at the working electrode and reacts with the sample. The indicator electrodes detect the changing potential or conductivity of the solution and the amount of change that takes place is measured and related to the concentration of the reactant in the sample. Figure 4.19 A constant current coulometry titration cell. The reagent is produced at the working electrode and reacts with the sample. The indicator electrodes detect the changing potential or conductivity of the solution and the amount of change that takes place is measured and related to the concentration of the reactant in the sample.
The following table lists some common coulometric (also known as constant-current coulometry) titrations.1 Since the titrant is generated electrolitically and reacted immediately, the method gets widespread applications. The generating electrolytic concentrations need to be only approximate, while unstable titrants are consumed as soon as they are formed. The technique is more accurate than methods where visual end points are required, such as in the case of indicators. The unstable titrants in the table below are marked with an asterisk. [Pg.300]

Controlled-current electrolysis in flowing solution has been extremely useful for analytical purposes. The prevalent techniques are constant-current coulometry and coulometric titrations, which are discussed in Chapter 25. [Pg.140]

The uses of constant-current coulometry for the determination of drugs in biological fluids are few, basically due to sensitivity restriction. Monforte and Purdy [46] have reported an assay for two allylic barbituric acid derivatives, sodium seconal and sodium sandoptal, with electrogenerated bromine as the titrant and biamperometry for endpoint detection. Quantitative bromination required an excess of bromine hence back titration with standard arsenite was performed. The assay required the formation of a protein-free filtrate of serum with tungstic acid, extraction into chloroform, and sample cleanup by back extraction, followed by coulometric titration with electrogenerated bromine. The protein precipitation step resulted in losses of compound due to coprecipitation. The recoveries of sodium seconal and sodium sandoptal carried through the serum assay were approximately 81 and 88%, respectively. Samples in the concentration range 7.5-50 pg/mL serum were analyzed by this procedure. [Pg.781]

Scale of Operation Coulometric methods of analysis can be used to analyze small absolute amounts of analyte. In controlled-current coulometry, for example, the moles of analyte consumed during an exhaustive electrolysis is given by equation 11.32. An electrolysis carried out with a constant current of 100 pA for 100 s, therefore, consumes only 1 X 10 mol of analyte if = 1. For an analyte with a molecular weight of 100 g/mol, 1 X 10 mol corresponds to only 10 pg. The concentration of analyte in the electrochemical cell, however, must be sufficient to allow an accurate determination of the end point. When using visual end points, coulometric titrations require solution concentrations greater than 10 M and, as with conventional titrations, are limited to major and minor analytes. A coulometric titration to a preset potentiometric end point is feasible even with solution concentrations of 10 M, making possible the analysis of trace analytes. [Pg.507]

Time, Cost, and Equipment Controlled-potential coulometry is a relatively time-consuming analysis, with a typical analysis requiring 30-60 min. Coulometric titrations, on the other hand, require only a few minutes and are easily adapted for automated analysis. Commercial instrumentation for both controlled-potential and controlled-current coulometry is available and is relatively inexpensive. Low-cost potentiostats and constant-current sources are available for less than 1000. [Pg.508]

Now returning to the coulometric analysis proper we can. say that any determination that can be carried out by voltammetry is also possible by coulometry whether it should be done by means of the controlled-potential or the titration (constant-current) method much depends on the electrochemical properties of the analyte itself and on additional circumstances both methods, because they are based on bulk electrolysis, require continuous stirring. [Pg.234]

Controlled-current coulometry is also called coulometric titration. An apparatus for controlled-current coulometry is shown in Fig. 5.35 for the case of determination of an acid. It consists of a constant current source, a timer, an end-point detector (pH meter), and a titration cell, which contains a generating electrode, a counter electrode in a diaphragm, and two electrodes for pH detection. The timer... [Pg.147]

Coulometry employs either a constant current or a controlled potential. Constant-current methods, like the preceding Br2/cyclohexene example, are called coulometric titrations. If we know the current and the time of reaction, we know how many coulombs have been delivered from Equation 17-2 q = / t. [Pg.357]

In coulometry, the moles of electrons needed for a chemical reaction are measured. In a coulometric (constant current) titration, the time needed for complete reaction measures the number of electrons consumed. Controlled-potential coulometry is more selective than... [Pg.372]

The Karl Fischer titration of water uses a buret to deliver reagent or coulometry to generate reagent. In bipotentiometric endpoint detection, the voltage needed to maintain a constant current between two Pt electrodes is measured. The voltage changes abruptly at the equivalence point, when one member of a redox couple is either created or destroyed. [Pg.373]

Coulometry comes in several flavors constant-potential or potentiostatic coulometry, constant-current or amperostatic coulometry, coulometric titrations, and electrogravimetry. [Pg.738]

There are essentially two different coulometric processes, namely potentio-static and galvanostatic coulometry. The former functions with constant, controlled electrode potential, whereas the galvanostatic method - also called coulometric titration - functions with constant current strength and uncontrolled potential. Fig. 13 shows the basic circuit diagram for potentiostatic coulometry. [Pg.76]

The accepted reference method for determining chloride in blood serum, plasma, urine, sweat, and other body fluids is the coulometric titration procedure. In this technique, silver ions are generated coulometri-cally. The silver ions then react with chloride ions to form insoluble silver chloride. The end point is usually detected by amperometry (see Section 25C-4) when a sudden increase in current occurs on the generation of a slight excess of Ag. In principle, the absolute amount of Ag needed to react quantitatively with CP can be obtained from application of Faraday s law. In practice, calibration is used. First, the time t, required to titrate a chloride standard solution with a known number of moles of chloride using a constant current / is measured. The same constant current is next used in the titration of the unknown solution, and the time is measured. The number of moles of chloride in the unknown is then obtained as follows ... [Pg.363]

Controlled-current coulometry uses a constant current. which passes through a cell until an indicator signals completion of the analytical reaction. The quantity of charge required to reach the end point is then calcuiated from the magnitude of the current and the time that the current passes. This method has enjoyed wider application than potentiostatic coulometry. It is frequently called a coulometric titration for reasons that we discuss in Section 24D. [Pg.886]

The coulometric titration or amperostatic coulometry method in which the current is maintained constant during the measurement by an amperostat until the signal which marks the end of the reaction with the analyte. This is the most simple form of coulometric measurement. [Pg.480]

Some typical important industrial applications of coulometry include the continuous monitoring of mercaptan concentration in the materials used in rubber manufacture. The sample continuously reacts with bromine, which is reduced to bromide. A third electrode measures the potential of B12 vs. Br and, based on the measurement, automatically regulates the coulometric generation of the bromine. Coulometry is used in commercial instruments for the continuous analysis and process control of the production of chlorinated hydrocarbons. The chlorinated hydrocarbons are passed through a hot furnace, which converts the organic chloride to HCl. The latter is dissolved in water and the Cl titrated with Ag" ". The Ag" " is generated by coulometry from a sUver electrode, Ag°. It is necessary for the sample flow rate to be constant at all times. Integration of the coulometric current needed to oxidize the silver to silver ion results in a measurement of the Cl concentration. [Pg.969]

Coulometry. Two methods of coulometry are used coulometry at controlled potential and coulometric titrations. The main advantage of the coulometric method is the elimination of the necessity of standardization as the Faraday constant is a standard. In analysis of complicated samples encountered in environmental analysis the coulometric titrations are more advantageous where 100% current efficiency can be more readily attained by suitable choice of the reagent-solvent system. Coulometric titrations are suitable for determining the amount of substance in the range 0.01 to 100 mg (and sometimes below 1 iJg). Under optimum conditions these titrations can be carried out with a precision and accuracy of 0.01%. Automatic coulometric analyzers for the determination of gaseous pollutants (SO2, O3, NO, etc.) have proven to be useful in environmental chemistry. [Pg.106]


See other pages where Constant-current coulometry titrations is mentioned: [Pg.113]    [Pg.534]    [Pg.340]    [Pg.262]    [Pg.773]    [Pg.777]    [Pg.262]    [Pg.28]    [Pg.259]    [Pg.401]    [Pg.28]    [Pg.771]    [Pg.288]    [Pg.860]    [Pg.238]    [Pg.288]    [Pg.53]    [Pg.106]    [Pg.342]    [Pg.814]    [Pg.396]    [Pg.1094]    [Pg.39]   


SEARCH



Constant current

Constant current coulometry

Coulometry

Coulometry titration

© 2024 chempedia.info