Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Condensation, test atmospheres

Of particular interest in the present chapter is the effect of test atmosphere on creep and creep damage mechanisms. While there are undoubtedly several factors that can promote creep cavitation and contribute to the observed changes in stress exponent and activation energy, the fact remains that the strain rates are substantially higher in air than in inert atmospheres, as shown in Fig. 8.12. This phenomenon is a direct consequence of the topotactic oxidation reaction of SiC whiskers exposed at the surface. As described by Porter and Chokshi,38 and subsequently by others,21,22 at high temperatures in air, a carbon-condensed oxidation displacement reaction occurs in which graphitic carbon and silica are formed at the whisker interface via... [Pg.288]

In addition to the condenser, a test facility must have adequate diffusion-pump capacity to handle leakage and outgassingof noncondensible gases. If liquid nitrogen is the condenser coolant, atmospheric gases and hydrogen or helium are essentially noncondensible gases. If liquid helium were used in all or part of the condenser, a reduction in required diffusion-pump capacity would result, but it is questionable whether such a helium system could be competitive with the diffusion pump on a cost basis. [Pg.14]

Painted Sheet Metal Corrosion Test Various Cyclic Corrosion Tests Condensation Water Test Atmospheres... [Pg.132]

DIN 50017 Atmospheres and there technical application Condensation water test atmospheres. [Pg.382]

Following the run, the lower housing was removed, and the specimens and test chamber were heated with an electric air heater to prevent condensation of atmospheric moisture. The test specimens were usually removed from the apparatus as quickly as possible to minimize the possibility of corrosion. [Pg.75]

To obtain an adequate amount of this isomer, double the amounts in the preparation in Sec. 11.5.3. Weigh out 5 g of the prepared compound, grind to a fine powder. Distil purest thionyl chloride and add 15 cm of the freshly distilled liquid to the powdered solid in a flask fitted with a reflux condenser. Test the evolved gases with acidified dichromate paper and a drop of cone, ammonia. When the evolution of SO2 and ammonia, as tested, appears to have stopped, reflux the contents under a nitrogen atmosphere for 1-2 hours. Remove the unreacted thionyl chloride by vacuum distillation and store the solid formed in a stoppered bottle in a vacuum desiccator. [Pg.156]

The oxime is freely soluble in water and in most organic liquids. Recrystallise the crude dry product from a minimum of 60-80 petrol or (less suitably) cyclohexane for this purpose first determine approximately, by means of a small-scale test-tube experiment, the minimum proportion of the hot solvent required to dissolve the oxime from about 0-5 g. of the crude material. Then place the bulk of the crude product in a small (100 ml.) round-bottomed or conical flask fitted with a reflux water-condenser, add the required amount of the solvent and boil the mixture on a water-bath. Then turn out the gas, and quickly filter the hot mixture through a fluted filter-paper into a conical flask the sodium chloride remains on the filter, whilst the filtrate on cooling in ice-water deposits the acetoxime as colourless crystals. These, when filtered anddried (either by pressing between drying-paper or by placing in an atmospheric desiccator) have m.p. 60 . Acetoxime sublimes rather readily when exposed to the air, and rapidly when warmed or when placed in a vacuum. Hence the necessity for an atmospheric desiccator for drying purposes. [Pg.94]

Place 35 ml. of a M solution of aluminium tsopropoxide or 7 g. of solid aluminium tsopropoxide, 450 ml. of dry isopropyl alcohol and 21 g. of purified benzaldehyde (Section IV,115) in a 1 litre round-bottomed flask. Fit a short reflux condenser (no water in the cooling jacket) or better a Hahn condenser (2) (containing a 1 cm. layer of ethyl alcohol in the iimer tube) to the flask and arrange for slow distillation from a water bath at the rate of 3-6 drops per minute. Continue the heating until a negative test for acetone is obtained after 5 minutes of total reflux (6-9 hours) if the volume of the mixture falls below 200 ml. during the reduction, add more isopropyl alcohol. Remove the reflux or Hahn condenser and distil off (Fig. II, 13, 3) most of the isopropyl alcohol under atmospheric pressure from a suitable oil bath. Hydrolyse the... [Pg.884]

Fluorescent ultraviolet lamps within an apparatus that allows condensation cycles rather than the water spray typical of xenon arc tests have been developed for plastics testing (279). The spectral cutoff wavelength of the lamps used in the apparatus determines the severity of the test. Ultraviolet B (UVB) 313 lamps allow a significant irradiance component below 290 nm, which is normally filtered out by the earth s atmosphere. Ultraviolet A (UVA)... [Pg.155]

Effect of mass The rate of rusting of steel in the atmosphere is affected to some extent by the mass of the part concerned, because this determines the speed at which the surface temperature adjusts itself to fluctuations in the ambient temperature, the amount of condensation during humid periods, and the time during which dew or rain remains in contact with the steel. For example, in a test over 12 months at the National Chemical Laboratory under sheltered conditions outdoors, thick steel plates rusted more than thin ones as is shown below. [Pg.496]

Assemble the ice bath. The vapors will condense and collect in the second test tube. Place enough ice in a 250-mL beaker to fill it half full. Place the second test tube in the ice bath. Insert the loose end of the tubing into the test tube. Make sure that this end of the tubing is open to the atmosphere. CAUTION Do not use a rubber stopper. [Pg.102]

The zinc hydroxide carbonate sample (Merck) was weighed on the thermobalance and linearly heated or cooled in the water vapor furnace. The H20-C02 atmosphere was generated by a flow of C02 through the water vaporizer into the sample chamber. The condensed water flows back into the flask, the C02 leaves through the gas outlet. An additional flow of C02 through the balance prevents any water condensation in the balance chamber or on the sample holder. The tests could only be started after both C02 gas flows were adjusted to a constant rate and the vaporizer showed a constant return flow of condensed water. Flowmeters were used to adjust and control the gas flow rates. [Pg.134]

The activity of calcined HTs was determined in self-condensation reaction of acetone (J.T. Baker) by using a fixed bed catalytic reactor with an on-line GC. Prior to the catalytic test, catalysts were pretreated in-situ under nitrogen atmosphere at 450°C for 5h. Acetone was supplied to the reactor by bubbling nitrogen gas through the acetone container at 0 °C. The reaction temperature was established at 200 C. The products were analyzed by means of GC (Varian CP-3800) using a WCOT Fused silica column, equipped with a FID detector. [Pg.59]

Further, steam and gas may escape into the atmosphere from wells undergoing tests, and from blow-out drillholes. Some gas may also escape in solution in the surplus condensate flowing from the cooling towers. [Pg.302]

It is also important to correct the raw vapour pressure data for any pad gas which was present in the test cell. This can be done by subtracting the partial pressure of any non-condensible pad gas which was present in the test cell, to obtain the vapour pressure (see A2.7.1). Because pressure transducers may not be very accurate at the bottom end of their range, it is advisable to vent the test cell to atmosphere, once it is filledand before sealing it and heating to the initial runaway temperature, so that a reliable initial pad gas pressure is known. (This may not always be compatible with the desire to simulate the runaway scenario within the test.) An alternative is to evacuate both test cell and containment vessel before the reactants are added so that there is no pad gas and no correction is needed,... [Pg.137]

The flash gas from the top of F-l passed through a wet-test meter to a sample tap and was vented to the atmosphere. Condensed liquid product was collected through a valve at the bottom of F-7, weighed, and analyzed. [Pg.51]


See other pages where Condensation, test atmospheres is mentioned: [Pg.2426]    [Pg.15]    [Pg.142]    [Pg.2181]    [Pg.2691]    [Pg.2668]    [Pg.2430]    [Pg.78]    [Pg.513]    [Pg.1324]    [Pg.127]    [Pg.62]    [Pg.189]    [Pg.231]    [Pg.1027]    [Pg.1069]    [Pg.1085]    [Pg.513]    [Pg.884]    [Pg.65]    [Pg.341]    [Pg.167]    [Pg.338]    [Pg.297]    [Pg.305]    [Pg.66]    [Pg.355]    [Pg.339]    [Pg.260]    [Pg.344]    [Pg.513]    [Pg.884]    [Pg.9]   
See also in sourсe #XX -- [ Pg.15 , Pg.16 ]




SEARCH



Condensation atmosphere

Condensers atmospheric

© 2024 chempedia.info