Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Column parameters, reduced

Column design involves the application of a number of specific equations (most of which have been previously derived and/or discussed) to determine the column parameters and operating conditions that will provide the analytical specifications necessary to achieve a specific separation. The characteristics of the separation will be defined by the reduced chromatogram of the particular sample of interest. First, it is necessary to calculate the efficiency required to separate the critical pair of the reduced chromatogram of the sample. This requires a knowledge of the capacity ratio of the first eluted peak of the critical pair and their separation ratio. Employing the Purnell equation (chapter 6, equation (16)). [Pg.367]

Performance Characterization of Liquid Chromatography Columns Using Reduced Parameters... [Pg.43]

Figure 4.17 Profiles of (a) Fe(II), (b) Fe(III) and (c) pH in columns of reduced soil exposed to O2 at one end for different times. Points are experimentally measured lines are predicted using the model described in the text with independently estimated parameter values (Kirk and Solivas, 1994). Reproduced by permission of Blackwell Publishing... Figure 4.17 Profiles of (a) Fe(II), (b) Fe(III) and (c) pH in columns of reduced soil exposed to O2 at one end for different times. Points are experimentally measured lines are predicted using the model described in the text with independently estimated parameter values (Kirk and Solivas, 1994). Reproduced by permission of Blackwell Publishing...
Thus the evaluation of the basic process parameters reduces to the determination of the amount of the element contained in the initial and resultant components of the raw material and the end product. Such values can be established in different ways, some of which were discussed by Pankov and Khripin [175], To obtain a value proportional to the amount of carbon in the key component, the zones of compounds separated in a chromatographic column are converted into carbon dioxide or methane. [Pg.242]

The traditional GC separation with a reduced run time, i.e., fast GC, is now the cutting edge and can be achieved by the use of shorter capillary columns, columns with reduced internal diameter, thinner stationary-phase films, H2 as the carrier gas, higher carrier gas velocities, faster oven temperature programming rates, and combinations of these parameters. However, one must be careful to understand the impact of these changes on the chromatographic resolution. Despite the unavoidable decrease in efficiency, the overall analytical results are obtained with 95% reduction in analytical time, typically less than 2... [Pg.844]

In Eq. (3), it is assumed that every possible compoimd may be present in the sample. When this is the case, each of the column parameters (H, S, etc.) can be important in affecting separation selectivity for that sample. If one or more of these column parameters has little or no effect on the separation, however, it is useful to reduce its contribution to the final value of F, because the resulting lower values of F make it more likely that a suitable column (i.e., one with F < 3) can be found. Specifically, the presence or absence of certain compound types leads to a re-weighting of the last two terms of Eq. (1) ... [Pg.329]

Method development remains the most challenging aspect of chiral chromatographic analysis, and the need for rapid method development is particularly acute in the pharmaceutical industry. To complicate matters, even structurally similar compounds may not be resolved under the same chromatographic conditions, or even on the same CSP. Rapid column equilibration in SFC speeds the column screening process, and automated systems accommodating multiple CSPs and modifiers now permit unattended method optimization in SFC [36]. Because more compounds are likely to be resolved with a single set of parameters in SFC than in LC, the analyst stands a greater chance of success on the first try in SFC [37]. The increased resolution obtained in SFC may also reduce the number of columns that must be evaluated to achieve the desired separation. [Pg.305]

To reduce or eliminate polymer solute/glass packing interactions the following parameters were optimized a) pH, ionic strength and concentrations of additives such as nonionic surfactants, b) selection of pore sizes in a column combination. [Pg.267]

The parameter p (= 7(5 ) in gas-liquid sy.stems plays the same role as V/Aex in catalytic reactions. This parameter amounts to 10-40 for a gas and liquid in film contact, and increases to lO -lO" for gas bubbles dispersed in a liquid. If the Hatta number (see section 5.4.3) is low (below I) this indicates a slow reaction, and high values of p (e.g. bubble columns) should be chosen. For instantaneous reactions Ha > 100, enhancement factor E = 10-50) a low p should be selected with a high degree of gas-phase turbulence. The sulphonation of aromatics with gaseous SO3 is an instantaneous reaction and is controlled by gas-phase mass transfer. In commercial thin-film sulphonators, the liquid reactant flows down as a thin film (low p) in contact with a highly turbulent gas stream (high ka). A thin-film reactor was chosen instead of a liquid droplet system due to the desire to remove heat generated in the liquid phase as a result of the exothermic reaction. Similar considerations are valid for liquid-liquid systems. Sometimes, practical considerations prevail over the decisions dictated from a transport-reaction analysis. Corrosive liquids should always be in the dispersed phase to reduce contact with the reactor walls. Hazardous liquids are usually dispensed to reduce their hold-up, i.e. their inventory inside the reactor. [Pg.388]

Increasing the speed of analysis has always been an important goal for GC separations. All other parameters being equal, the time of GC separations can be decreased in a number of ways (1) shorten the column (2) increase the carrier gas flow rate (3) reduce the column film thickness (4) reduce the carrier gas viscosity (5) increase the column diameter and/or (6) heat the column more quickly. The trade-off for increased speed, however, is reduced sample capacity, higher detection limits, and/or worse separation efficiency. [Pg.763]

TYPICAL REDUCED PARAMETERS FOR DIFFERENT COLUMN TYPES... [Pg.44]

Column Type Minimum Reduced Plate Height Minimum Reduced Velocity flow Resistance Parameter Separation Impedance... [Pg.44]

Having chosen the test mixture and mobile diase composition, the chromatogram is run, usually at a fairly fast chart speed to reduce errors associated with the measurement of peak widths, etc.. Figure 4.10. The parameters calculated from the chromatogram are the retention volume and capacity factor of each component, the plate count for the unretained peak and at least one of the retained peaks, the peak asymmetry factor for each component, and the separation factor for at least one pair of solutes. The pressure drop for the column at the optimum test flow rate should also be noted. This data is then used to determine two types of performance criteria. These are kinetic parameters, which indicate how well the column is physically packed, and thermodynamic parameters, which indicate whether the column packing material meets the manufacturer s specifications. Examples of such thermodynamic parameters are whether the percentage oi bonded... [Pg.184]

Reduced parameters (section 1.7.10) can -be used to compare the potential of open tubular and packed columns to deliver a certain separation potential in SFC [8,43,53-56]. The Golay equation, equation 6.1, can be rewritten as... [Pg.822]

Upon substitution of the reduced parameters given above the separation time for a packed column and an open tubular column would be Identical if d 1.73 dp given the current limitations of open tubular column technology the column diameter cannot be reduced to the point %diere these columns can compete with packed columns for fast separations. This is illustrated by the practical txanple in Figure 6.3 (57). Ihe separation speed cannot be Increased for an open tubular column by increasing the reduced velocity since the reduced plate height is increased... [Pg.823]

The maximum number of theoretical plates that can be achieved for a particular column depends on the pressure drop per theoretical plate and the maximum pressure drop allowed over the column, (ising the same reduced parameters given earlier, the maximum number of theoretical plates obtained for an open tubular... [Pg.823]


See other pages where Column parameters, reduced is mentioned: [Pg.394]    [Pg.562]    [Pg.563]    [Pg.565]    [Pg.292]    [Pg.196]    [Pg.39]    [Pg.395]    [Pg.240]    [Pg.399]    [Pg.1313]    [Pg.121]    [Pg.227]    [Pg.305]    [Pg.216]    [Pg.179]    [Pg.197]    [Pg.177]    [Pg.432]    [Pg.36]    [Pg.44]    [Pg.259]    [Pg.311]    [Pg.699]    [Pg.733]    [Pg.847]    [Pg.378]    [Pg.191]    [Pg.194]    [Pg.235]    [Pg.239]   
See also in sourсe #XX -- [ Pg.31 ]




SEARCH



Reduced parameters

© 2024 chempedia.info