Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Column-packing equipment

Please also make sure of your standard laboratory safety regulations and operating safety instructions for hydraulic equipment are followed when using the column packing equipment. [Pg.435]

In order to obtain a dense, uniform packing of the small-diameter particles, it is recommended that the packing is slurried in a liquid of similar density to the packing. This slurry is then rapidly pumped into the column to form a uniform bed. If specialised column packing equipment is not available, it is recommended that pre-packed commercial columns are purchased. [Pg.220]

Samples and calibration standards are prepared for analysis using a 10-mL syringe. Add 10.00 mL of each sample and standard to separate 14-mL screw-cap vials containing 2.00 mL of pentane. Shake vigorously for 1 min to effect the separation. Wait 60 s for the phases to separate. Inject 3.0-pL aliquots of the pentane layer into a GC equipped with a 2-mm internal diameter, 2-m long glass column packed with a stationary phase of 10% squalane on a packing material of 80/100 mesh Chromosorb WAW. Operate the column at 67 °C and a flow rate of 25 mL/min. [Pg.576]

The principal uses of PCTFE plastics remain in the areas of aeronautical and space, electrical/electronics, cryogenic, chemical, and medical instmmentation industries. AppHcations include chemically resistant electrical insulation and components cryogenic seals, gaskets, valve seats (56,57) and liners instmment parts for medical and chemical equipment (58), and medical packaging fiber optic appHcations (see Fiber optics) seals for the petrochemical /oil industry and electrodes, sample containers, and column packing in analytical chemistry and equipment (59). [Pg.394]

The commercial production equipment consists of a furnace, heat-exchanger tubes, a fractionating column packed with Rachig rings, a KCl feed, a waste removal system, and a vapor condensing system (Fig. 1). [Pg.516]

A 60-cm. x 1-cm. externally heated column packed with 4-mm. glass helices and equipped with a total-reflux variable take-off head was used. [Pg.88]

Fig.2 shows the experimental results of Co(NH3)6 reduction, which is done in a stirred reactor at 80 °C and pH 4.1. It can be seen that activated carbon can promote the [Co(NH3)6] reduction significantly. The [Co(NH3)6] " conversion reaches 81.2% with6.7g/l activated carbon in the aqueous solution while only 8.18% of [Co(NH3)6] is reduced when there is no activated carbon added. Thus, a regeneration column packed with the activated carbon should be equipped with the absorber. [Pg.231]

No expensive equipment is required for OCC however, the separation efhciency depends on the analyst s experience since a new column has to be packed for each analysis. In addition, depending on the packing type (powder or slurry), stationary phase, and purpose of the separation, the separation can take from 30 min to 4 hr. The AOAC official method for the determination of carotenoids still uses OCC." Separation of carotenoids from many foods was developed on a column packed with a mixture of MgO and HyfloSupercel (or celite or diatomaceous earth) at 1 1... [Pg.454]

Few laboratories have developed their own facilities for packing conventional and small bore columns, preferring to rely on commercial manufacturers for their needs. This is understandable since packed columns containing any common phase can be obtained at an acceptable cost and with a guarantee of acceptable performance and stability. Column packing generally requires access to equipment not readily available in all laboratories and... [Pg.176]

Chromatographic system. (Follow the method described in the general procedure <621 >.) The gas chromatograph is equipped with a flame ionization detector and a 1.2 m x 2 mm column packed with 3% phase G32 on support S1A. The injection port, detector, and column temperatures are maintained at about 250, 300, and 250 °C, respectively, and helium is used as the carrier gas, flowing at rate of about 50 mL/min. The relative retention times for cholestane and miconazole nitrate are about 0.44 and 1, respectively. Chromatograph the Standard preparation, and record the peak responses as directed for procedure The resolution, R, between cholestane and miconazole nitrate is not less than 2 and the relative standard deviation of replicate injections is not more than 3%. [Pg.35]

Chromatographic system. The gas chromatograph is equipped with a flamioniza-tion detector and a 2 mm x 1.8 m glass column packed with 10% phase G34 on 80- to 100-mesh support SI A. The column temperature is maintained at about 150 °C, and the injection port and the detector block temperatures are maintained at about 250 °C. Dry helium is used as the carrier gas at a flow rate of about 40 mL/min. Chromatograph the Standard preparation, measure the peak responses, and calculate the ratio, Rs, as directed for procedure the relative retention times are about 0.5 for valproic acid and 1.0 for biphenyl the resolution, R, between valproic acid and biphenyl is not less than 3.0 the relative standard deviation for replicate injections is not more than 2.0%. [Pg.227]

Modem technologies provide many techniques for expanding the throughput of an analytical laboratory. The task that needs to be accomplished and the possible drawbacks should be carefully considered. Optimized LC equipment can utilize columns packed with much smaller stationary phase particles to achieve significant reductions in gradient time while still achieving the same or even better peak capacities than conventional methods. [Pg.117]

The submitters concentrated the dry ether solution to a volume of 80-100 ml. by distillation through a 1.0-cm. x 40-cm. column packed with glass helices and equipped with a total-... [Pg.28]

Note A mixture of two organic liquids is used for this study. The specific liquids will be selected by your instructor. They should be available in a vial equipped with a rubber septum. Your instructor has also selected the initial column packing and length. The two liquids should be in a ratio of approximately 1 1 by volume. [Pg.360]

Colon and Maloney demonstrated another packing method that also avoids pumping the slurry through the column [51]. They used centripetal force to drive beads, which have a higher density than the liquid contained in solvent slurry, through the capillary. Their packing equipment enables a rotation speed of up to 3000 rpm at which the packing time is only 5-15 min. [Pg.16]


See other pages where Column-packing equipment is mentioned: [Pg.96]    [Pg.58]    [Pg.59]    [Pg.59]    [Pg.228]    [Pg.81]    [Pg.865]    [Pg.53]    [Pg.4]    [Pg.75]    [Pg.139]    [Pg.159]    [Pg.550]    [Pg.227]    [Pg.1289]    [Pg.253]    [Pg.628]    [Pg.239]    [Pg.704]    [Pg.178]    [Pg.90]    [Pg.136]    [Pg.138]    [Pg.116]    [Pg.252]    [Pg.259]    [Pg.340]    [Pg.365]    [Pg.98]    [Pg.573]    [Pg.263]    [Pg.289]    [Pg.63]   
See also in sourсe #XX -- [ Pg.21 , Pg.46 ]




SEARCH



Equipment packed columns

Equipment packed columns

Equipment packing

High-performance liquid chromatography column-packing equipment

Packed columns

Packed columns, packing

© 2024 chempedia.info