Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Chart recorders 10% resolution

The mixture is identical in each example. The peaks are shown separated by 2, 3, 4, 5 and 6 (a) and it is clear that a separation of 6a would appear to be ideal for accurate quantitative results. Such a resolution, however, will often require very high efficiencies which will be accompanied by very long analysis times. Furthermore, a separation of 6o is not necessary for accurate quantitative analysis. Even with manual measurements made directly on the chromatogram from a strip chart recorder, accurate quantitative results can be obtained with a separation of only 4a. That is to say that duplicate measurements of peak area or peak height should not differ by more than 2%. (A separation of 4a means that the distance between the maxima of the two peaks is equal to twice the peak widths). If the chromatographic data is acquired and processed by a computer, then with modem software, a separation of 4a is quite adequate. [Pg.109]

The reaction manifold describing the automated determination of ammonia is shown in Fig. 6.1. Two alternative modes of sampling are shown discrete and continuous. Discrete 5 ml samples contained in ashed (450 °C) glass vials are sampled from an autosampler (Hook and Tucker model A40-11 1.5 min sam-ple/wash). For high-resolution work in the estuary, the continuous sampling mode is preferred. The indophenol blue complex was measured at 630 nm with a colorimeter and the absorbance recorded on a chart recorder. [Pg.329]

Detect the PTC-amino acids as they elute with a spectrophotometer set at 254 nm. When the signal output (10 mV maximum) is relayed to a chart recorder with a speed of 0.5 cm/min, the PTC-amino acid peaks will show good height and good resolution. [Pg.119]

Figure 8.3 Chart recorder pen response and 10% valley measurement of resolution. Figure 8.3 Chart recorder pen response and 10% valley measurement of resolution.
The earliest work in high-resolution mass spectrometry produced spectra similar in appearance to that in Figure 3.3a. Chart records were obtained by slowly varying the magnetic field of a Nier-Johnson mass spectrometer. The time (or distance along the chart) between the centers of unknown and reference peaks and the known reference masses was used to determine the exact masses of the unknown ions. Current instruments, with their computerized data acquisition systems, automatically identify the reference masses, generate the internal calibration table, and calculate the accurate masses. [Pg.126]

Frequently, it is desirable to use the window mode in conjunction with a pulse-height-selector scanner in order to measure the spectrum of Fig. 4.16(a) on a strip-chart recorder. If the resolution of Fig. 4.16(a) is to be faithfully reproduced in Fig. 4.16(b), the window width must be small compared to the peak width in Fig. 4.16(a). That is, the window width must be small compared to the pulse height resolution inherent in the amplifier pulse height spectrum. A good practical choice is to make the window width less than one-fourth of the width of the narrowest peak in the amplifier pulse height spectrum—the width being taken as the width of the peak at half the peak amplitude above background. [Pg.114]

Several method performance indicators are tracked, monitored, and recorded, including the date of analysis, identification of equipment, identification of the analyst, number and type of samples analyzed, the system precision, the critical resolution or tailing factor, the recovery at the reporting threshold level, the recovery of a second reference weighing, the recovery for the control references (repeated reference injections for evaluation of system drift), the separation quality, blank issues, out of spec issues, carry over issues, and other nonconformances. The quantitative indicators are additionally visualized by plotting on control charts (Figure 23). [Pg.93]

You will note some structure in the Stokes band near 460 which is due to chlorine isotopic frequency shifts. (See Exp. 37 for a discussion of isotope effects in diatomic molecules.) Rescan this region at higher resolution (1 cm or less) with an expansion of the chart display and measure the frequencies of each of the components. Record the ambient temperature near the Raman cell. [Pg.405]

The procedures described next were developed for the deconvolution of electronic absorption spectra (UV-visible spectra) but are equally applicable to the deconvolution of infrared, Raman or NMR spectra. UV-visible spectra differ from vibrational spectra in that the number of bands is much smaller and the bandwidths are much wider. Band shape may also be different. UV-visible spectra are also usually recorded under conditions of high resolution and high signal-to-noise ratio. Spectra from older instruments usually require manual digitization from a spectrum on chart paper, at e.g., 10 nm intervals. With the widespread use of computer-controlled instruments, it is a simple matter to obtain a file of spectral data at, e.g., 1 nm intervals. In fact, it may be necessary to reduce the size of the data set to speed up calculations. [Pg.344]

To be a successful monitor, the sponsor representative should know how to interpret hospital/ clinic records/charts, laboratory tests and interpretations, query resolution procedures, protocol and CRF data requirements, medical nomenclature, SAE procedures and health authority requirements. In addition, a monitor needs to have excellent interpersonal communication and problemsolving skills. [Pg.34]

Figure 16.1 Fragmentation spectrum and mass spectrum presented in graphical or tabular form. (a) Fragmentation spectrum of methanol (b) Unconventional representation of the same spectrum in the form of a circular diagram (hollow pie chart) statistically, for 321 ions formed, there are 100 of mass 31 u (Da), 72 of mass 29, etc. The various ions constitute many different populations (c) Section of a high-resolution recording of a compound M presenting two ions with very close m/z ratios (one due to loss of CO and the other due to loss of C2H4). Figure 16.1 Fragmentation spectrum and mass spectrum presented in graphical or tabular form. (a) Fragmentation spectrum of methanol (b) Unconventional representation of the same spectrum in the form of a circular diagram (hollow pie chart) statistically, for 321 ions formed, there are 100 of mass 31 u (Da), 72 of mass 29, etc. The various ions constitute many different populations (c) Section of a high-resolution recording of a compound M presenting two ions with very close m/z ratios (one due to loss of CO and the other due to loss of C2H4).
A sampler/wash ratio of 1 2 was originally proposed by Eivazi et al. (1982). To be able to use a computer program (Sigmacan) to read the charts, the sampler/wash ratio of 1 2 was changed to 1 3 timing (Sims and Crutchfield, 1992). Although that reduces the daily output by 50%, a better baseline resolution between peaks on the recorder is achieved, which allows more accurate interpolation of the data. Analysis of 100 digested samples per day is possible even at the slower rate. [Pg.98]

These improvements and the matching of the time resolution throughout the recording channel enabled us to determine the integrated intensify down to the level of 2-4 electrons and to record the diffuse background directly on a potentiometer chart. This made it possible to reveal fine details of the structure. [Pg.35]

Apart from keeping a record of the state of the equipment, the routine calibration parameters should be monitored by means of control charts. One might monitor, for example, the resolution at two energies, the energy calibration factors, and the full-energy peak efficiency at two energies. A simple example is shown in Figure 15.2. [Pg.298]


See other pages where Chart recorders 10% resolution is mentioned: [Pg.207]    [Pg.252]    [Pg.22]    [Pg.252]    [Pg.85]    [Pg.568]    [Pg.248]    [Pg.201]    [Pg.214]    [Pg.347]    [Pg.99]    [Pg.129]    [Pg.353]    [Pg.605]    [Pg.447]    [Pg.54]    [Pg.93]    [Pg.365]    [Pg.536]    [Pg.51]    [Pg.14]    [Pg.10]    [Pg.234]    [Pg.385]    [Pg.96]    [Pg.506]    [Pg.298]    [Pg.168]    [Pg.654]    [Pg.3]    [Pg.1911]    [Pg.1604]    [Pg.121]    [Pg.114]   
See also in sourсe #XX -- [ Pg.402 ]




SEARCH



Chart recorder

© 2024 chempedia.info