Nier-Johnson mass spectrometer

Schematic diagram of an electric and a magnetic sector as they are arranged in a forward geometry double-focusing mass spectrometer (Nier-Johnson geometry). (Reprinted from Ref. 7 with permission of John V ffley Sons)... Schematic diagram of an electric and a magnetic sector as they are arranged in a forward geometry double-focusing mass spectrometer (Nier-Johnson geometry). (Reprinted from Ref. 7 with permission of John V ffley Sons)...
Johnson and Nier published an ion optic with a 90° electric and 60° magnetic sector, which, because of the outstanding focussing properties, was to become the basis for many high-resolution, organic mass spectrometers (Nier/Johnson analyser). [Pg.5]

Fig. 1. Schematic diagram of a double-focusing Nier-Johnson magnetic sector mass spectrometer where ( " ) represents paths of ions having slightly... Fig. 1. Schematic diagram of a double-focusing Nier-Johnson magnetic sector mass spectrometer where ( " ) represents paths of ions having slightly...
The most recent metastable defocusing technique, which is referred to as mass-analyzed ion kinetic spectroscopy (MIKES) by Beynon (77, 270, 277) and as direct analysis of daughter ions (DADI) by Maurer (278) requires the interchange of the source and collector positions in a double-focusing mass spectrometer of Nier-Johnson geometry. With... [Pg.271]

Figure 3.5 Double-focusing sector field mass spectrometer with Nier-johnson geometry. (77. Kienitz (ed.), Massenspektrometrie (1968), Verlag Chemie, Weinheim. Reproduced by permission ofWiley-VCH)... Figure 3.5 Double-focusing sector field mass spectrometer with Nier-johnson geometry. (77. Kienitz (ed.), Massenspektrometrie (1968), Verlag Chemie, Weinheim. Reproduced by permission ofWiley-VCH)...
Figure 5.29 Instrumental setup of a radiofrequency (rf) glow discharge mass spectrometer by combining an ion source developed in house with a commercial double-focusing sector field mass spectrometer with reverse Nier-Johnson geometry (Element, Thermo Fisher Scientific, Bremen, Germany), (f S. Becker et al., IntJ. Mass Spectrom., 164, 81 (1997). Produced by permission of Elsevier.)... Figure 5.29 Instrumental setup of a radiofrequency (rf) glow discharge mass spectrometer by combining an ion source developed in house with a commercial double-focusing sector field mass spectrometer with reverse Nier-Johnson geometry (Element, Thermo Fisher Scientific, Bremen, Germany), (f S. Becker et al., IntJ. Mass Spectrom., 164, 81 (1997). Produced by permission of Elsevier.)...
Figure 16.5—Magnetic analyser mass spectrometer, a) Nier Johnson system, b) directional focusing properties of the magnetic field, c) principle of a double focusing mass spectrometer d) Mattauch-Herzog system. Figure 16.5—Magnetic analyser mass spectrometer, a) Nier Johnson system, b) directional focusing properties of the magnetic field, c) principle of a double focusing mass spectrometer d) Mattauch-Herzog system.
A double-focusing mass spectrometer with a Nier-Johnson configuration (with = = 90°)... [Pg.85]


See other pages where Nier-Johnson mass spectrometer is mentioned: [Pg.158]    [Pg.158]    [Pg.158]    [Pg.158]    [Pg.239]    [Pg.540]    [Pg.543]    [Pg.48]    [Pg.42]    [Pg.45]    [Pg.351]    [Pg.158]    [Pg.137]    [Pg.231]    [Pg.272]    [Pg.20]    [Pg.84]    [Pg.84]    [Pg.85]    [Pg.131]    [Pg.132]    [Pg.135]    [Pg.138]    [Pg.157]    [Pg.161]    [Pg.166]    [Pg.269]    [Pg.275]    [Pg.476]    [Pg.53]    [Pg.166]    [Pg.440]    [Pg.20]    [Pg.84]    [Pg.84]    [Pg.84]    [Pg.85]    [Pg.131]    [Pg.132]    [Pg.135]    [Pg.137]    [Pg.138]    [Pg.157]   
See also in sourсe #XX -- [ Pg.452 ]