Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Capillary current

The shape of the mass-analyzed gas phase /tot curve is seen to be very similar to that of the capillary current I (Figure 3). An approximate proportionality between the two currents when the concentration is increased is generally observed in this concentration range.45,51 For an analyte concentration above 10 5 M, the intensity of the impurity ions B+ is observed to decrease. This decrease is a consequence of the weak dependence of the capillary current I on the total electrolyte concentration. [Pg.272]

Besides the three modes of ionic migration discussed above for an electrolyte filled open silica capillary, current in a packed column can also flow via conduction through the stationary phase. However, when ionic conduction through the bulk electrolyte solution is still the primary form of ionic migration, conductivity of a packed column, apacked, can be expressed along similar lines as Eq. 1.1 in the following manner [18]... [Pg.19]

As a result, their intensity can be adjusted by modifying either the applied electrospray voltage or the position of the CE capillary outlet. Prior to starting a sequence, analyses have to be performed to ensure that these electric currents remain stable, otherwise, the capillary position has to be slightly adjusted to achieve good quantitative results, furthermore, the capillary current monitoring can be a diagnostic tool of the chiral selector s entrance into the nebulisation chamber. [Pg.277]

ESI. Flow for negative (-) ESI was the same, with a needle voltage of -2 kV and a heated capillary current of 4 amps. A linear octopole ion trap was used to accumulate ions, which were then transferred to a Penning ion trap via radio frequency (RF)-only multipoles. Mass-to-charge ratio (m/z) was determined by a quad-rupolar electric trapping potential approximation and used to generate the m/z spectrum [7]. [Pg.709]

The narrow bore of the capillary column and the relative thickness of the capillary s walls are important. When an electric field is applied to a capillary containing a conductive medium, such as a buffer solution, current flows through the capillary. This current leads to Joule heating, the extent of which is proportional to the capillary s radius and the magnitude of the electric field. Joule heating is a problem because it changes the buffer solution s viscosity, with the solution at the center of the... [Pg.601]

Ohm s law the statement that the current moving through a circuit is proportional to the applied potential and inversely proportional to the circuit s resistance (E = iR). (p. 463) on-column injection the direct injection of thermally unstable samples onto a capillary column, (p. 568) one-taUed significance test significance test in which the null hypothesis is rejected for values at only one end of the normal distribution, (p. 84)... [Pg.776]

A number of viscometers have been developed for securing viscosity data at temperatures as low as 0 °C (58,59). The most popular instmments in current use are the cone plate (ASTM D3205), parallel plate, and capillary instmments (ASTM D2171 and ASTM D2170). The cone plate can be used for the deterniination of viscosities in the range of 10 to over 10 Pa-s (10 P) at temperatures of 0—70°C and at shear rates from 10 to 10 5 . Capillary viscometers are commonly used for the deterniination of viscosities at 60 —135°C. [Pg.369]

One of the important problems in the diagnosis of different disease in their early stages is the determination of bio-active substances in biological fluids. We are currently interested in applying capillary electrophoresis (CE) as technique for the rapid and highly efficient separation of corticosteroids in semm and urine. Steroids can analyze by MEKC. [Pg.250]

The potential dependence of the velocity of an electrochemical phase boundary reaction is represented by a current-potential curve I(U). It is convenient to relate such curves to the geometric electrode surface area S, i.e., to present them as current-density-potential curves J(U). The determination of such curves is represented schematically in Fig. 2-3. A current is conducted to the counterelectrode Ej in the electrolyte by means of an external circuit (voltage source Uq, ammeter, resistances R and R") and via the electrode E, to be measured, back to the external circuit. In the diagram, the current indicated (0) is positive. The potential of E, is measured with a high-resistance voltmeter as the voltage difference of electrodes El and E2. To accomplish this, the reference electrode, E2, must be equipped with a Haber-Luggin capillary whose probe end must be brought as close as possible to... [Pg.40]

Since usually the reference electrode is not equipped with a capillary probe (see Fig. 2-3), there is an error in the potential measurement given by Eq. (2-34) in this connection see the data in Section 3.3.1 on IR-free potential measurement. The switching method described there can also be applied in a modified form to potential-controlled protection current devices. Interrupter potentiostats are used that periodically switch off the protection current for short intervals [5]. The switch-off phase is for a few tens of microseconds and the switch-on phase lasts several hundred microseconds. [Pg.448]

There are a number of tissue adhesives currently approved as adjuncts to hemostasis. Each agent has specific properties which enhance intraoperative hemostasis in particular situations. All forms of surgical bleeding are not alike. Bleeding can be fast or slow, diffuse or punctate, arterial or venous, capillary or larger vessel. Thus one agent may be more efficacious than another in a specific clinical situation. [Pg.1112]

The major advantage of the capillary hydrodynamic chromatography is that the mobile phase does not need to have similar solubility parameter as the sample and packing material. (In SEC, nonsize exclusion effects may be observed if the solubility parameter of the sample, packing material, or mobile phase is considerably different.) Therefore, the hydrodynamic size of polymers can be studied in a 0 solvent and even in a solvent that is not compatible with any currently available SEC packing material (9). Figure 22.4 is an example of polystyrene separation in both THF and diethyl malonate. Diethyl malonate is the 0 solvent of polystyrene at 31-36 C. [Pg.600]

SFC has been performed with either open capillary columns similar to those used in GC or packed columns transferred from LC, and the instrumentation requirements differ for these two approaches [12]. This chapter will focus on the use of packed column technology because of its dominance in the area of pharmaceutical compound separations. Current commercial instrumentation for packed column SFC utilizes many of the same components as traditional LC instruments, including pumps, injection valves, and detectors. In fact, most modem packed column SFC instm-ments can also be used to perform LC separations, and many of the same stationary phases can be used in both LC and SFC [9]. [Pg.302]

Measurements of the corrosion potential of a single metal corroding uniformly do not involve an IR drop, but similar considerations do not apply when the metal is polarised by an external e.m.f., and under these circumstances the IR drop must be minimised by using a Luggin capillary placed close to the surface of the electrode (see Fig. 1.22, Section 1.4). Even so, the IR drop is not completely eliminated by this method, and a further error is introduced by the capillary shielding the surface from the current flow... [Pg.1007]

Accurate control of potential, stability, frequency response and uniform current distribution required the following low resistance of the cell and reference electrode small stray capacitances small working electrode area small solution resistance between specimen and point at which potential is measured and a symmetrical electrode arrangement. Their design appears to have eliminated the need for the usual Luggin capillary probe. [Pg.1125]


See other pages where Capillary current is mentioned: [Pg.266]    [Pg.271]    [Pg.13]    [Pg.229]    [Pg.498]    [Pg.131]    [Pg.325]    [Pg.541]    [Pg.594]    [Pg.106]    [Pg.12]    [Pg.158]    [Pg.266]    [Pg.271]    [Pg.13]    [Pg.229]    [Pg.498]    [Pg.131]    [Pg.325]    [Pg.541]    [Pg.594]    [Pg.106]    [Pg.12]    [Pg.158]    [Pg.21]    [Pg.21]    [Pg.665]    [Pg.1926]    [Pg.766]    [Pg.603]    [Pg.84]    [Pg.60]    [Pg.344]    [Pg.298]    [Pg.108]    [Pg.25]    [Pg.51]    [Pg.179]    [Pg.182]    [Pg.337]    [Pg.346]    [Pg.501]    [Pg.430]    [Pg.265]    [Pg.48]    [Pg.1008]    [Pg.230]   


SEARCH



Capillaries streaming current

Capillary electrophoresis counter current

Capillary electrophoresis current

Capillary zone electrophoresis current

© 2024 chempedia.info