Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Capillary position

It was found, that at standard gas-chromatograph sampling of 1 pL of analyte solution the limit of detection for different amines was measured as 0.1-3 ng/ml, or of about 1 femtomole of analyte in the probe. This detection limit is better of published data, obtained by conventional GC-MS technique. Evidently, that both the increasing of the laser spot size and the optimization of GC-capillary position can strongly improve the detection limit. [Pg.103]

The nebulizer capillary position may be adjustable on a screw thread to permit optimization of sample uptake and drop size. Alternatively or additionally, an impact bead may be placed in the path of the initial aerosol to provide a secondary fragmentation and so improve the efficiency of nebulization. Such a device is illustrated in Fig. 2.9. [Pg.28]

Capillary electrophoresis separations are dependent on the relative mobilities of analytes under the influence of an electric field and do not depend on mobile phase/stationary phase interactions. A fused silica capillary is filled with a buffer and both ends submerged into two reservoirs of the buffer. A platinum electrode is immersed in each reservoir and a potential difference (5-30 kV) is applied across the electrode. An aliquot of sample of a few nanoliters is injected onto the capillary by either hydrostatic or electrokinetic injection, and the components migrate to the negative electrode. Separations of analytes arise from differences in the electrophoretic mobilities, which are dependent on the mass-to-charge ratio of the components, physical size of the analyte, and buffer/analyte interactions. An electro-osmotic flow (EOF) of the buffer occurs in the capillary and arises as a result of interactions of the buffer with dissociated functional groups on the surface of the capillary. Positive ions from the buffer solution are attracted to negative ions... [Pg.399]

As in all flame spectrometric techniques, care should be taken to make sure that the nebulizer performance, if the nebulizer is adjustable, is optimized. This may involve direct adjustment of the nebulizer capillary position or minor adjustments to the position of the impactor, if fitted. [Pg.56]

A schemahc diagram of the DEMS apparatus is shown in Fig. 5. The electrochemistry compartment corrsists of a circular block of passivated htanirrm (a) that rests above a stainless-steel support (1) cormected to the mass spectrometer. The space between the cell body and the snpport is a Teflon membrane (j) embedded on a steel mesh (k) the membrane is 75 pm thick, has 50% porosity and pore width of 0.02 pm. The single-crystal disk (h) is the working electrode its face is in contact with the electrolyte solution and separated from the cell body by another Teflon membrane (i) that functions as a spacer to form a ca 100-pm thick electrolyte layer (j). Stop-flow or continnons-flow electrolysis can be performed with this arrangement. For the latter, flow rates have to be minimal, ca 1 pL/s, to allow ample time (ca 2 s) for the electrogenerated products to diffuse to the upper Teflon membrane. Two capillaries positioned at opposite sides of the cell body (b, e) serve as electrolyte inlet and outlet as well as connection ports to the reference (f) and two auxiliary Pt-wire electrodes (d, f). [Pg.285]

Some electrospray parameters are known to be critical for achieving stable conditions and thereby good quantitative results. These parameters are the sheath liquid composition and flow rate, the nebulizing gas pressure, the applied electrospray voltage and the capillary outlet position. On the other hand, in previous studies, the impact of drying gas flow rate and temperature on stability and sensitivity were demonstrated to be moderate [3, 76]. Most of the quoted parameters are well described in the literature, apart from the capillary position which is disregarded for CE-ESI-... [Pg.276]

As a result, their intensity can be adjusted by modifying either the applied electrospray voltage or the position of the CE capillary outlet. Prior to starting a sequence, analyses have to be performed to ensure that these electric currents remain stable, otherwise, the capillary position has to be slightly adjusted to achieve good quantitative results, furthermore, the capillary current monitoring can be a diagnostic tool of the chiral selector s entrance into the nebulisation chamber. [Pg.277]

Figure 3.7 First coupling of a nanoLC column to the ESI Chip using a liquid electrode to apply spray voltage to the effluent exiting a capillary positioned at the inlet to the ESI Chip. Make-up flow was optionally applied to adjust solvent composition to enhance ionization. Figure 3.7 First coupling of a nanoLC column to the ESI Chip using a liquid electrode to apply spray voltage to the effluent exiting a capillary positioned at the inlet to the ESI Chip. Make-up flow was optionally applied to adjust solvent composition to enhance ionization.
Figure 4.4.17. Description of the principal construction of a Knauer membrane osmometer A 300 1 - head thermostat, 2 - channel for syringe, 3 - calibration device with suction tube, 4 - calibration glass, 5 - c illary position MEASUREMENT, 6 - capillary position CALIBRATION, 7 - tension screws, 8 - cell retaining disc, 9 - upper half of measuring cell, 10 - sample introduction system, 11 -semipermeable membrane, 12 - lower half of measuring cell, 13 -pressure measuring system, 14 - cell thermostat, 15 - suction of calibration bottle. [Reprinted from the operating manual with permission from Dr. H. Knauer GmbH (Germany)]. Figure 4.4.17. Description of the principal construction of a Knauer membrane osmometer A 300 1 - head thermostat, 2 - channel for syringe, 3 - calibration device with suction tube, 4 - calibration glass, 5 - c illary position MEASUREMENT, 6 - capillary position CALIBRATION, 7 - tension screws, 8 - cell retaining disc, 9 - upper half of measuring cell, 10 - sample introduction system, 11 -semipermeable membrane, 12 - lower half of measuring cell, 13 -pressure measuring system, 14 - cell thermostat, 15 - suction of calibration bottle. [Reprinted from the operating manual with permission from Dr. H. Knauer GmbH (Germany)].
When the time period is plotted against the length of the capillary, positive slope is obtained. Furthermore, when time period is plotted against log d, straight lines with a negative slope are obtained where d = 2a. [Pg.206]

The following procedure helps to prevent breakage of the micropipette tip during its centering in the microscope field. It is applicable when using the Zeiss automated injection system but can be easily adapted to manual injection. All steps should be performed in the capillary position Cell (AIS program menu). [Pg.25]

Switch the capillary position to Wait (AIS program menu). [Pg.26]

To 50 ml of sample are added 0.1 mL of sulphuric acid (reagent 1) and bromine water (reagent 8) drop by drop until the solution remains pale yellow for about 5 min. The excess bromine is removed by a fast stream of air fi-om a capillary positioned above the surface of the solution. For precision measurements, correct for the dilution caused by the added sulphuric acid and bromine water (about 0.4 %). [Pg.196]

Figure D.3 Schematic of a Luggin capillary positioning a reference electrode in close proximity to an electrochemical cell working electrode. Figure D.3 Schematic of a Luggin capillary positioning a reference electrode in close proximity to an electrochemical cell working electrode.

See other pages where Capillary position is mentioned: [Pg.295]    [Pg.45]    [Pg.493]    [Pg.993]    [Pg.192]    [Pg.69]    [Pg.200]    [Pg.310]    [Pg.102]    [Pg.288]    [Pg.276]    [Pg.488]    [Pg.20]    [Pg.533]    [Pg.26]    [Pg.292]    [Pg.447]    [Pg.111]    [Pg.913]    [Pg.38]    [Pg.127]    [Pg.243]    [Pg.75]    [Pg.1550]   
See also in sourсe #XX -- [ Pg.7 ]




SEARCH



© 2024 chempedia.info