Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Calibration alternative models

In Table IV we present the results of fitting alternative models to the pattern of weights and the calibration curve. Before using the results in Tables III and IV to calculate detection limits. [Pg.62]

Figure 7 Model of the descent of life following the standard model of Woese (1987), as calibrated by the geological evidence (source Shen et al, 2001, and other evidence). See Figure 11 for alternative model. Figure 7 Model of the descent of life following the standard model of Woese (1987), as calibrated by the geological evidence (source Shen et al, 2001, and other evidence). See Figure 11 for alternative model.
Alternative models may be developed to include additional covariates which are not measured with error, e.g., X = f(R,Z). The classical model is used when an attempt to measure x is made but cannot be done so due to various measurement errors. An example of this is the measurement of blood pressure. There is only one true blood pressure reading for a subject at a particular point in time, but due to minor calibration errors in the instrument, transient increases in blood pressure due to diet, etc., possible recording errors and reading errors by the nurse, etc., blood pressure is a composite variable that can vary substantially both within and between days. In this case it makes sense to try and model the observed blood pressure using Eq. (2.84). Under this model, the expected value of X is x. In regression calibration problems, the focus is on the distribution of x given X. For purposes herein, the focus will be on the classical error model. The reader is referred to Fuller (1987) and Carroll et al. (1995) for a more complete exposition of the problem. [Pg.80]

The two previous chapters introduced and described a fractiOTi of the most important research into interest-rate models that has been carried out since the first model, presented by Oldrich Vasicek, appeared in 1977. These models can be used to price derivative seciuities, and equitibrium models can be used to assess fair value in the bond market. Before this can take place however, a model must be fitted to the yield curve, or calibrated In practice, this is carried out in two ways the most popular approach involves calibrating the model against market interest rates given by instruments such as cash Libor deposits, futures, swaps and bonds. The alternative method is to model the yield curve from the market rates and then calibrate the model to this fitted yield curve. The first approach is common when using, for example extended Vasicek... [Pg.85]

Alternatively, if we think that interest rates will be higher in the future, we can assume a drift factor greater than 0%. In this case, we can calibrate the model yield curve with market yield curve. In other words, we adopt an iterative procedure in which we set the pricing error between the model and market yield curve equal to 0 by changing the drift factor. [Pg.227]

Appropriate techniques, tools and processes Design tool calibration and validation Quantitative analysis of alternatives Modeling, simulation and test Analytical refinement of the design... [Pg.284]

The trend is that the closer the residual variances of the two models are, the more calibrators are required by Mandel s test to exceed the critical F values (to reject the null hypothesis i.e. to conclude that the variance explained by the additional term of the alternative model is significant). This is due to the similarity between the models. [Pg.129]

If the variances of the two models are similar (which may be a common practical situation e.g. when the linear and non-linear models are very similar and they differ by 2%), the number of calibration solutions to recognize such a difference would be so high ( = 145 or 283, see Table Al.l and Figure Al.l), that we will rarely perform such a calibration and we will not be able to reject the null hypothesis i.e. the alternative model will not be significant). At the limit, if the variances tend to be almost the same, the quadratic term (or other alternative polynomial) will not be significant, FexpjupAc will b 0 Fexp.Mandei = 1, and the difference will obviously never exceed Ftab(cc%,l,n - 3). [Pg.129]

The first one we mention is the question of the validity of a test set. We all know and agree (at least, we hope that we all do) that the best way to test a calibration model, whether it is a quantitative or a qualitative model, is to have some samples in reserve, that are not included among the ones on which the calibration calculations are based, and use those samples as validation samples (sometimes called test samples or prediction samples or known samples). The question is, how can we define a proper validation set Alternatively, what criteria can we use to ascertain whether a given set of samples constitutes an adequate set for testing the calibration model at hand ... [Pg.135]

In the text which follows we shall examine in numerical detail the decision levels and detection limits for the Fenval-erate calibration data set ( set-B ) provided by D. Kurtz (17). In order to calculate said detection limits it was necessary to assign and fit models both to the variance as a function of concentration and the response (i.e., calibration curve) as a function of concentration. No simple model (2, 3 parameter) was found that was consistent with the empirical calibration curve and the replication error, so several alternative simple functions were used to illustrate the approach for calibration curve detection limits. A more appropriate treatment would require a new design including real blanks and Fenvalerate standards spanning the region from zero to a few times the detection limit. Detailed calculations are given in the Appendix and summarized in Table V. [Pg.58]

Two procedures for improving precision in calibration curve-based-analysis are described. A multiple curve procedure is used to compensate for poor mathematical models. A weighted least squares procedure is used to compensate for non-constant variance. Confidence band statistics are used to choose between alternative calibration strategies and to measure precision and dynamic range. [Pg.115]

Bauer et al. describe the use of a noncontact probe coupled by fiber optics to an FT-Raman system to measure the percentage of dry extractibles and styrene monomer in a styrene/butadiene latex emulsion polymerization reaction using PLS models [201]. Elizalde et al. have examined the use of Raman spectroscopy to monitor the emulsion polymerization of n-butyl acrylate with methyl methacrylate under starved, or low monomer [202], and with high soUds-content [203] conditions. In both cases, models could be built to predict multiple properties, including solids content, residual monomer, and cumulative copolymer composition. Another study compared reaction calorimetry and Raman spectroscopy for monitoring n-butyl acrylate/methyl methacrylate and for vinyl acetate/butyl acrylate, under conditions of normal and instantaneous conversion [204], Both techniques performed well for normal conversion conditions and for overall conversion estimate, but Raman spectroscopy was better at estimating free monomer concentration and instantaneous conversion rate. However, the authors also point out that in certain situations, alternative techniques such as calorimetry can be cheaper, faster, and often easier to maintain accurate models for than Raman spectroscopy, hi a subsequent article, Elizalde et al. found that updating calibration models after... [Pg.223]

In Other words, the assurance of quality by measurement of process impurities in the end product has been replaced by assurance of quality by the removal of variance in the process (by continuous monitoring of a continuous process). Naturally, whether online process analysis is being used as a surrogate for an alternative off-line technique to measure specific analytes or as a monitor to reduce process variance it needs calibration and validation. These stages require measurement of process analytes by a reference off-line technique, usually HPLC, and subsequent demonstration that the resulting calibration model has reliable predictive power. [Pg.252]

We have used density functional calculations to calibrate intensity factors for the DNA and RNA bases (43). While these are not necessarily optimal values, they are used here to provide an initial idea of the performance of ring-current models in explaining DNA shift patterns. A more extensive test of alternative values is in progress. [Pg.200]


See other pages where Calibration alternative models is mentioned: [Pg.156]    [Pg.403]    [Pg.2]    [Pg.355]    [Pg.334]    [Pg.222]    [Pg.286]    [Pg.18]    [Pg.95]    [Pg.349]    [Pg.513]    [Pg.131]    [Pg.353]    [Pg.378]    [Pg.182]    [Pg.522]    [Pg.302]    [Pg.208]    [Pg.70]    [Pg.62]    [Pg.93]    [Pg.144]    [Pg.210]    [Pg.218]    [Pg.119]    [Pg.423]    [Pg.397]    [Pg.399]    [Pg.418]    [Pg.162]    [Pg.233]    [Pg.265]    [Pg.271]    [Pg.11]    [Pg.85]    [Pg.157]    [Pg.186]   
See also in sourсe #XX -- [ Pg.62 ]




SEARCH



Alternate models

Alternative models

© 2024 chempedia.info