Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Analytical variables, measures

Dimensionality of analytical data. Analytical data are present either in the form of measured values yt or analytical results x,. Multivariate data, i.e., results of m variables (e.g., analyte concentrations) measured at n different samples, are mostly represented in the form of data sets and data matrices ... [Pg.85]

Horwitz claims that irrespective of the complexity found within various analytical methods the limits of analytical variability can be expressed or summarized by plotting the calculated mean coefficient of variation (CV), expressed as powers of two [ordinate], against the analyte level measured, expressed as powers of 10 [abscissa]. In an analysis of 150 independent Association of Official Analytical Chemists (AOAC) interlaboratory collaborative studies covering numerous methods, such as chromatography, atomic absorption, molecular absorption spectroscopy, spectrophotometry, and bioassay, it appears that the relationship describing the CV of an analytical method and the absolute analyte concentration is independent of the analyte type or the method used for detection. [Pg.483]

To learn that the changes in concentration caused by current flow will follow Faraday s laws, so the analytical variable during measurement is current, where / oc Canaiyte-... [Pg.131]

Limit of detection (LOD) sounds like a term that is easily defined and measured. It presumably is the smallest concentration of analyte that can be determined to be actually present, even if the quantification has large uncertainty. The problem is the need to balance false positives (concluding the analyte is present, when it is not) and false negatives (concluding the analyte is absent, when it is really present). The International Union of Pure and Applied Chemistry (IUPAC) and ISO both shy away from the words limit of detection, arguing that this term implies a clearly defined cutoff above which the analyte is measured and below which it is not. The IUPAC and ISO prefer minimum detectable (true) value and minimum detectable value of the net state variable, which in analytical chemistry would become minimum detectable net concentration. Note that the LOD will depend on the matrix and therefore must be validated for any matrices likely to be encountered in the use of the method. These will, of course, be described in the method validation document. [Pg.238]

SNR/i is used to denote the analyte shot noise limit, and a. Da, La refer only to analyte variables. Recall that ts is the measurement time for each resolution element and does not necessarily equal the total measurement time As expected, SNR varies with... [Pg.63]

Among the different analytical variables listed in Table 5.10, specific gravity, refractive index and optical rotation are widely used as quality indexes. The values of these properties allow us to estimate the (/-limonene content, general composition and adulterations in essential oils. Aldehyde content is also a widely used index in the characterization of essential oils, because it is a measure of the aromatic fraction values increase in mature fruit and decrease during fruit storage (Lopez, 1995). The UV absorption index indicates general quality and adulterations, and shows differences between the different types of essential oils (Lopez, 1995). [Pg.180]

Many analytical variables must be controlled carefully to assure accurate measurements by analytical methods. Reliable analytical methods are obtained by a careful process of selection, evaluation, implementation, maintenance, and control (see Chapter 14). Efficient, effective, and uninterrupted laboratory service requires many procedures aimed at preventing the occurrence of problems. Laboratories may experience different problems with the same analytical methods owing to different amounts of effort being allocated to the care and support of those methods. [Pg.494]

Preanalyticai and analytical variables significantly reduce the usefulness of measurements of markers of bone formation and resorption. The long-term, within-individual variability of urine markers is generally higher (15% to 60%) than that of serum markers (5% to ioo/o).3i> i >345.550,586... [Pg.1936]

Many analytical measures cannot be represented as a time-series in the form of a spectrum, but are comprised of discrete measurements, e.g. compositional or trace analysis. Data reduction can still play an important role in such cases. The interpretation of many multivariate problems can be simplified by considering not only the original variables but also linear combinations of them. That is, a new set of variables can be constructed each of which contains a sum of the original variables each suitably weighted. These linear combinations can be derived on an ad hoc basis or more formally using established mathematical techniques. Whatever the method used, however, the aim is to reduce the number of variables considered in subsequent analysis and obtain an improved representation of the original data. The number of variables measured is not reduced. [Pg.64]

One physician who measured serum alkaline phosphatase in himself repeatedly over a period of 8 years, found that despite the prolonged period of observation, the relative standard deviation of the values obtained was only 6.5% (Mc3). When this figure is compared with the relative standard deviation of 4-5 % for the analytical variability of the method employed, it seems highly probable that the true intrasubject variation was smaller than indicated by the relative standard deviation and that the observed fluctuations were due, in part, to test imprecision. [Pg.176]

Sensitivity and precision are interrelated in an assay system. Precision is defined as the reproducibility of replicate analyses at different levels of the analyte, within assay and between assays. The variability of the measurement is dependent on the concentration of the analyte being measured. Sensitivity is defined as the level of measure for a nonzero quantity that can be measured with a predetermined precision. The term low limit of quantitation (LLOQ) is often used as... [Pg.242]

The highest precision for quant iiaiivc chromatography is obtained using internal standards because the uncertainties introduced by sample injection are avoided In this procedure, a carefully measured quantity of an internal-standard subsLitnee is introduced into each standard and. sample, and the ratio of analyte to Internal standard peak areas (or heights) serves as the analytical variable. For this method to be successful, the internal-standard peak must be well separated from the peaks of all other components of the sample H > 1.2.5) the internal-standard peak should, on the other hand, appear do.se to the analyte peak. With a suitable internal standard, precisions of belter than 1 % relative can usually be achieved. [Pg.783]

In my lecture I am warning you that the analytical chemist is nowhere near as good as he thinks he is or that he makes it appear that he is. For those who do not wish to be confused by variability, the analytical chemist will give you a number. But the variability is still there. To properly interpret chemical values in terms of biological phenomena, the analytical variability must be removed to assure that the final results are truly of toxicological significance and not merely the analytical error of the chemist. The toxicologist must be particularly wary when the chemist operates near the limits of measurement. It appears... [Pg.333]

A diagnosis procedure based on evaluation of physical constraint laws derived from bond graph models is described here. Symbolically written constraints, called analytical redundancy relations (ARRs), are expressed in terms of known variables (measurements and inputs). ARRs are static or dynamic constraints which link the time evolution of the known variables when a system operates according to its normal operation model. The error or deviation from the constraint model is called a residual. The objective of quantitative diagnosis is to evaluate the residuals and associate the fault symptoms with deviations of residuals. [Pg.244]

Analytical variables are associated with devices designed to measure the composition of a substance. For example, process technicians use analyzers to determine the percentage of a substance in a process stream. Analyzers come in a variety of shapes and designs, and can measure the concentration of a specific chemical or element. Other examples of analytical process variables include pH or parts per million (ppm). These variables are frequently tracked on a cooling water system. Plastic plant technicians check for melt flow, color, and the concentration of special additives. Figure 7-5 shows an assortment of basic instruments used in the chemical processing industry. [Pg.173]


See other pages where Analytical variables, measures is mentioned: [Pg.254]    [Pg.255]    [Pg.72]    [Pg.105]    [Pg.391]    [Pg.14]    [Pg.82]    [Pg.250]    [Pg.246]    [Pg.129]    [Pg.96]    [Pg.170]    [Pg.3]    [Pg.513]    [Pg.551]    [Pg.672]    [Pg.2]    [Pg.436]    [Pg.277]    [Pg.283]    [Pg.321]    [Pg.167]    [Pg.426]    [Pg.366]    [Pg.344]    [Pg.497]    [Pg.178]    [Pg.301]    [Pg.620]    [Pg.512]   
See also in sourсe #XX -- [ Pg.173 , Pg.196 ]




SEARCH



Analyte Analytical measurement

Analyte, measurement

Analytical measurement

Variability measurement

Variability, measures

Variables measurement

© 2024 chempedia.info