Amperometry amperometric

Ampere, Andr6 current, electricity current amp (unit of current) amperometry amperometric voltammetry... [Pg.610]

Amperometry. Amperometric methods measure the current produced at a working electrode in response to an applied potential. Amperometric enzyme assays rely on the production of an oxidizable or reducible species from an enzyme-catalyzed reaction. The applied potential is extreme enough to completely oxidize (at positive potentials) or reduce (at negative potentials) any analyte that contacts the working electrode. In stirred or unstirred solutions, the current produced under such mass-transport-controlled conditions is directly proportional to analyte concentration. [Pg.53]

Amperometric Sensors A number of voltammetric systems are produced commercially for the determination of specific species of interest in industry and research. These are usually based on measuring the limiting current at a constant applied potential and relating the measured current to concentration. This technique is often called amperometry. Amperometric devices are sometimes called electrodes but are, in fact, complete voltammetric cells and are better referred to as sensors. Two of these devices are described here. [Pg.681]

One important application of amperometry is in the construction of chemical sensors. One of the first amperometric sensors to be developed was for dissolved O2 in blood, which was developed in 1956 by L. C. Clark. The design of the amperometric sensor is shown in Figure 11.38 and is similar to potentiometric membrane electrodes. A gas-permeable membrane is stretched across the end of the sensor and is separated from the working and counter electrodes by a thin solution of KCl. The working electrode is a Pt disk cathode, and an Ag ring anode is the... [Pg.519]

Selecting the Voltammetric Technique The choice of which voltammetric technique to use depends on the sample s characteristics, including the analyte s expected concentration and the location of the sample. Amperometry is best suited for use as a detector in flow systems or as a selective sensor for the rapid analysis of a single analyte. The portability of amperometric sensors, which are similar to po-tentiometric sensors, make them ideal for field studies. [Pg.520]

Electrochemical Detectors Another common group of HPLC detectors are those based on electrochemical measurements such as amperometry, voltammetry, coulometry, and conductivity. Figure 12.29b, for example, shows an amperometric flow cell. Effluent from the column passes over the working electrode, which is held at a potential favorable for oxidizing or reducing the analytes. The potential is held constant relative to a downstream reference electrode, and the current flowing between the working and auxiliary electrodes is measured. Detection limits for amperometric electrochemical detection are 10 pg-1 ng of injected analyte. [Pg.585]

Amperometry refers to measurement of current under a constant applied voltage and under these conditions it is the concentration of the analyte which determines the magnitude of the current. Such measurements may be used to follow the change in concentration of a given ion during a titration, and thus to fix the end point this procedure is referred to as amperometric titration. [Pg.591]

Other important alternate electrochemical methods under study for pCO rely on measuring current associated with the direct reduction of CO. The electrochemistry of COj in both aqueous and non-aqueous media has been documented for some time 27-29) interferences from more easily reduced species such as O2 as well as many commonly used inhalation anesthetics have made the direct amperometric approach difficult to implement. One recently described attempt to circumvent some of these interference problems employs a two cathode configuration in which one electrode is used to scrub the sample of O by exhaustive reduction prior to COj amperometry at the second electrode. The response time and sensitivity of the approach may prove to be adequate for blood ps applications, but the issue of interfering anesthetics must be addressed more thorou ly in order to make the technique a truly viable alternative to the presently used indirect potentiometric electrode. [Pg.55]

SECM employs a mobile UME tip (Fig. 3) to probe the properties of a target interface. Although both amperometric and potentiometric electrodes have found application in SECM, amperometry - in which a target species is consumed or generated at the probe UME - has found the most widespread use in kinetic studies at liquid interfaces, as... [Pg.293]

The use of non-inert and chemically modified electrodes and other strategies for the detection of species that are difficult to analyze with the normal electrode materials have been reviewed.55 Photosensitization prior to amperometric detection is another tactic that has proved useful for the analysis of substances that are normally considered to be electrochemically inert.56 The use of pulsed amperometry has recently been reviewed.57... [Pg.221]

Coulometry and amperometry can be distinguished by the extent to which the analyte undergoes a Faradaic reaction at the working electrode, namely complete and partial, respectively. Coulometry is essentially high-efficiency amperometry with working electrodes of large surface area. Successful coulometric or amperometric detection can result only if the applied potential is chosen correctly. [Pg.673]

Again for the titration of Ce(IV) with Fe(II) we shall now consider constant-potential amperometry at one Pt indicator electrode and do so on the basis of the voltammetric curves in Fig. 3.71. One can make a choice from three potentials eu e2 and e3, where the curves are virtually horizontal. Fig. 3.74 shows the current changes concerned during titration at e1 there is no deflection at all as it concerns Fe(III) and Fe(II) only at e2 and e3 there is a deflection at A = 1 but only to an extent determined by the ratio of the it values of the Ce and Fe redox couples. The establishment of the deflection point is easiest at e2 as it simply agrees with the intersection with the zero-current abscissa as being the equivalence point in fact, no deflection is needed in order to determine this intersection point, but if there is a deflection, the amperometric method is not useful compared with the non-faradaic potentiometric titration unless the concentration of analyte is too low. [Pg.214]

Z. Samec, E. Samcova, and H.H. Girault, Ion amperometry at the interface between two immiscible electrolyte solutions in view of realizing the amperometric ion-selective electrode. Talcmta 63, 21—32 (2004). [Pg.135]

The concept of amperometry can also be applied to a titration experiment, much like potential measurements were in Section 14.5 (potentiometric titration). Such an experiment is called an amperometric titration, a titration in which the end point is detected through the measurement of the current flowing at an electrode. [Pg.407]

An amperometric technique relies on the current passing through a polarizable electrode. The magnitude of the current is in direct proportion to the concentration of the electroanalyte, with the most common amperometric techniques being polarography and voltammetry. The apparatus needed for amperometric measurement tends to be more expensive than those used for potentiometric measurements alone. It should also be noted that amperometric measurements can be overly sensitive to impurities such as gaseous oxygen dissolved in the solution, and to capacitance effects at the electrode. Nevertheless, amperometry is a much more versatile tool than potentiometry. [Pg.3]

The differences between potentiometry and amperometry are summarized in Table 1.1. It will be seen that amperometric measurements are generally more precise and more versatile than those made by using potentiometry, so the majority of this book will therefore be concerned with amperometric measurements. [Pg.3]

Amperometry The techniques and methodology of determining a concentration as a function of current the most popular amperometric measurement technique is voltammetry. [Pg.337]

The analytic principles that have been applied to accumulate air quality data are colorimetry, amperometry, chemiluminescence, and ultraviolet absorption. Calorimetric and amperometric continuous analyzers that use wet chemical techniques (reagent solutions) have been in use as ambient-air monitors for many years. Chemiluminescent analyzers, which measure the amount of chemiluminescence produced when ozone reacts with a gas or solid, were developed to provide a specific and sensitive analysis for ozone and have also been field-tested. Ultraviolet-absorption analyzers are based on a physical detection principle, the absorption of ultraviolet radiation by a substance. They do not use chemical reagents, gases, or solids in their operation and have only recently been field-tested. Ultraviolet-absorption analyzers are ideal as transfer standards, but, as discussed earlier, they have limitations as air monitors, because aerosols, mercury vapor, and some hydrocarbons could, interfere with the accuracy of ozone measurements made in polluted air. [Pg.262]

It is used in IC systems when the amperometric process confers selectivity to the determination of the analytes. The operative modes employed in the amperometric techniques for detection in flow systems include those at (1) constant potential, where the current is measured in continuous mode, (2) at pulsed potential with sampling of the current at dehned periods of time (pulsed amperometry, PAD), or (3) at pulsed potential with integration of the current at defined periods of time (integrated pulsed amperometry, IPAD). Amperometric techniques are successfully employed for the determination of carbohydrates, catecholamines, phenols, cyanide, iodide, amines, etc., even if, for optimal detection, it is often required to change the mobile-phase conditions. This is the case of the detection of biogenic amines separated by cation-exchange in acidic eluent and detected by IPAD at the Au electrode after the post-column addition of a pH modiher (NaOH) [262]. [Pg.409]

In amperometry, the current at the working electrode is proportional to analyte concentration. The amperometric glucose monitor generates H202 by enzymatic oxidation of glucose and the H202 is measured by oxidation at an electrode. A mediator is employed to rapidly shuttle electrons between electrode and analyte. [Pg.372]

It is now clear that electrochemical measurements can often have significant advantages over the classical spectroscopic approaches. Amperometry can be more specific therefore, lower detection limits are often feasible. Because electrochemical detectors do not require optical carriers, they can be much less expensive than UV absorption or fluorescence detectors. This is especially true when one considers that electrochemical detectors are inherently tunable without the need for such things as monochrometers or filters. On the other hand, there can be significant problems with reliability, and, more often than not, there is a lack of acceptance by chemists weaned on Beer s law. Amperometric methods in biochemistry are just beginning to be commercialized, and it is now almost certain that they will come into widespread use. [Pg.815]

All of the fat-soluble vitamins, including provitamin carotenoids, exhibit some form of electrochemical activity. Both amperometry and coulometry have been applied to electrochemical detection. In amperometric detectors, only a small proportion (usually <20%) of the electroactive solute is reduced or oxidized at the surface of a glassy carbon or similar nonporous electrode in coulometric detectors, the solute is completely reduced or oxidized within the pores of a graphite electrode. The operation of an electrochemical detector requires a semiaqueous or alcoholic mobile phase to support the electrolyte needed to conduct a current. This restricts its use to reverse-phase HPLC (but not NARP) unless the electrolyte is added postcolumn. Electrochemical detection is incompatible with NARP chromatography, because the mobile phase is insufficiently polar to dissolve the electrolyte. A stringent requirement for electrochemical detection is that the solvent delivery system be virtually pulse-free. [Pg.356]

A dual EC detection employing amperometry/conductivity allows measurements of a wider range of analytes per separation (ionic via CCD mode and electroactive compounds via amperometric detection) has also been reported [73]. This scheme also gave two simultaneous signals for the same component that is both charged and electroactive. This is the only report in which both principles are combined. Most of the CE-EC microchips are based in the amperometric detection mode that is more deeply considered in next section. [Pg.836]

Amperometry is the most widely reported EC detection mode for CE microchips, which primarily relies on oxidation or reduction of elect-rochemically active species by applying a constant potential to a working electrode. The current is then monitored as a function of time. Since it is based on the redox reaction that occurs at the electrode surface, electrodes can be miniaturised without loss in sensitivity. The relevance of this simple technique is reported in several reviews [48,74], In this section, a general overview of the combination of this detection technique to CE microchips together with special sections for different amperometric techniques and electrode materials and types are considered. [Pg.837]


See other pages where Amperometry amperometric is mentioned: [Pg.1575]    [Pg.250]    [Pg.1575]    [Pg.250]    [Pg.103]    [Pg.295]    [Pg.221]    [Pg.360]    [Pg.379]    [Pg.379]    [Pg.254]    [Pg.99]    [Pg.271]    [Pg.16]    [Pg.103]    [Pg.685]    [Pg.815]    [Pg.816]    [Pg.387]    [Pg.733]    [Pg.844]    [Pg.881]    [Pg.1067]    [Pg.81]    [Pg.208]   