Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Zero control

Fill a test-tube with distilled water or with a blank solution to a depth of not less than 3 cm, and set to zero by means of the zero control. [Pg.728]

LOAD, LOADl, LOAD2 = terms in the process model to accommodate non-zero steady state measurements (present values) at zero controller output, i.e. if the heaters are off (0 power) the temperature will not be 0. Also can be changed to represent undesired disturbances to the system affecting control. [Pg.502]

Chase GW and Long AR (1997) Nutritional Metrology, PART 4, The Role of a Zero Control Reference Material. Food Testing and Analysis, Jun/Jul 1997, 3 30-33. [Pg.291]

The GCZero control. This is a control that helps set the zero position on the chart recorder, but it is not to be confused with the zero control on the chart recorder. [Pg.236]

Now set the pen on the chart recorder to zero using only the chart recorder zero control. Once you do that, leave the chart recorder alone. [Pg.237]

If the pen on the chart recorder moves off zero, use the GC zero control only to bring the pen back to the zero line on the chart recorder paper. [Pg.237]

Do not touch the chart recorder zero. Use the GC zero control only. [Pg.237]

As the coarse attenuator gets to more sensitive settings (lower numbers), it becomes more difficult to adjust the chart recorder pen to zero using only the GC zero control. Do the best you can at the lowest attenuation (highest sensitivity) you can hold a zero steady at. [Pg.237]

Insert the Blank solution in the nephelometer and adjust to zero reading of the scale by the aid of zero-control-knob,... [Pg.290]

A test can meet these standards by correctly identifying 30 of 30 zero control samples and 30 of 30 samples containing each claimed drug at its tolerance/safe level. With this selectivity standard for acceptance, the probability is low that an accepted screening test would produce a positive result on a sample that does not contain any of the drug it is designed to detect. [Pg.776]

FIG. 14 Effect of lycopene on ALP activity p < 0.005, comparison with zero control a < 0.05, b < 0.01 (Kim et al., 2003). (Lycopene II - Effect on osteoblasts The carotenoid lycopene stimulates cell proliferation and alkaline phosphotase activity of SaOS-2 cells. Reprinted from Journal of Medicinal Food. 2003 6, pp. 79-86 by permission of Mary Ann Liebert, Inc., Publishers.)... [Pg.137]

Set the 02-electrode system to amplifier zero, and adjust the amplifier zero control until the recorder registers zero. This establishes the reading that occurs in the absence of 02. You can confirm this setting by later using an 02-free water solution, such as water that has been boiled vigorously and cooled or sparged with an inert gas. [Pg.236]

Alternatively, wells can be photographed by transmitted light under a dissecting microscope. Total areas of cell outgrowth of both time zero control and test wells are digitized using a Sigma-Scan... [Pg.79]

Fig. 4-6. Detector circuits for vapor-phase chromatography, (a) Thermistor detector Di, D=, Victory Eng. Corp. 32A12 thermistors Ri, Ri, 1,0000 wire wound resistors Rz, 1,0000 Helipot Rt, 10,0000 1% carbon film resistor Rs, 5,0000 1 % carbon film resistor Rt, Rj, 2,5000 1 % carbon film resistor Sw, single-pole four-position switch, (b) Hot-wire detector Rz, filament current control, to adjust filament current between 150—300 ma ( 20 ohm 5w) Ri, R4, reference detectors Rs, R, sample detectors Re, zero control 20 Re, 600 1 % carbon film resistor R, 300 1 /, carbon film resistor R, 150 1% carbon film resistor Rse, 7.5Q 1 % carbon film resistor Rs, 7.SCI 1 % carbon film resistor M, 300 ma meter,- S, single-pole six-position sv/itch. Fig. 4-6. Detector circuits for vapor-phase chromatography, (a) Thermistor detector Di, D=, Victory Eng. Corp. 32A12 thermistors Ri, Ri, 1,0000 wire wound resistors Rz, 1,0000 Helipot Rt, 10,0000 1% carbon film resistor Rs, 5,0000 1 % carbon film resistor Rt, Rj, 2,5000 1 % carbon film resistor Sw, single-pole four-position switch, (b) Hot-wire detector Rz, filament current control, to adjust filament current between 150—300 ma ( 20 ohm 5w) Ri, R4, reference detectors Rs, R, sample detectors Re, zero control 20 Re, 600 1 % carbon film resistor R, 300 1 /, carbon film resistor R, 150 1% carbon film resistor Rse, 7.5Q 1 % carbon film resistor Rs, 7.SCI 1 % carbon film resistor M, 300 ma meter,- S, single-pole six-position sv/itch.
The message level was assumed to be at steady-state at time zero (control animals), yielding the baseline equation... [Pg.518]

The 120 cps output signal from the detector is fed to the converter, where it is converted into a 0-10 MVDC signal compatible with 0-10 MVDC recorder. Span and zero controls are located at the convertor. The laboratory set-up is shown in Figure 1. [Pg.293]

Czeizel and Kiraly (1976) carried out chromosome tests in male labourers 20-61 years of age engaged in the manufacture of TCPE and Buvinol . Using the method of Moorhead et al. (1960) they investigated the lymphocytes of peripheral blood after 48 hours culture. The Edinburgh method (Buckton et al, 1962 Court-Brown et al., 1967 Buckton and Evans, 1973) was used for evaluation, that is, the numerical distribution of chromosomes and the chromatid and chromosome-type injuries were recorded. The modal and nonmodal chromosome numbers of zero-control and factory control did not differ considerably. [Pg.538]

Clean conventional 1 liter jars were used. Each jar contained 1 vial containing the sample to be tested firmly placed in soil, 1 humidifying vial containing distilled water and 1 vial containing the exact quantity of the known titrated solution of NaOH. The tests are carried out in triplicate against the zero control and the micronized eellulose (Figure 14.12). [Pg.326]

Set the zero control and 100% transmittance controls according to the operating manual instructions using DDL Record transmittance values for the blank calibration standards, ICV measured in triplicate, and unknown sample. [Pg.559]

If, as occasionally happens, the first tube in the full gradient has less ss DNA than the ice control, we then use it as the zero % control for normalizing. This is illustrated in the second data set, BAR-CHI. [Pg.402]


See other pages where Zero control is mentioned: [Pg.729]    [Pg.240]    [Pg.119]    [Pg.329]    [Pg.296]    [Pg.167]    [Pg.164]    [Pg.280]    [Pg.80]    [Pg.228]    [Pg.523]    [Pg.325]    [Pg.349]    [Pg.328]    [Pg.104]    [Pg.1924]    [Pg.1925]    [Pg.40]    [Pg.65]    [Pg.718]    [Pg.589]    [Pg.273]    [Pg.183]    [Pg.83]    [Pg.451]    [Pg.30]   
See also in sourсe #XX -- [ Pg.280 ]




SEARCH



Oral drug delivery zero-order controlled release

Release zero-order controlled

Side-effects zero-order controlled release

© 2024 chempedia.info