Uses of Array Collectors

The major advantage of array detectors over point ion detectors lies in their ability to measure a range of m/z values and the corresponding ion abundances all at one time, rather than sequentially. For example, suppose it takes 10 msec to measure one m/z value and the associated number of ions (abundance). To measure 100 such ions sequentially with a point ion detector would necessitate 1000 msec (1 sec) for the array detector, the time is still 10 msec because all ions arrive at the same time. Therefore, when it is important to be able to measure a range of ion m/z values in a short space of time, the array detector is advantageous. [Pg.209]

There are two common occasions when rapid measurement is preferable. The first is with ionization sources using laser desorption or radionuclides. A pulse of ions is produced in a very short interval of time, often of the order of a few nanoseconds. If the mass spectrometer takes 1 sec to attempt to scan the range of ions produced, then clearly there will be no ions left by the time the scan has completed more than a few nanoseconds (ion traps excluded). If a point ion detector were to be used for this type of pulsed ionization, then after the beginning of the scan no more ions would reach the collector because there would not be any left The array collector overcomes this difficulty by detecting the ions produced all at the same instant. [Pg.209]

A second use of arrays arises in the detection of traces of material introduced into a mass spectrometer. For such very small quantities, it may well be that the tiny amount of substance will have disappeared by the time a scan has been has been completed by a mass spectrometer equipped [Pg.209]

An array ion collector (detector) consists of a large number of miniature electron multiplier elements arranged side by side along a plane. Point ion collectors gather and detect ions sequentially (all ions are focused at one point one after another), but array collectors gather and detect all ions simultaneously (all ions are focused onto the array elements at the same time). Array detectors are particularly useful for situations in which ionization occurs within a very short space of time, as with some ionization sources, or in which only trace quantities of a substance are available. For these very short time scales, only the array collector can measure a whole spectrum or part of a spectrum satisfactorily in the time available. [Pg.210]

Collectors (Detectors) of Ions Arrays and MicroChannel Plates [Pg.211]


A second use of arrays arises in the detection of trace components of material introduced into a mass spectrometer. For such very small quantities, it may well be that, by the time a scan has been carried out by a mass spectrometer with a point ion collector, the tiny amount of substance may have disappeared before the scan has been completed. An array collector overcomes this problem. Often, the problem of detecting trace amounts of a substance using a point ion collector is overcome by measuring not the whole mass spectrum but only one characteristic m/z value (single ion monitoring or single ion detection). However, unlike array detection, this single-ion detection method does not provide the whole spectrum, and an identification based on only one m/z value may well be open to misinterpretation and error. [Pg.216]

For either the in-line or hybrid analyzers, the ions injected into the TOF section must all begin their flight down the TOF tube at the same instant if arrival times of ions at a detector are to be used to measure m/z values (see Chapter 26, TOF Ion Optics ). For the hybrid TOF instruments, the ion detector is usually a microchannel plate ion counter (see Chapter 30, Comparison of Multipoint Collectors (Detectors) of Ions Arrays and MicroChannel Plates ). [Pg.153]

Other types of mass spectrometer may use point, array, or both types of collector. The time-of-flight (TOF) instrument uses a special multichannel plate collector an ion trap can record ion arrivals either sequentially in time or all at once a Fourier-transform ion cyclotron resonance (FTICR) instrument can record ion arrivals in either time or frequency domains which are interconvertible (by the Fourier-transform technique). [Pg.201]

In modem mass spectrometry, ion collectors (detectors) are generally based on the electron multiplier and can be separated into two classes those that detect the arrival of all ions sequentially at a point (a single-point ion collector) and those that detect the arrival of all ions simultaneously (an array or multipoint collector). This chapter compares the uses of single- and multipoint ion collectors. For more detailed discussions of their construction and operation, see Chapter 28, Point Ion Collectors (Detectors), and Chapter 29, Array Collectors (Detectors). In some forms of mass spectrometry, other methods of ion detection can be used, as with ion cyclotron instmments, but these are not considered here. [Pg.211]

Other types of mass spectrometer can use point, array, or both types of ion detection. Ion trap mass spectrometers can detect ions sequentially or simultaneously and in some cases, as with ion cyclotron resonance (ICR), may not use a formal electron multiplier type of ion collector at all the ions can be detected by their different electric field frequencies in flight. [Pg.212]

To differentiate tteir functions and modes of operation, the array collector of spatially dispersed m/z values is still called an array collector for historical reasons, but the other multipoint detector of a temporally dispersed range of m/z values is called a microchannel plate (typically used in time-of-flight instruments). [Pg.410]

Advances in TIMS-techniques and the introduction of multiple collector-ICP-MS (MC-ICP-MS) techniques have enabled the research on natural variations of a wide range of transition and heavy metal systems for the first time, which so far could not have been measured with the necessary precision. The advent of MC-ICP-MS has improved the precision on isotope measurements to about 40 ppm on elements such as Zn, Cu, Fe, Cr, Mo, and Tl. The technique combines the strength of the ICP technique (high ionization efficiency for nearly all elements) with the high precision of thermal ion source mass spectrometry equipped with an array of Faraday collectors. The uptake of elements from solution and ionization in a plasma allows correction for instrument-dependent mass fractionations by addition of external spikes or the comparison of standards with samples under identical operating conditions. All MC-ICP-MS instruments need Ar as the plasma support gas, in a similar manner to that commonly used in conventional ICP-MS. Mass interferences are thus an inherent feature of this technique, which may be circumvented by using desolvating nebulisers. [Pg.33]

Flat-plate PV collectors contain an array of individual cells connected in a series/parallel circuit and encapsulated within a sandwich structure, the front of which is glass or plastic. Unlike thermal collectors, the back of these collectors is not insulated, because for best performance, they need to be cooled by the atmosphere. If this energy could be used and thereby this loss could be eliminated in new designs, the conversion efficiency could be much improved. [Pg.89]


See other pages where Uses of Array Collectors is mentioned: [Pg.209]    [Pg.209]    [Pg.209]    [Pg.209]    [Pg.216]    [Pg.216]    [Pg.205]    [Pg.205]    [Pg.210]    [Pg.212]    [Pg.216]    [Pg.219]    [Pg.44]    [Pg.247]    [Pg.495]    [Pg.323]    [Pg.355]    [Pg.137]    [Pg.188]    [Pg.1587]    [Pg.495]    [Pg.205]    [Pg.205]    [Pg.210]    [Pg.212]    [Pg.216]   