Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Spreads sources

FIGURE 19.75 Paper response curves. The lower curve is typical of coated stock, which minimizes dot spread. (Source Durbeck, R.C. and Sherr, S. 1988. Hardcopy Output Devices. Academic Press, San Diego, CA. With per-... [Pg.2154]

Localized AE sources appear during load variations, startups or shutdowns, but their positions are uniformly spread over the length of the two bodies of the header this can be seen from the histogram of the localized AE events for the front body (fig.S) and for the rear body (fig.9). [Pg.78]

At steady electrical load the background noise is normally low and fairly constant along the SH headers and with time no AE sources come up on the Unit 3 header, while the very few localized events recorded on Unit 4 are spread out over the whole length of the header. [Pg.79]

If the spreading is into a limited surface area, as in a laboratory experiment, the film front rather quickly reaches the boundaries of the trough. The film pressure at this stage is low, and the now essentially uniform film more slowly increases in v to the final equilibrium value. The rate of this second-stage process is mainly determined by the rate of release of material from the source, for example a crystal, and the surface concentration F [46]. Franses and co-workers [47] found that the rate of dissolution of hexadecanol particles sprinkled at the water surface controlled the increase in surface pressure here the slight solubility of hexadecanol in the bulk plays a role. [Pg.111]

Another consideration when gaseous samples are ionized is the variation in where the ions are fonned in the source. The above arguments assumed that the ions were all fonned at a connnon initial position, but in practice they may be fonned anywhere in the acceleration zone. The result is an additional spread in the final TOF distributions, smce ions... [Pg.1353]

The critical requirements for the ion source are that the ions have a small energy spread, there are no fast neutrals in the beam and the available energy is 1-10 keV. Both noble gas and alkali ion sources are conunon. Por TOP experunents, it is necessary to pulse the ion beam by deflecting it past an aperture. A beam line for such experiments is shown in figure B1.23.5 it is capable of producing ion pulse widths of 15 ns. [Pg.1807]

Equation (2.28), being statistical in nature, requires a large number of particles to be measured, especially if the spread of particle size is wide. The possibility of error from this source is stressed by Arnell and Henneberry who found that in a particular sample of finely ground quartz, two particles in a total of 335 had a diameter about twenty times the most probable diameter, and that if these were overlooked the calculated value of A would be nearly doubled. [Pg.63]

Precision is a measure of the spread of data about a central value and may be expressed as the range, the standard deviation, or the variance. Precision is commonly divided into two categories repeatability and reproducibility. Repeatability is the precision obtained when all measurements are made by the same analyst during a single period of laboratory work, using the same solutions and equipment. Reproducibility, on the other hand, is the precision obtained under any other set of conditions, including that between analysts, or between laboratory sessions for a single analyst. Since reproducibility includes additional sources of variability, the reproducibility of an analysis can be no better than its repeatability. [Pg.62]

Evaluating Indeterminate Error Although it is impossible to eliminate indeterminate error, its effect can be minimized if the sources and relative magnitudes of the indeterminate error are known. Indeterminate errors may be estimated by an appropriate measure of spread. Typically, a standard deviation is used, although in some cases estimated values are used. The contribution from analytical instruments and equipment are easily measured or estimated. Indeterminate errors introduced by the analyst, such as inconsistencies in the treatment of individual samples, are more difficult to estimate. [Pg.63]

As m increases, At becomes progressively smaller (compare the difference between the square roots of 1 and 2 (= 0.4) with the difference between 100 and 101 (= 0.05). Thus, the difference in arrival times of ions arriving at the detector become increasingly smaller and more difficult to differentiate as mass increases. This inherent problem is a severe restriction even without the second difficulty, which is that not all ions of any one given m/z value reach the same velocity after acceleration nor are they all formed at exactly the same point in the ion source. Therefore, even for any one m/z value, ions at each m/z reach the detector over an interval of time instead of all at one time. Clearly, where separation of flight times is very short, as with TOF instruments, the spread for individual ion m/z values means there will be overlap in arrival times between ions of closely similar m/z values. This effect (Figure 26.2) decreases available (theoretical) resolution, but it can be ameliorated by modifying the instrument to include a reflectron. [Pg.191]

TOF mass spectrometers are very robust and usable with a wide variety of ion sources and inlet systems. Having only simple electrostatic and no magnetic fields, their construction, maintenance, and calibration are usually straightforward. There is no upper theoretical mass limitation all ions can be made to proceed from source to detector. In practice, there is a mass limitation in that it becomes increasingly difficult to discriminate between times of arrival at the detector as the m/z value becomes large. This effect, coupled with the spread in arrival times for any one m/z value, means that discrimination between unit masses becomes difficult at about m/z 3000. At m/z 50,000, overlap of 50 mass units is more typical i.e., mass accuracy is no better than about 50-100 mass... [Pg.191]

The electron sources used in most sems are thermionic sources in which electrons are emitted from very hot filaments made of either tungsten (W) or lanthanum boride (LaB ). W sources are typically heated to ca 2500—3000 K in order to achieve an adequate electron brightness. LaB sources require lower temperatures to achieve the same brightness, although they need a better vacuum than W sources. Once created, these primary electrons are accelerated to some desired energy with an energy spread (which ultimately determines lateral resolution) on the order of ca 1.5 eV. [Pg.271]

Once the primary electron beam is created, it must be demagnified with condenser lenses and then focused onto the sample with objective lenses. These electron lenses are electromagnetic in nature and use electric and magnetic fields to steer the electrons. Such lenses are subject to severe spherical and chromatic aberrations. Therefore, a point primary beam source is blurred into a primary beam disk to an extent dependent on the energy and energy spread of the primary electrons. In addition, these lenses are also subject to astigmatism. AH three of these effects ultimately limit the primary beam spot size and hence, the lateral resolution achievable with sem. [Pg.271]

Acrylonitrile is combustible and ignites readily, producing toxic combustion products such as hydrogen cyanide, nitrogen oxides, and carbon monoxide. It forms explosive mixtures with air and must be handled in weU-ventilated areas and kept away from any source of ignition, since the vapor can spread to distant ignition sources and flash back. [Pg.185]

Once the source modeling is complete, the quantitative result is used in a consequence analysis to determine the impact of the release. This typically includes dispersion modeling to describe the movement of materials through the air, or a fire and explosion model to describe the consequences of a fire or explosion. Other consequence models are available to describe the spread of material through rivers and lakes, groundwater, and other media. [Pg.475]

Disposable polypropylene nonwoven fabrics are widely used as the coverstock for disposable baby diapers. The expansion of the disposable diaper market throughout the world has been the primary source of growth in the consumption of polypropylene in the fiber market. In addition, nonwoven polypropylene fabrics are used in a variety of other disposable sanitary products, such as baby wipes, adult incontinence, and feminine hygiene products. Use of polypropylene nonwovens in disposable medical apparel, such as surgical gowns, has increased as a means of reducing the spread of infection. [Pg.421]


See other pages where Spreads sources is mentioned: [Pg.909]    [Pg.323]    [Pg.129]    [Pg.131]    [Pg.909]    [Pg.323]    [Pg.129]    [Pg.131]    [Pg.170]    [Pg.444]    [Pg.507]    [Pg.932]    [Pg.1106]    [Pg.1309]    [Pg.1624]    [Pg.2061]    [Pg.2062]    [Pg.2065]    [Pg.2066]    [Pg.37]    [Pg.121]    [Pg.170]    [Pg.177]    [Pg.185]    [Pg.192]    [Pg.227]    [Pg.345]    [Pg.361]    [Pg.94]    [Pg.270]    [Pg.551]    [Pg.174]    [Pg.118]    [Pg.50]    [Pg.139]    [Pg.84]    [Pg.49]   
See also in sourсe #XX -- [ Pg.2 , Pg.268 ]




SEARCH



Spreads source oils

© 2024 chempedia.info