Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Sampler switch

Sampler Injection volume variation Sample aspirating needle and loading/transfer port Contact between sampler switching valve and sample... [Pg.103]

Consider the sampler switch shown in Fig. 7.88a where f(t) is the continuous input. Suppose the switch closes every ST units of time (the sampling time) and opens again instantaneously. Then the output of the sampler (generally written as f (0) will be a sequence of impulses each of magnitude equal to that of f(r) at that particular sampling time (Figs 7.886 and 7.88c). [In practice this will not be so as a... [Pg.672]

Fig. 7.88. Sampling a continous signal (a) sampler switch (b) continuous signal—input to sampler (c) sampled signal—output from sampler... Fig. 7.88. Sampling a continous signal (a) sampler switch (b) continuous signal—input to sampler (c) sampled signal—output from sampler...
Consider a sampler switch followed by an element for which we know the continuous transfer function G(s) as in Figure 7.89. [Pg.675]

The particles most likely to cause adverse health effects are the fine particulates, in particular, particles smaller than 10 p and 2.5 mm in aerodynamic diameter, respectively. They are sampled using (a) a high-volume sampler with a size-selective inlet using a quartz filter or (b) a dichotomous sampler that operates at a slower flow rate, separating on a Teflon filter particles smaller than 2.5 mm and sizes between 2.5 mm and 10 mm. No generally accepted conversion method exists between TSP and PM,o, which may constitute between 40% and 70% of TSP. In 1987, the USEPA switched its air quality standards from TSP to PMk,. PM,q standards have also been adopted in, for example, Brazil, Japan, and the Philippines. In light of the emerging evidence on the health impacts of fine particulates, the USEPA has proposed that U.S. ambient standards for airborne particulates be defined in terms of fine particulate matter. [Pg.16]

Perkin-Elmer Model lSS-100 automatic sampler Kratos ABI Spectroflow Model 783 UV/VIS detector, VlCl EQ60 LC switching valve Supelcosil LC-CN, 33 x 4.6-mm i.d., 5-pm particle size... [Pg.1305]

As well as the methods discussed above headspace samplers, pyrolyzers, thermal desorbers, and column switching devices could be considered as specialized sample inlet systems. These are treated separately in Chapter 8. [Pg.643]

For HPLC, the injector is a valve. In the charge position, a 50- jL syringe is used to fill the sample loop that holds a specific volume of sample solution. The valve is switched to the run position, and the eluent carries the sample out of the sample loop and into the column (auto samplers are also available for HPLC). A recording of the detector output is automatically started at the time of injection and produces a chromatogram of the separated components. [Pg.281]

Starting the instrument after the required volume has been set, the air sampler is started by pressing the [ON/OEF] key or using the remote control. Starting the instrument with the remote control is possible only after the instrument has been switched on. [Pg.765]

Figure 4 shows a simple version of a feedback flow control sampler system. An orifice and pressure switch are used to monitor air flow. [Pg.492]

When the phase increment Ml L is greater than one, and the original signal is being reduced in sample rate, classical sample rate techniques require that the cutoff frequency of the prototype filter change to. One approach is to time- scale the polyphase subfilters, but this increases the computation rate, which is undesirable in a typical VLSI implementation. Another approach is to switch filter tables. In practice, may sampler implementations only have one filter table, and pitch shifting up is restricted to be less than one octave. With this restriction, it is usually OK to use only one filter table with its cutoff equal to f. The artifacts resulting from this compromise are usually acceptable. [Pg.180]

Sometimes, harmful substances are released into a water body for short periods, but only at irregular intervals. In such situations event-controlled sampling is useful (Figure 1.3).7 When some selected parameter (e.g., conductivity and temperature) reaches its threshold value, then an automatic pump sampler is switched on and a sample is collected. [Pg.4]

An adjustable glass column of 9mm ID, packed with 5g of Fractosil 200 was connected to a Spectra Physics pump connected to solvent reservoirs with a six way rotary switch. The sample was injected onto the column with a rotary loop sampler. The effluent from the column was collected with an Isco fraction collector set so that 10 ml of eluate was collected for each... [Pg.215]

In ICP-MS (Fig. 112) the ions formed in the ICP are extracted with the aid of a conical water-cooled sampler into the first vacuum stage where a pressure of a few mbar is maintained. A supersonic beam is formed and a number of collision processes take place as well as an adiabatic expansion. A fraction is sampled from this beam through the conical skimmer placed a few cm away from the sampler. Behind the skimmer, ion lenses focus the ion beam now entering a vacuum of 10-5. This was originally done with the aid of oil diffusion pumps or cryopumps, respectively, but very quickly all manufacturers switched to turbomolecular pumps backed by roughing pumps. [Pg.255]

The most widely used sampler in the United States networks today is comprised of two buckets, and a rain-activated switch to operate a movable cover. During non-precipitating periods, the cover remains tightly sealed on one bucket. Precipitation falling on the sensitive switch completes an electrical circuit activating a motor which lifts the cover from one side and places it on the opposite bucket. The sampler thus provides a dry sample as well as a wet sample. This is the standard instrument used throughout the NADP/NTN network. [Pg.26]

Consider a line carrying a measurement signal y, which varies continuously in time as shown in Figure 27.1a. The line is interrupted by a switch, called sampler, which closes every T seconds and remains closed for an infinitesimally short period of time (theoretically, a time point). The x s of Figure 27.1b show the value of the signal at the other end of the line, when T = 1 second. We notice that the resulting signal has values only when the time is a multiple of T, that is, at time points... [Pg.652]

For stations or wells that have flow, no flow operations, the sampler should be turned off with a flow switch when the flow is off. Sampling should be stopped when there is no flow in every case. After a number of years and many test locations, it is recommended that in locations where gas has a heating value of 1025 BTU or above should be considered as prime locations for the use of a continuous sampler. [Pg.74]


See other pages where Sampler switch is mentioned: [Pg.100]    [Pg.313]    [Pg.330]    [Pg.428]    [Pg.55]    [Pg.216]    [Pg.2]    [Pg.140]    [Pg.145]    [Pg.237]    [Pg.496]    [Pg.674]    [Pg.118]    [Pg.65]    [Pg.100]    [Pg.696]    [Pg.25]    [Pg.36]    [Pg.140]    [Pg.118]    [Pg.728]    [Pg.190]    [Pg.254]    [Pg.115]    [Pg.409]    [Pg.289]    [Pg.654]    [Pg.216]    [Pg.869]    [Pg.4336]    [Pg.21]    [Pg.988]    [Pg.1138]   
See also in sourсe #XX -- [ Pg.673 ]




SEARCH



© 2024 chempedia.info