Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Preamplifiers operational

Limitations in the digitizer s dynamic range can be overcome by using multiple transient recorders operating at diflerent sensitivities, or by adding logarithmic preamplifiers in the detection system. From the preceding discussion it appears, however, that quantitative analysis is not the primary area of application of LIMS. Semiquantitative and qualitative applications of LIMS have been developed and are discussed in the remainder of this article. [Pg.590]

The preamplifier amplifies the voltage pulse. Further amplification is obtained by sending the signal through an amplifier circuit (typically about 10 volts maximum). The pulse size is then determined by a single channel analyzer. Figure 10 shows the operation of a single channel analyzer. [Pg.48]

Ge(Li) Detector Characteristics. Resolution measurements for the 18-cm.8 Ge(Li) detector were made with the anticoincidence shield in the inoperative mode, with a normal operating bias of 1700 volts, and with a preamplifier designed in our Laboratory (3, 4), and operated in conjunction with a Tennelec TC-200 linear amplifier. Resolution at 1.33 M.e.v. was 2.62 k.e.v., FWHM (Figure 4). The electronic pulser resolution for the amplifier system at a slightly lower energy was 1.86 k.e.v., the total capacitance of the detector was 28 pF, the noise slope was 0.035 k.e.v./pF, and the leakage current at 1700 volts was 0.5 X 10"9 amp. [Pg.216]

The electronic instrumentation necessary for the operation of the proportional counter is shown in Figure 18.6. Pulses from the detector pass through a preamplifier and amplifier, where they are shaped and amplified. Emerging from the amplifier, the pulses go to a discriminator. The discriminator is set so as not to trip on noise pulses but rather to trip on radiation pulses of any larger size. The number of discriminator pulses produced is recorded by the scaler. [Pg.546]

The most important factors in determining the s/n ratio or sensitivity of an NMR experiment in a cryogenic NMR probe, insofar as the hardware itself is concerned, are the temperature of the coil, 7/ the resistance of the coil, Rc the temperature of the sample, Ts and the resistance added to the coil by the presence of the sample or the sample resistance, Rs. The temperature of the rf coil in a cryogenic NMR probe is typically in the range of 15-30K commercial examples of the Varian Cold-probe operate at 25 K. The preamplifier noise temperature is generally in the range of 10-15 K and the coil resistance is small compared to a conventional room temperature NMR probe. The first two terms in the expression below provide the basis for the vast improvement in the performance of a cryogenic relative to a conventional NMR probe. The sample temperature, Ts, and sample resistance, Rs,... [Pg.21]

The operation principle of pulse channel for measuring the coimt rate is based on the neutron registration by the detector of ionization fission chamber or neutron counter, pulse transmission from the detector via the communication lines to the preamplifier inlet and their following amplification, formation and processing by means of the auxiliary electronic equipment. [Pg.215]

The bridge should be operated with as low a power level and for as short a time as possible to prevent polarization effects.f If necessary, a preamplifier can be used to increase the imbalance signal before it is applied to the vertical input terminals of the oscilloscope. Be careful that the output leads of the preamplifier are properly connected to the input terminals of the scope, one of which is usually grounded. In order to optimize the accuracy of the measurements, choose Rj values whenever possible so that the variable terminal D is near the center of the slide wire, i.e., the reading Xis near 500 and A3/A4 — 1). [Pg.240]

In general, the detectors are combined with a preamplifier, an amplifier and a multichannel analyser, in which the pulses are sorted according to their pulse heights. Frequently, the multichannel analyser is operated by a computer and a program for peak search, peak net area calculation, energy calibration and radionuclide identification. [Pg.113]

X-ray spectrometry is generally carried out with Si(Li) detectors. The set-up is similar to that applied to y-ray spectrometry with i-Ge or Ge(Li) detectors cooling of the detector in a cryostat, operation in combination with a preamplifier, an amplifier and a multichannel analyser. The energy resolution is very good, as already mentioned in section 7.6, and makes it possible to distinguish the characteristic X rays of neighbouring elements. Some X-ray emitters that may be used for calibration purposes are listed in Table 7.6. [Pg.115]

A preamplifier is necessary. This is normally an integral part of the detector unit and it is located very near the detector in order to take advantage of the cooling which is necessary for the operation of the detector and which helps the preamplifier to operate with low noise. [Pg.170]

A biased high voltage power supply is required to supply high voltage to the detector through the preamplifier. A power supply of 1500-5000 V is adequate for the operation of germanium detectors. [Pg.170]

E is the energy of characteristic X-ray line and Fisa constant called the Fano factor, which has a value of 0.12 for Si(Li). electronic noise factor, plays an important role in the resolution. Reduction of the electronic noise will improve the resolution of the EDS detector. Thus, the Si(Li) diode and the preamplifier are mounted in a cylindrical column (the cryostat) so that they can operate at the temperature of liquid nitrogen (-196°C) in order to reduce the electrical noise and increase the signal-to-noise ratio. [Pg.184]

Fig. 7-20 Si(Li) counter and FET preamplifier, very schematic. Both are in a cooled evacuated space, and x-rays enter through a beryllium window. The counter is operated at about 1000 volts. 0 = electron, = hole. Fig. 7-20 Si(Li) counter and FET preamplifier, very schematic. Both are in a cooled evacuated space, and x-rays enter through a beryllium window. The counter is operated at about 1000 volts. 0 = electron, = hole.
ISPD. Very simple operation with minimum variable parameters. MCP photocathode voltage is adjustable from approximately 160-240 V, but is typically fixed. Phosphor voltage is adjustable from 5-6 KV, and is also typically fixed. MCP voltage can be varied to adjust the gain. As a rule of thumb, a 50 V increase in voltage will double the gain. With respect to SPDs, the phases and preamplifier gain need usually be adjusted only once. Scan time is usually fixed. [Pg.10]

SPD. The SPD is rather simple to operate. Adjustments are mainly those of phases and odd-even preamplifier gain setting. The array does not provide any amplification gain (as SIT or MCP/SPD devices do). Scan rate can also be adjusted from a few KHz to a few MHz. [Pg.10]

The move to all-digital instrument components and digital oversampling (as used in CD players, for example) added further benefits. Commercial systems now offer cryogenic probes, which improve signal-to-noise (S/N) by reducing the operating temperature of the coil and the preamplifier. The increase in the S/N ratio by a factor of 4, as compared to conventional probes, leads to a possible reduction in experiment time of 16-fold or a reduction in required sample concentration by a factor of... [Pg.160]

The output from a voltage sensitive or resistor feedback preamplifier is a tail pulse with a rather long decay time. Hence, some pulse pile-up is unavoidable, except at very low count rates. Pile-up will cause the average level of this signal to increase with pulse rate, which may approach the limit of linear operation of the preamplifier. [Pg.224]

Currently manufaetured hybrid PMTs have gains of the order of 10,000. Therefore eleetronie noise from the matching resistor and from the preamplifier impairs the time resolution of single photon detection. The relatively large distance between the photoeathode and diode chip may also cause transit-time spread. In praetieal applieations, the extremely high operating voltage also causes problems. Currently hybrid PMTs are not routinely used for TCSPC experiments. [Pg.221]


See other pages where Preamplifiers operational is mentioned: [Pg.189]    [Pg.189]    [Pg.1423]    [Pg.70]    [Pg.68]    [Pg.72]    [Pg.75]    [Pg.175]    [Pg.578]    [Pg.216]    [Pg.62]    [Pg.53]    [Pg.175]    [Pg.428]    [Pg.12]    [Pg.26]    [Pg.125]    [Pg.129]    [Pg.598]    [Pg.256]    [Pg.21]    [Pg.139]    [Pg.1423]    [Pg.292]    [Pg.181]    [Pg.201]    [Pg.223]    [Pg.224]    [Pg.140]    [Pg.185]    [Pg.233]    [Pg.238]    [Pg.243]    [Pg.245]   
See also in sourсe #XX -- [ Pg.164 ]




SEARCH



Preamplifiers

© 2024 chempedia.info