Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Potentiometric conditions

This value does not express the actual result since side and/or parallel reactions (e.g., H+ or 02 reduction) are not considered, but it does demonstrate the completeness of the cementation process and the effectiveness of this liquid-liquid extraction. During this extraction no external current flows through the phase boundary Hg (amalgam)/solution thereby establishing a potentiometric condition. The question of the potential difference at the phase boundary can be answered by constructing the experimentally accessible current-voltage curves for the reactions ... [Pg.230]

This figure demonstrates that also under potentiometric conditions (- no external current flow) electrochemical net reactions occur. The EMF of the zinc-amalgam in a given Zn2 -ion solution depends on the current-voltage characteristic of other ions (in this example, Cu2 and Pb2 are interfering ions with respect to the Zn2 equilibrium potential) at the amalgam electrode. EMF drifts are thus explainable. [Pg.231]

Zirconia cells similar to the ones employed in the present study, have been used i) by Mason et al (18) to electrochemically remove oxygen from Pt and Au catalysts used for NO decomposition. It was shown that electrochemical oxygen pumping causes a dramatic increase in the rate of NO decomposition (18,19), ii) by Farr and Vayenas to electrochemically oxidize ammonia and cogenerate NO and electrical energy (20,21), iii) by Vayenas et al (11,12,22,23) to study the mechanism of several metal catalyzed oxidations under open circuit (potentiometric) conditions. [Pg.184]

Many - gas sensors based on - solid electrolytes operate under potentiometric conditions [iii]. The sensors for oxygen use oxide -> conductors, such as ZrC>2 -based ceramic, those for halogens use halide conductors (e.g., KAg s), while -> hydrogen sensors use protonic conductors. There are sensors for C02, N02, NH3, S03) H2S, HCN, HF, etc. (see -> lambda probe). [Pg.544]

Provided that the redox equilibrium (7.5) is electrochemically reversible under potentiometric conditions at the inert electrode, the electrode will assume a potential given by the following equation (Nemst equation) ... [Pg.197]

In potentiometry the potential of an electrochemical cell is measured under static conditions. Because no current, or only a negligible current, flows while measuring a solution s potential, its composition remains unchanged. For this reason, potentiometry is a useful quantitative method. The first quantitative potentiometric applications appeared soon after the formulation, in 1889, of the Nernst equation relating an electrochemical cell s potential to the concentration of electroactive species in the cell. ... [Pg.465]

Most potentiometric electrodes are selective for only the free, uncomplexed analyte and do not respond to complexed forms of the analyte. Solution conditions, therefore, must be carefully controlled if the purpose of the analysis is to determine the analyte s total concentration. On the other hand, this selectivity provides a significant advantage over other quantitative methods of analysis when it is necessary to determine the concentration of free ions. For example, calcium is present in urine both as free Ca + ions and as protein-bound Ca + ions. If a urine sample is analyzed by atomic absorption spectroscopy, the signal is proportional to the total concentration of Ca +, since both free and bound calcium are atomized. Analysis with a Ca + ISE, however, gives a signal that is a function of only free Ca + ions since the protein-bound ions cannot interact with the electrode s membrane. [Pg.489]

Precision The precision of a potentiometric measurement is limited by variations in temperature and the sensitivity of the potentiometer. Under most conditions, and with simple, general-purpose potentiometers, the potential can be measured with a repeatability of +0.1 mV. From Table 11.7 this result corresponds to an uncertainty of +0.4% for monovalent analytes, and +0.8% for divalent analytes. The reproducibility of potentiometric measurements is about a factor of 10 poorer. [Pg.495]

The holistic thermodynamic approach based on material (charge, concentration and electron) balances is a firm and valuable tool for a choice of the best a priori conditions of chemical analyses performed in electrolytic systems. Such an approach has been already presented in a series of papers issued in recent years, see [1-4] and references cited therein. In this communication, the approach will be exemplified with electrolytic systems, with special emphasis put on the complex systems where all particular types (acid-base, redox, complexation and precipitation) of chemical equilibria occur in parallel and/or sequentially. All attainable physicochemical knowledge can be involved in calculations and none simplifying assumptions are needed. All analytical prescriptions can be followed. The approach enables all possible (from thermodynamic viewpoint) reactions to be included and all effects resulting from activation barrier(s) and incomplete set of equilibrium data presumed can be tested. The problems involved are presented on some examples of analytical systems considered lately, concerning potentiometric titrations in complex titrand + titrant systems. All calculations were done with use of iterative computer programs MATLAB and DELPHI. [Pg.28]

The sufficient selectivity to a principal component is the most important condition determining the possibility of ion-selective electrodes (ISEs) practical appliances. In this work, the relationship between the potentiometric selectivity of alkylammonium-selective electrodes and factors such as the nature of plasticizer, ion-exchanger and substitution degree of cationic nitrogen atoms of the principal and foreign ions, is discussed. [Pg.314]

It may happen that AH is not available for the buffer substance used in the kinetic studies moreover the thermodynamic quantity A//° is not precisely the correct quantity to use in Eq. (6-37) because it does not apply to the experimental solvent composition. Then the experimentalist can determine AH. The most direct method is to measure AH calorimetrically however, few laboratories Eire equipped for this measurement. An alternative approach is to measure K, under the kinetic conditions of temperature and solvent this can be done potentiometrically or by potentiometry combined with spectrophotometry. Then, from the slope of the plot of log K a against l/T, AH is calculated. Although this value is not thermodynamically defined (since it is based on the assumption that AH is temperature independent), it will be valid for the present purpose over the temperature range studied. [Pg.258]

Standard potentials are determined with full consideration of activity effects, and are really limiting values. They are rarely, if ever, observed directly in a potentiometric measurement. In practice, measured potentials determined under defined concentration conditions (formal potentials) are very useful for predicting the possibilities of redox processes. Further details are given in Section 10.90. [Pg.65]

Standard potentials Ee are evaluated with full regard to activity effects and with all ions present in simple form they are really limiting or ideal values and are rarely observed in a potentiometric measurement. In practice, the solutions may be quite concentrated and frequently contain other electrolytes under these conditions the activities of the pertinent species are much smaller than the concentrations, and consequently the use of the latter may lead to unreliable conclusions. Also, the actual active species present (see example below) may differ from those to which the ideal standard potentials apply. For these reasons formal potentials have been proposed to supplement standard potentials. The formal potential is the potential observed experimentally in a solution containing one mole each of the oxidised and reduced substances together with other specified substances at specified concentrations. It is found that formal potentials vary appreciably, for example, with the nature and concentration of the acid that is present. The formal potential incorporates in one value the effects resulting from variation of activity coefficients with ionic strength, acid-base dissociation, complexation, liquid-junction potentials, etc., and thus has a real practical value. Formal potentials do not have the theoretical significance of standard potentials, but they are observed values in actual potentiometric measurements. In dilute solutions they usually obey the Nernst equation fairly closely in the form ... [Pg.363]

In electro-gravimetric analysis the element to be determined is deposited electroly tically upon a suitable electrode. Filtration is not required, and provided the experimental conditions are carefully controlled, the co-deposition of two metals can often be avoided. Although this procedure has to a large extent been superseded by potentiometric methods based upon the use of ion-selective electrodes (see Chapter 15), the method, when applicable has many advantages. The theory of the process is briefly discussed below in order to understand how and when it may be applied for a more detailed treatment see Refs 1-9. [Pg.503]

The relative change of conductance of the solution during the reaction and upon the addition of an excess of reagent largely determines the accuracy of the titration under optimum conditions this is about 0.5 per cent. Large amounts of foreign electrolytes, which do not take part in the reaction, must be absent, since these have a considerable effect upon the accuracy. In consequence, the conductimetric method has much more limited application than visual, potentiometric, or amperometric procedures. [Pg.523]

A sample of polyester (ca. 1 g, exactly weighed) is dissolved in 20 mL toluene-ethanol mixture (1/1 vol.) and titrated by a solution of KOH in ethanol (0.05 mol/L) using a potentiometric titrator. A blank titration must be performed under the same conditions. Hardly soluble polyesters (e.g., PET) must be dissolved in an o-cresol-chloroform mixture or in hot benzyl alcohol.417 The result (acid content) is normally expressed in mmol COOH/g polyester but may also be given as the acid number, defined as the number of milligrams of KOH required to neutralize 1 g of polyester. [Acid number = (number of mmol COOH/g polyester) x 56.106.]... [Pg.94]

Alternatively, the 3- and 4-hydroxy sulfonates may be converted to the corresponding sultones by treatment with a strong mineral acid. An ether extract concentrates the organic components, sultones, and alkenesulfonic acid, which can be weighed and titrated potentiometrically with sodium hydroxide. 2-Hydroxyalkanesulfonate will not dehydrate to the sultone under these conditions and is not measured. [Pg.435]

The potentiometric titration was carried out in order to determine the functional groups present in the biomass surface. During the titration experiments, the C02-free condition was always maintained to avoid the influence of inorganic carbon on the solution pH. Detailed potentiometric titration procedure and estimation method of functional groups are available in the previous reports [4,6]. [Pg.162]

Extensive data are given in the Uterature for the potentiometric titration of polymer acids which may be used to study the behaviour of polyelectrolyte systems under different conditions. For poly(a-D) galacturonic acid there are few data of this kind, especially in connection with the occurrence of a conformational transition induced by pH variations, or with the effect brought about by the addition or the exchange of counterions. Since for a polyacid not exhibiting a conformational transition in the course of titration, pK K denoting the apparent dissociation constant) increases monotonously with degree... [Pg.609]

Potentiometric methods are based on the measurement of the potential of an electrochemical cell consisting of two electrodes immersed in a solution. Since the cell potential is measured under the condition of zero cmrent, usually with a pH/mV meter, potentiometry is an equilibrium method. One electrode, the indicator electrode, is chosen to respond to a particular species in solution whose activity or concentration is to be measured. The other electrode is a reference electrode whose half-cell potential is invariant. [Pg.3]

However, under most conditions the activity coefficients cannot be neglected, certainly for a single redox couple where the ox/red concentration ratio cannot be simply calculated from the true standard potential and the potential directly observed. In order to overcome this difficulty the concept of the formal potential was introduced, which represents a formal standard potential E ° measured in an actual potentiometric calibration and obeying the Nernst equation, E = E ° + (0.05916/n) log ([ox]/[red]) at 25° C, E"0 must meet the conditions under which the analytical measurements have to be made. Sometimes the formal potential values are decisive for the direction of the reaction between two redox couples even when the E° values do not differ markedly10. [Pg.48]

Fig. 2.21. Potentiometric complex-formation titration (under conditions favouring the overall reaction Cu2+ + Y4- -> CuY2-). [Pg.106]


See other pages where Potentiometric conditions is mentioned: [Pg.230]    [Pg.230]    [Pg.373]    [Pg.80]    [Pg.304]    [Pg.52]    [Pg.136]    [Pg.236]    [Pg.610]    [Pg.876]    [Pg.542]    [Pg.196]    [Pg.117]    [Pg.163]    [Pg.336]    [Pg.117]    [Pg.3]    [Pg.348]    [Pg.744]    [Pg.154]    [Pg.668]    [Pg.44]    [Pg.200]    [Pg.213]    [Pg.164]   
See also in sourсe #XX -- [ Pg.110 ]




SEARCH



Potentiometric

© 2024 chempedia.info