Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Potentials, potentiometer

There are two procedures for doing this. The first makes use of a metal probe coated with an emitter such as polonium or Am (around 1 mCi) and placed above the surface. The resulting air ionization makes the gap between the probe and the liquid sufficiently conducting that the potential difference can be measured by means of a high-impedance dc voltmeter that serves as a null indicator in a standard potentiometer circuit. A submerged reference electrode may be a silver-silver chloride electrode. One generally compares the potential of the film-covered surface with that of the film-free one [83, 84]. [Pg.116]

Potentiometers Measuring the potential of an electrochemical cell under conditions of zero current is accomplished using a potentiometer. A schematic diagram of a manual potentiometer is shown in Figure 11.2. The current in the upper half of the circuit is... [Pg.464]

The potential of the working electrode, which changes as the composition of the electrochemical cell changes, is monitored by including a reference electrode and a high-impedance potentiometer. [Pg.465]

Potentiometric measurements are made using a potentiometer to determine the difference in potential between a working or, indicator, electrode and a counter electrode (see Figure 11.2). Since no significant current flows in potentiometry, the role of the counter electrode is reduced to that of supplying a reference potential thus, the counter electrode is usually called the reference electrode. In this section we introduce the conventions used in describing potentiometric electrochemical cells and the relationship between the measured potential and concentration. [Pg.466]

Precision The precision of a potentiometric measurement is limited by variations in temperature and the sensitivity of the potentiometer. Under most conditions, and with simple, general-purpose potentiometers, the potential can be measured with a repeatability of +0.1 mV. From Table 11.7 this result corresponds to an uncertainty of +0.4% for monovalent analytes, and +0.8% for divalent analytes. The reproducibility of potentiometric measurements is about a factor of 10 poorer. [Pg.495]

Where stray currents are involved, several measurements have to be taken that are continually changing with time, simultaneously with each other. A double recorder is most suitable for this. Linear recorders with direct indication of the measurements cannot be used for potential measurements because the torque of the mechanism is too small to overcome the friction of the pen on the paper. Amplified recorders or potentiometer recorders are used to record potentials. In amplified recorders, as in amplified voltmeters, the measured signal is converted into a load-independent impressed current and transmitted to the measuring mechanism, which consists of a torque motor with a preamplifier. The amplifier results in an... [Pg.100]

Figure 8-5 shows the main circuit diagram of a potential control rectifier provided with magnetic amplifiers (transducers). The chosen potential is set at the nominal value with a potentiometer. The actual potential is compared with this value, which corresponds to the voltage between a reference electrode and the protected object. [Pg.234]

An instrument mechanic was asked to test the trip on A furnace. He put the controller on manual and then went behind the panel. His next step was to take the cover off the back of the controller, disconnect one of the leads, applj a gradually increasing potential from a potentiometer, and note the reading at which the trip would operate if it was on auto control. [Pg.83]

This method involves very simple and inexpensive equipment that could be set up m any laboratory [9, 10] The equipment consists of a 250-mL beaker (used as an external half-cell), two platinum foil electrodes, a glass tube with asbestos fiber sealed m the bottom (used as an internal half-cell), a microburet, a stirrer, and a portable potentiometer The asbestos fiber may be substituted with a membrane This method has been used to determine the fluoride ion concentration in many binary and complex fluondes and has been applied to unbuffered solutions from Willard-Winter distillation, to lon-exchange eluant, and to pyrohydrolysis distil lates obtained from oxygen-flask or tube combustions The solution concentrations range from 0 1 to 5 X 10 M This method is based on complexing by fluonde ions of one of the oxidation states of the redox couple, and the potential difference measured is that between the two half-cells Initially, each cell contains the same ratio of cerium(IV) and cerium(tll) ions... [Pg.1026]

Voltmeters and potentiometers The instruments described here are generally referred to as corrosion voltmeters. As mentioned previously, the current flowing through any potential-measurement circuit must be small to avoid errors due to polarisation. Moreover, if the current flow is too large, errors will be introduced owing to the voltage drop caused by the contact resistance between the reference electrode and the electrolyte. It is thus clear that the prime requirement of a potential measurement circuit is high resistance. [Pg.246]

To measure structure/electrolyte potentials with electrolyte resistivities in excess of 2 kQ cm, a high-resistance potentiometer unit as shown in Fig. 10.43 or a potentiometric voltmeter as illustrated in Fig. 10.44 may be used. [Pg.246]

For normal field work the potentiometric voltmeter is the more popular instrument, being usually of lighter construction and not requiring calibration against a standard cell. Where extremely small potentials (usually potential shifts) of the order of 1 mV are to be measured, however, the potentiometer is more suitable and accurate. [Pg.248]

The films were then soaked in water and removed from the plates. Portions were mounted in glass cells which were filled with potassium chloride solution two Ag/AgCl electrodes were inserted into the limbs of the cells and the unit was placed in a thermostat. The resistance of the films was determined, from time to time, by connecting the cells in series with a known resistance and applying a potential of 1 V to the combination the potential drop across the standard resistance was measured by means of a valve potentiometer. [Pg.599]

Potential measurements—a reference electrode and a potentiometer or electrometer which require only a small current to give a measurement of e.m.f. and thus minimise polarisation of the electrodes. [Pg.1005]

The preferred potential-measuring instruments are potentiometers or electrometers, either of which permit measurements to be made without flow of sufficient current to polarise the electrodes during the determinations. It is also possible to use millivoltmeters if the internal resistance of the instrument is high enough to avoid any appreciable flow of current. [Pg.1019]

If it is desired to use the biamperometric method for detecting the end point, then the calomel electrode and also the silver rod (if used) must be removed and replaced by two platinum plates 1.25 cm x 1.25 cm. The potentiometer (or pH meter) used to measure the e.m.f. must also be removed, and one of the indicator electrodes is then joined to a sensitive galvanometer fitted with a variable shunt. The indicator circuit is completed through a potential divider... [Pg.541]

If the values of the a.c. current are plotted against the potential applied by the potentiometer, a series of peaks are obtained as illustrated in Fig. 16.11(a) the normal d.c. polarogram of the same solution is also shown (b). [Pg.613]

A simple apparatus suitable for this procedure is shown in Fig. 16.17. B is a 3 volt torch battery or 2 volt accumulator, M is a micro-ammeter, R is a 500 ohm, 0.5 watt potentiometer, and E,E are platinum electrodes. The potentiometer is set so that there is a potential drop of about 80-100 millivolts across the electrodes. [Pg.635]

Galvani potentials between two conductors of different types cannot be measured by any means. Methods in which the force acting on a test charge is measured cannot actually be used here, since any values that could be measured would be distorted by the chemical forces. The same holds true for determinations of the work of transfer. At least one more interface is formed when a measuring device such as a voltmeter or potentiometer is connected, and the Galvani potential of that interface will be... [Pg.25]


See other pages where Potentials, potentiometer is mentioned: [Pg.306]    [Pg.116]    [Pg.198]    [Pg.212]    [Pg.366]    [Pg.942]    [Pg.464]    [Pg.465]    [Pg.494]    [Pg.505]    [Pg.776]    [Pg.472]    [Pg.82]    [Pg.107]    [Pg.676]    [Pg.676]    [Pg.248]    [Pg.1171]    [Pg.1220]    [Pg.1239]    [Pg.583]    [Pg.593]    [Pg.659]    [Pg.362]    [Pg.1037]    [Pg.35]    [Pg.17]    [Pg.23]    [Pg.195]    [Pg.210]   
See also in sourсe #XX -- [ Pg.288 ]




SEARCH



© 2024 chempedia.info