Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Positive electrode installation

Place the taped rod with the Teflon insert into the fitting, put on the ferrules and nut, and then tighten. [Pg.143]

Insert and position the Teflon liner back into the nipple. The threaded rod will now protrude through the holes in the liner when it is positioned correctly. [Pg.143]

Put a nickel washer on the inside of each electrode rod, over the mesh. [Pg.145]

Screw the 8-32 nuts onto the rod over the nickel washers to fasten and tighten. [Pg.145]

Apply thread seal and Teflon tape to the NPT ends of the straight 34 NPT x % tubing Yor-Lok fittings, and screw and tighten them into the four holes in the top and bottom blind flanges. [Pg.146]


Tap both holes for 34" NPT thread. Do not tap too far. The fittings for the positive electrodes should only extend slightly past the inner surface of the pipe wall, so that they do not interfere with the liner that will be installed later. [Pg.129]

Figure 12-9 shows on and off potentials that were measured around the circumference of a flat-bottomed tank 100 m in diameter. These values, however, give no information on the tank/soil potentials at the center of the container or at points away from the edge of the tank. In new tank constructions, long-life reference electrodes are therefore installed in the center of the base where the most positive potentials are found [15]. [Pg.322]

The sample treatment is executed as follows The sample is located inside the chamber. The SFM scanner is retracted by about 8 mm from the sample. Then the chamber is closed by the lid. In order to anneal the sample successfully, we perform the treatment in a hermetically sealed reactor chamber. The sample is exposed to chloroform vapor while a voltage is applied across the electrodes on the glass substrate. The gas flow rate through the system is controlled by mass flow controllers and additionally checked by a flow meter installed at the outlet of the chamber. After a certain annealing time, the solvent is removed with a flow of pure argon while the voltage is still applied in order to quench the structure in the presence of the electric field. Subsequently the lid is opened, the voltage is switched off, and the SFM scanner is moved towards the sample. This procedure is repeated several times. The accomplishable position stability as a key performance feature of the quasi in situ SFM amounts to almost the same value as reported in [32] (low pressure plasma treatments). The increase in lateral drift is only about 23 nm per process step. The... [Pg.11]

After installation in concrete, it is almost impossible to check the potential accurately. Measurement of the potential of the embedded electrode with a reference electrode placed on the concrete surface may give highly variable results, depending on the condition of the concrete cover (moisture, carbonation depth and chloride content). A better practice is to install two reference electrodes in approximately the same position in the concrete and compare their mutual potential difference. [Pg.33]

The electrochemical and chemical stability of diamond makes it an ideal electrode material for electrochemical fluorination reactions. The installation of fluorine a to heteroatom-substituted positions can be anodically performed by hydrogen fluoride/triethylamine mixtures. The Fuchigami group studied several electrode materials for the fluorination of oxindole 20. In this transformation to 21 only a... [Pg.13]

As several installations are running since 2—3 years with good feedbacks from users, Adamant has gained a medium-/long-term field experience with BDD electrodes and DiaCell -based units. These positive references are found in ... [Pg.160]

Figure 8.1b portrays the process of filtering out the ions in solution. Inside the tank, cation and anion membranes are installed alternate to each other. Two electrodes are put on each side of the tank. By impressing electricity on these electrodes, the positive anode attracts negative ions and the negative cathode attracts positive ions. This impression of electricity is the reason why the respective ions replace their like... [Pg.388]

Figure 2.16 Electron capture detector(ECD) (a) and photo-ionization detector (PID) (b). The BCD must be installed in an well ventilated position owing to it containing a radioactive source. The PID contains a UV source from which the photons are emitted, having a pre-selected energy, using a filter which prevents undesired carrier gas ionization M + hv M+ - - e ). Examples of filters LiF at ll.SeV, MgFj at 9.6-10 eV, sapphire at 8.4 eV. On contact with the electrodes the molecules return to uncharged state, ionization being therefore reversible. The use of the make-up gas provides an optimal flow. Figure 2.16 Electron capture detector(ECD) (a) and photo-ionization detector (PID) (b). The BCD must be installed in an well ventilated position owing to it containing a radioactive source. The PID contains a UV source from which the photons are emitted, having a pre-selected energy, using a filter which prevents undesired carrier gas ionization M + hv M+ - - e ). Examples of filters LiF at ll.SeV, MgFj at 9.6-10 eV, sapphire at 8.4 eV. On contact with the electrodes the molecules return to uncharged state, ionization being therefore reversible. The use of the make-up gas provides an optimal flow.
By a proper treatment of the electrodes, one can obtain a texture with a uniform orientation of the smectic normal in one direction within the cell plane. Between the crossed polarizers such a cell will be black if a polarizer is installed parallel to the smectic normal. Upon application of the electric field, the antiferroelectric structure becomes distorted. At low voltages of any polarity, the electrooptic response is proportional to E the bottom part of the curves has symmetric parabolic form [35] shown in Fig. 13.24b. Above the AF-F transition, the director acquires one of the two symmetric angular positions ( 9 on the conical surface) typical of the SmC phase. At these two extreme positions the transmission is maximum. With increasing temperature from T toTi the AF-F threshold decreases due to a decrease of the potential barrier separating structures with alternating and uniform tilt. It is natural because within the SmC A phase T1 is closer to the range of the SmC phase than T2 or T3. [Pg.421]

The technician on board relies entirely on the information transmitted by the diver, which is not always reliable and reproducible. Therefore, such measurements are carried out with the additional aid of a television camera so that the technician on board can record the position and measurement on videotape. The arrangement shown in Fig. 16-12A is advantageous because the reference electrode can be coupled with the TV camera. The state of the anodes, their possible passivation and material loss can be investigated at the same time as the potential measurements, and the marine growths can be removed if they threaten to smother the anodes. Platforms are given an annual visual examination. Potential measurements are also carried out on these occasions. Impressed current installations are also subject to continual monitoring so that defects can be detected at an early stage and measures to repair them instituted. [Pg.386]


See other pages where Positive electrode installation is mentioned: [Pg.143]    [Pg.194]    [Pg.143]    [Pg.194]    [Pg.45]    [Pg.155]    [Pg.170]    [Pg.107]    [Pg.47]    [Pg.149]    [Pg.45]    [Pg.136]    [Pg.867]    [Pg.63]    [Pg.86]    [Pg.257]    [Pg.47]    [Pg.256]    [Pg.292]    [Pg.479]    [Pg.340]    [Pg.145]    [Pg.395]    [Pg.74]    [Pg.90]    [Pg.117]    [Pg.12]    [Pg.592]    [Pg.24]    [Pg.232]    [Pg.53]    [Pg.217]    [Pg.297]    [Pg.390]    [Pg.1251]    [Pg.55]    [Pg.47]    [Pg.256]    [Pg.292]    [Pg.479]    [Pg.549]   


SEARCH



Electrode positive

Electrodes installation

Electrodes installed

© 2024 chempedia.info