Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Peak critical, separation

Figure 2.96 shows the splitting of the net peak under increasing of the dimensionless electrode kinetic parameter for a given film thickness. The potential separation between split peaks increases in proportion to the electrode kinetic parameter and the amplitude of the potential modulation. The dependence of the peak potential separation on the amplitude is separately illustrated in Fig. 2.97. The analysis of the splitting by varying the amplitude is particularly appeahng, since this instrumental parameter affects solely the split peak without altering the film thickness parameter. Table 2.7 lists the critical intervals of the film thickness and the electrode kinetic parameters attributed with the splitting. Figure 2.96 shows the splitting of the net peak under increasing of the dimensionless electrode kinetic parameter for a given film thickness. The potential separation between split peaks increases in proportion to the electrode kinetic parameter and the amplitude of the potential modulation. The dependence of the peak potential separation on the amplitude is separately illustrated in Fig. 2.97. The analysis of the splitting by varying the amplitude is particularly appeahng, since this instrumental parameter affects solely the split peak without altering the film thickness parameter. Table 2.7 lists the critical intervals of the film thickness and the electrode kinetic parameters attributed with the splitting.
Detailed aspects of analytical and preparative chromatography are discussed in two fundamental monographs [11, 12]. The identification of a chromatographic system providing adequate selectivity (a), that is, sufficient separation of the compound of interest from the closest eluting peak (critical pair), often is the most challenging task given the solubility - often lack thereof - of the feed, as discussed in Section 7.4.3. [Pg.223]

Due to the presence of the internal standard, it is critical to ensure that the analyte peak be separated from the internal standard peak. A minimum of baseline separation (resolution >1.5) of these two peaks is required to give reliable quantitation. In addition, to quantitate the responses of internal standard accurately, the internal standard should be baseline resolved from any significant related substances and should have a peak height or area similar to that of the standard peak. [Pg.14]

Critical separations in chromatography should be investigated at the appropriate level. Specificity can best be demonstrated by the resolution of two chromographic peaks that elute close to each other. In the potency assay, one of the peaks would be the analyte peak. Figure 2.4 illustrates the selectivity of a method to resolve known degradation peaks from the parent peak. Based on the... [Pg.21]

Critical separations in chromatography should be investigated at an appropriate level. For critical separations, selectivity can be demonstrated by the resolution of the two components that elute closest to each other. Peak purity tests using diode array or mass spectrometric detectors may be useful to show that the analyte chromatographic peak is not attributable to more than one component. [Pg.751]

A critical problem with adhesion layers arises from grain boundary diffusion. Deposited films tend to be polycrystalline and granular. The electrochemistry of the adhesion film is frequently much less desirable than the electrochemistry of the primary film. Moreover, minute contamination of the primary metal film surface by adhesion components can dramatically degrade the electron transfer properties (e.g., electrochemical reversibility, as evidenced by cyclic voltammetric peak potential separation) of the film [58], Thus it is essential that the adhesion layer is not exposed to solution. While the rate of diffusion of adhesion metals through the bulk of the primary layer is quite slow, grain boundary diffusion along the surfaces of grains is much faster. In many cases, the adhesion layer can seriously compromise the performance of the electrode. This is particularly a problem for chromium underlayers. Recently a codeposited Ti/W adhesion layer has been recommended as an alternative to chromium, with reportedly better adhesion and fewer interferences than Cr. A procedure was also described to recondition these electrodes to minimize interference by adhesion layer metals [58]. [Pg.346]

In physical terms, the formation of LiH in the ground state from its constituent atoms occurs by means of a transfer of an electron from the Li atom to H when the internuclear distance decreases below a critical separation R. This same concept underlies the harpoon mechanism which is used to explain the very large cross-sections for reaction which are observed for such processes as K -i- Br2 - KBr + Br. As the reactants approach, the covalent K + Br2 potential surface is intersected by an ionic K Br surface. Accordingly, an electron transfers from K to Br2. Subsequent production of KBr and Br is immediate. This model is also in accord with the observation in beam scattering experiments that the distribution of KBr product is strongly forward-peaked . [Pg.322]

Resolution. The resolution (Kj) is a measure of how well two peaks are separated. For reliable quantitation, well-separated peaks are essential. The separation of all peaks of interest is checked visually using a synthetic sample solution. The resolution factor (K) between the critical peak pair is calculated according to the formulas described in USP 24 and the European Pharmacopoeia (EP, 3rd ed.). [Pg.426]

The simplest and most common temperature program is the single linear ramp illustrated earher. More complex programs, with multiple ramps interrupted by isothermal intervals may be designed to optimize critical separations and minimize overall analysis time. Environmental and biological samples often contain very slowly eluting heavier background contaminants of no interest to the analysis, which must be cleared off the column before the next injection, lest they elute with and interfere with analyte peaks from... [Pg.767]

The worst resolved peak pair is BC hence the critical separation factor is 1.55. [Pg.36]

The second and third peaks will be the pair of peaks in the mixture that are eluted closest together and, thus, the most difficult to separate (usually given the term the critical pair as they define the severity of the separation). Finally, the fourth peak will be that which is eluted last from the mixture and will determine when the analysis is complete and establishes the total analysis time. The chromatographic system must be designed to separate the critical pair and, as this is the pair that is eluted closest together, all other peaks should also be resolved... [Pg.362]

Column design involves the application of a number of specific equations (most of which have been previously derived and/or discussed) to determine the column parameters and operating conditions that will provide the analytical specifications necessary to achieve a specific separation. The characteristics of the separation will be defined by the reduced chromatogram of the particular sample of interest. First, it is necessary to calculate the efficiency required to separate the critical pair of the reduced chromatogram of the sample. This requires a knowledge of the capacity ratio of the first eluted peak of the critical pair and their separation ratio. Employing the Purnell equation (chapter 6, equation (16)). [Pg.367]

Eor low molecular weight polymers the separation of the system peaks from the low molecular weight end of the polymer peak is very critical in obtaining accurate MWD and the percentage of low molecular weight materials in the polymer. The water/methanol mixture is a better solvent for PVP than water. PVP K-15 and K-30 should be better separated from the system peaks in the water/methanol mixture than in water because the difference in hydrodynamic volumes between PVP K-15 or K-30 and system peaks is larger in... [Pg.518]

For preparative or semipreparative-scale enantiomer separations, the enantiose-lectivity and column saturation capacity are the critical factors determining the throughput of pure enantiomer that can be achieved. The above-described MICSPs are stable, they can be reproducibly synthesized, and they exhibit high selectivities - all of which are attractive features for such applications. However, most MICSPs have only moderate saturation capacities, and isocratic elution leads to excessive peak tailing which precludes many preparative applications. Nevertheless, with the L-PA MICSP described above, mobile phases can be chosen leading to acceptable resolution, saturation capacities and relatively short elution times also in the isocratic mode (Fig. 6-6). [Pg.164]

The time that the last peak is eluted in the reduced chromatogram represents the total analysis time after which, the analysis can be terminated. The two peaks that are eluted closest together are the most difficult to separate and which, for obvious reasons, are called the "critical pair". [Pg.107]

The columns must be designed or chosen such that the critical pair are separated and, as a second priority, the last peak must be eluted in a reasonable time. The first peak in the chromatogram is not considered part of the reduced chromatogram and is included as the dead volume marker from which the capacity factors of each solute can be calculated, together with the separation ratio of the critical pair. [Pg.107]

The above considerations apply to samples where all the components are of interest and all need to be separated and quantitatively assessed. In practice, for many samples, only specific components of the mixture are important and only those need to be separated from the matrix and be analyzed. The components of the matrix need not be resolved and they are of no interest. It follows, that under these circumstances the critical pair will be comprised of the component of interest that has the closest neighbor and the neighbor itself. Such a situation usually greatly simplifies the separation problem but it should be noted that the last peak must still be eluted before the next analysis can be carried out and so the analysis time may not be significantly reduced. [Pg.108]

The column that is required for an analysis must have the efficiency necessary to resolve the critical pair. However, it is necessary to decide what constitutes resolution, before the column efficiency can be calculated. How narrow must the peaks be maintained relative to their separation to permit an accurate quantitative analysis In figure 7, five pairs of peaks are shown, the area of the smaller peak being half that of the larger peak. [Pg.108]

Having defined that the resolution required to separate the critical pair in a specific sample is 4a it is now possible to calculate the number of theoretical plates that are necessary to provide adequate quantitative accuracy. This can be easily carried out using the information provided by the Plate Theory in the chapter 2. Restating figure 10 from chapter 2 as figure 8, it is seen that the retention volume difference between the peaks (Av) is... [Pg.109]

Particle separation can be characterized by the separation factor, Rp, which is the ratio of eluant to particle elution volumes, or, by the difference in elution voliame, AV, between particle and eluant marker turbidity peaks. For polystyrene monodisperse standards, a linear relationship occ irs between the log of the particle diameter and AV, with a series of parallel lines resulting for different concentration of either salt or surfactant below its critical micelle concentration (IT>18,19) The separation factor has also been shown to be independent of eluant... [Pg.2]

The position of the peak (Figure 16) is of critical importance in distinguishing a composition based separation. The large axial dispersion in GPC 1 was attributed to the sample loadings being more than the 9 silica filled columns could handle. This had potentially serious consequences in terms of chromatogram sampling effects. [Pg.177]

A common criticism of CE is the poor consistency of migration times for peaks compared with HPLC or GC. The effects of matrix components and small differences in ionic strength and pH can have significant effects in the separation. The use of a... [Pg.783]

Since reproducibility of the flow system is critical to obtaining reproducibility, one approach has been to substitute lower-performance columns (50-to 100-p packings) operated at higher temperatures.1 Often, improvements in detection and data reduction can substitute for resolution. Chemometric principles are a way to sacrifice chromatographic efficiency but still obtain the desired chemical information. An example of how meaningful information can be derived indirectly from chromatographic separation is the use of system or vacancy peaks to monitor chemical reactions such as the titration of aniline and the hydrolysis of aspirin to salicylic acid.18... [Pg.92]


See other pages where Peak critical, separation is mentioned: [Pg.22]    [Pg.161]    [Pg.94]    [Pg.39]    [Pg.76]    [Pg.892]    [Pg.34]    [Pg.65]    [Pg.663]    [Pg.101]    [Pg.816]    [Pg.365]    [Pg.383]    [Pg.407]    [Pg.71]    [Pg.433]    [Pg.301]    [Pg.156]    [Pg.182]    [Pg.50]    [Pg.263]    [Pg.766]    [Pg.996]    [Pg.201]   
See also in sourсe #XX -- [ Pg.660 ]




SEARCH



Peak Separation

© 2024 chempedia.info