Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Mercury instrument

Fig. 11.11 Photographs showing non-mercury instruments for the determination of porosity and related data, (a) Capillary flow po-rometer for gas as well as liquid permeability and the testing of filter integrity (b) gas pycnometer (c) bulk/absolute density analyzer (d) BET sorptometer (courtesy PMI, Ithaca, NY, USA). Fig. 11.11 Photographs showing non-mercury instruments for the determination of porosity and related data, (a) Capillary flow po-rometer for gas as well as liquid permeability and the testing of filter integrity (b) gas pycnometer (c) bulk/absolute density analyzer (d) BET sorptometer (courtesy PMI, Ithaca, NY, USA).
Mercury instruments, electronics, chemicals, use curtailed because of toxicity... [Pg.289]

The commercial form of Cartesian manostat, model 7A, is depicted in Fig. II, 23, 7 it is normally charged with mercury except for very low pressures when di-w-butyl phthalate is employed. The manostat is highly sensitive in its action furthermore, once the pressure has been set in the instrument, the system may be shut down without disturbing the setting. [Pg.116]

The metal is widely used in laboratory work for making thermometers, barometers, diffusion pumps, and many other instruments. It is used in making mercury-vapor lamps and advertising... [Pg.81]

Miscellaneous Atomization Methods A few elements may be atomized by a chemical reaction that produces a volatile product. Elements such as As, Se, Sb, Bi, Ge, Sn, Te, and Pb form volatile hydrides when reacted with NaBH4 in acid. An inert gas carries the volatile hydrides to either a flame or to a heated quartz observation tube situated in the optical path. Mercury is determined by the cold-vapor method in which it is reduced to elemental mercury with SnCb- The volatile Hg is carried by an inert gas to an unheated observation tube situated in the instrument s optical path. [Pg.415]

Industrial and Control Instruments. Mercury is used in many industrial and medical instmments to measure or control reactions and equipment functions, including thermometers, manometers (flow meters), barometers and other pressure-sensing devices, gauges, valves, seals, and navigational devices (see Pressure measurements Process control Temperature measurement). Whereas mercury fever thermometers are being replaced by... [Pg.109]

As atomic fluorescence spectrometer a mercury analyzer Mercur , (Analytik-Jena, Germany) was used. In the amalgamation mode an increase of sensitivity by a factor of approximately 7-8 is obtained compared with direct introduction, resulting in a detection limit of 0,09 ng/1. This detection limit has been improved further by pre-concentration of larger volumes of samples and optimization of instrumental parameters. Detection limit 0,02 ng/1 was achieved, RSD = 1-6 %. [Pg.171]

Direct reading samplers include simple devices such as colorimetric indicating tubes in which a color change indicates the presence of the contaminant in air passed through the tube, or instruments which are more or less specific for a particular substance. In the latter category are carbon monoxide indicators, combustible gas indicators (explosimeters) and mercury vapor meters, as well as a number of other instruments. [Pg.266]

Liquid-in-glass thermometers measure the thermal expansion of a liquid, which is placed in a solid container, on a length scale. The mercury thermometer is one example of liquid thermometers. Alcohol is also used with this type of instrument. The temperature range is -80 to a-330 °C depending on the liquid. The quality, stability, and accuracy vary considerably. The advantages are a simple construction and low price. A disadvantage is that they are not compatible for connection to monitoring systems. [Pg.1137]

From the ventilation point of view, the fixed points -38.83 °C (triple-point of mercury), 0.010 °C (triple-point of water), 29.76 °C (melting point of gallium), and 156.60 °C (freezing point of indium) are of relevance. The triple-point of water is relatively simple to achieve and maintain with a triple-point apparatus. Some freezing point cells are covered in standards. In practical temperature calibration of measuring instruments, the lTS-90 fixed points are not used directly. [Pg.1140]

U-Tube manometer An instrument used for measuring pressure differences in a fluid or a gas by means of a LJ-shaped tube containing a fluid such as mercury or oil. [Pg.1484]

The Nernst equation shows that the glass electrode potential for a given pH value will be dependent upon the temperature of the solution. A pH meter, therefore, includes a biasing control so that the scale of the meter can be adjusted to correspond to the temperature of the solution under test. This may take the form of a manual control, calibrated in 0 C, and which is set to the temperature of the solution as determined with an ordinary mercury thermometer. In some instruments, arrangements are made for automatic temperature compensation by inserting a temperature probe (a resistance thermometer) into the solution, and the output from this is fed into the pH meter circuit. [Pg.566]

Now commence the voltage sweep using a scan rate of 5 mV per second, or with a manual polarograph, increase the voltage in steps of 0.05 V. The recorder plot will take the form shown in Fig. 16.4 if a manual instrument is used, then since the current oscillates as mercury drops grow and then fall away, the plot will have a saw-tooth appearance, and for measurement purposes a smooth curve must be drawn through the midpoint of the peaks of the plot. [Pg.617]

Instruments for the measurement of fluorescence are known as fluorimeters or spectrofluorimeters. The essential parts of a simple fluorimeter are shown in Fig. 18.1. The light from a mercury-vapour lamp (or other source of ultraviolet light) is passed through a condensing lens, a primary filter (to permit the light band required for excitation to pass), a sample container, a secondary filter (selected to absorb the primary radiant energy but transmit the fluorescent... [Pg.733]

Before each wavelength scan, the grating locates itself by finding the 365 nm triplet of a built-in mercury light source, and then moves sequentially to the wavelengths for the elements to be determined. As an option the instrument... [Pg.777]

The first successful application of the continuous wave (CW) He-Ne gas laser as a Raman excitation source by Kogelnik and Porto (14) was reported in 1963. Since that time, significant improvements in instrumentation have been continually achieved which have circumvented a great number of problems encountered with mercury lamp sources. The renaissance of Raman spectroscopy has also been due to improvements in the design of monochromators and photoelectric recording systems. [Pg.306]

The pressure of the atmosphere can be measured with a barometer, an instrument invented in the seventeenth century by Evangelista Torricelli, a student of Galileo. Torricelli (whose name coincidentally means little tower in Italian) formed a little tower of liquid mercury. He sealed a long glass tube at one end, filled it with mercury, and inverted it into a beaker (Fig. 4.4). The column of mercury fell until the pressure that it exerted at its base matched the pressure exerted by the atmosphere. To interpret measurements with a barometer, we need to find how the height of the column of mercury depends on the atmospheric pressure. [Pg.263]

In the single wavelength detector, 254 nm and 280 nm are strong lines from the mercury source and hence, are selected in most comnercial instruments. Of the three types of U.V. detectors, this one gives the lowest noise level (down to 0.00002 O.D.), but, of course, some flexibility is lost in not being able to work at other wavelengths. Nevertheless, when one purchases an instrument, this is the first detector selected. [Pg.235]

Fig. 15 Instrumented Zanasi LZ-64 showing mercury swivel for signal removal (A) dosator containing strain-gauged piston (B) mercury swivel. (From Ref. 38.)... [Pg.358]

Mercury is contained in several types of instruments that are commonly used by electrical utilities, municipalities, and households. Among others, these devices include barometers, meters, temperature... [Pg.1223]

The experiments described here were performed with a Digilab FTS40 Fourier transform instrument, with a liquid nitrogen-cooled Mercury Cadmium Telluride, (MCT), detector. The instrument is provided with a computer for data acquisition, storage and mathematical treatment. P-polarized incident light was obtained by means of an A1 wire-grid polarizer supported on a BaF2 substrate. [Pg.137]


See other pages where Mercury instrument is mentioned: [Pg.3007]    [Pg.3007]    [Pg.144]    [Pg.428]    [Pg.498]    [Pg.102]    [Pg.279]    [Pg.318]    [Pg.1317]    [Pg.9]    [Pg.53]    [Pg.592]    [Pg.612]    [Pg.676]    [Pg.181]    [Pg.136]    [Pg.401]    [Pg.168]    [Pg.148]    [Pg.366]    [Pg.1006]    [Pg.222]    [Pg.196]    [Pg.333]    [Pg.357]    [Pg.359]    [Pg.10]    [Pg.1222]    [Pg.327]    [Pg.433]   
See also in sourсe #XX -- [ Pg.114 , Pg.115 ]




SEARCH



Instrumentation, mercury porosimetry

© 2024 chempedia.info