Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Measured calorimeter types

For each of these control modes, different calorimeter types with different construction principles and different measuring principles exist (see Chapters 1 and 5). For reasons of clarity, we restrict ourselves in what follows to these three operation modes and collect the different calorimeters in three groups without further detailed classification characteristics. Numerous calorimeters belong to these groups, and it is not possible to present all of them in particular, we are not able to list all calorimeters that are commercially available but only typical examples to explain the respective characteristic properties. [Pg.157]

For measurement of the heat exchanged between the sample in the calorimeter vessel and the surroundings, different methods exist, which have already been explained in Part One of this book. We will not repeat the fundamentals in this chapter unless the details are of essential importance for the respective calorimeter type. [Pg.157]

Combustion or bomb calorimetry is used primary to derive enthalpy of fonuation values and measurements are usually made at 298.15 K. Bomb calorimeters can be subdivided into tluee types (1) static, where the bomb or entire calorimeter (together with the bomb) remains motionless during the experiment (2) rotating-... [Pg.1907]

Thermochemistry is concerned with the study of thermal effects associated with phase changes, formation of chemical compouncls or solutions, and chemical reactions in general. The amount of heat (Q) liberated (or absorbed) is usually measured either in a batch-type bomb calorimeter at fixed volume or in a steady-flow calorimeter at constant pressure. Under these operating conditions, Q= Q, = AU (net change in the internal energy of the system) for the bomb calorimeter, while Q Qp = AH (net change in the enthalpy of the system) for the flow calorimeter. For a pure substance. [Pg.351]

In the combustion reaction as carried out in the calorimeter of Figure 7-2, the volume of the system is kept constant and pressure may change because the reaction chamber is sealed. In the laboratory experiments you have conducted, you kept the pressure constant by leaving the system open to the surroundings. In such an experiment, the volume may change. There is a small difference between these two types of measurements. The difference arises from the energy used when a system expands against the pressure of the atmosphere. In a constant volume calorimeter, there is no such expansion hence, this contribution to the reaction heat is not present. Experiments show that this difference is usually small. However, the symbol AH represents the heat effect that accompanies a chemical reaction carried out at constant pressure—the condition we usually have when the reaction occurs in an open beaker. [Pg.112]

Figure 6-17 illustrates a constant-volume calorimeter of a type that is often used to measure q for combustion reactions. A sample of the substance to be burned is placed inside the sealed calorimeter in the presence of excess oxygen gas. When the sample bums, energy flows from the chemicals to the calorimeter. As in a constant-pressure calorimeter, the calorimeter is well insulated from its surroundings, so all the heat released by the chemicals is absorbed by the calorimeter. The temperature change of the calorimeter, with the calorimeter s heat capacity, gives the amount of heat released in the reaction. [Pg.393]

A survey of the literature shows that although very different calorimeters or microcalorimeters have been used for measuring heats of adsorption, most of them were of the adiabatic type, only a few were isothermal, and until recently (14, 15), none were typical heat-flow calorimeters. This results probably from the fact that heat-flow calorimetry was developed more recently than isothermal or adiabatic calorimetry (16, 17). We believe, however, from our experience, that heat-flow calorimeters present, for the measurement of heats of adsorption, qualities and advantages which are not met by other calorimeters. Without entering, at this point, upon a discussion of the respective merits of different adsorption calorimeters, let us indicate briefly that heat-flow calorimeters are particularly adapted to the investigation (1) of slow adsorption or reaction processes, (2) at moderate or high temperatures, and (3) on solids which present a poor thermal diffusivity. Heat-flow calorimetry appears thus to allow the study of adsorption or reaction processes which cannot be studied conveniently with the usual adiabatic or pseudoadiabatic, adsorption calorimeters. In this respect, heat-flow calorimetry should be considered, actually, as a new tool in adsorption and heterogeneous catalysis research. [Pg.193]

When heat is liberated or absorbed in the calorimeter vessel, a thermal flux is established in the heat conductor and heat flows until the thermal equilibrium of the calorimetric system is restored. The heat capacity of the surrounding medium (heat sink) is supposed to be infinitely large and its temperature is not modified by the amount of heat flowing in or out. The quantity of heat flowing along the heat conductor is evaluated, as a function of time, from the intensity of a physical modification produced in the conductor by the heat flux. Usually, the temperature difference 0 between the ends of the conductor is measured. Since heat is transferred by conduction along the heat conductor, calorimeters of this type are often also named conduction calorimeters (20a). [Pg.195]

Adiabatic calorimeters are complex home-made instruments, and the measurements are time-consuming. Less accurate but easy to use commercial differential scanning calorimeters (DSCs) [18, 19] are a frequently used alternative. The method involves measurement of the temperature of both a sample and a reference sample and the differential emphasizes the difference between the sample and the reference. The two main types of DSC are heat flux and power-compensated instruments. In a heat flux DSC, as in the older differential thermal analyzers (DTA), the... [Pg.310]

Proper calibration of the DSC instruments is crucial. The basis of the enthalpy calibration is generally the enthalpy of fusion of a standard material [21,22], but electrical calibration is an alternative. A resistor is placed in or attached to the calorimeter cell and heat peaks are produced by electrical means just before and after a comparable effect caused by the sample. The different heat transfer conditions during calibration and measurement put limits on the improvement. DSCs are usually limited to temperatures from liquid nitrogen to 873 K, but recent instrumentation with maximum temperatures close to 1800 K is now commercially available. The accuracy of these instruments depends heavily on the instrumentation, on the calibration procedures, on the type of measurements to be performed, on the temperature regime and on the... [Pg.311]

Heat capacities at high temperatures, T > 1000 K, are most accurately determined by drop calorimetry [23, 24], Here a sample is heated to a known temperature and is then dropped into a receiving calorimeter, which is usually operated around room temperature. The calorimeter measures the heat evolved in cooling the sample to the calorimeter temperature. The main sources of error relate to temperature measurement and the attainment of equilibrium in the furnace, to evaluation of heat losses during drop, to the measurements of the heat release in the calorimeter, and to the reproducibility of the initial and final states of the sample. This type of calorimeter is in principle unsurpassed for enthalpy increment determinations of substances with negligible intrinsic or extrinsic defect concentrations... [Pg.312]

The measurement of an enthalpy change is based either on the law of conservation of energy or on the Newton and Stefan-Boltzmann laws for the rate of heat transfer. In the latter case, the heat flow between a sample and a heat sink maintained at isothermal conditions is measured. Most of these isoperibol heat flux calorimeters are of the twin type with two sample chambers, each surrounded by a thermopile linking it to a constant temperature metal block or another type of heat reservoir. A reaction is initiated in one sample chamber after obtaining a stable stationary state defining the baseline from the thermopiles. The other sample chamber acts as a reference. As the reaction proceeds, the thermopile measures the temperature difference between the sample chamber and the reference cell. The rate of heat flow between the calorimeter and its surroundings is proportional to the temperature difference between the sample and the heat sink and the total heat effect is proportional to the integrated area under the calorimetric peak. A calibration is thus... [Pg.313]

Measurements based on the law of conservation of energy are of two main types. In phase change calorimetry the enthalpy of the reaction is exactly balanced by the enthalpy of a phase change of a contained compound surrounded by a larger reservoir of the same compound used to maintain isothermal conditions in the calorimeter. The latter enthalpy, the measurand, is often displayed indirectly through the change in the volumetric properties of the heat reservoir compound, e.g. ice/water. [Pg.314]

The Contalab, initially supplied by Contraves, was purchased by Mettler-Toledo, which is now placing less emphasis on this design than on the RC1. Some comments here are appropriate, however, since it is another type of bench-scale calorimeter, and units continue to be used. Its measuring system is based on the heat balance principle, in which a heat balance is applied over the cooling/heating medium. For this purpose, both the flow rate of the coolant and its inlet and outlet temperatures must be known accurately. Figure 3.12 is a schematic plan of the Contalab. [Pg.119]

Certain gaseous fluorides have been regarded as stable to hydrolysis, and it was therefore unexpected when Cady showed that C103F and S02F2 could be rapidly hydrolyzed in dilute alkali solutions (47). This was confirmed calorimetrically when it was shown that the rate of hydrolysis measured calorimetrically was dependent on mass transfer of gas across the gas-water interface. A bell-type calorimeter was used to overcome this problem (49,51). This type of calorimeter can be used for any gas-liquid reaction and is much more effective than passage of gas through sintered discs into solution. [Pg.24]

In the course of a long and thorough study of the polymerisation of isobutene (IB) by syncatalytic systems based on aluminium-organic compounds (Magagnini et al., 1977 and preceding papers) measurements were made by a Biddulph-Plesch type reaction calorimeter fitted with conductivity (k) electrodes on polymerisations of IB by Et2AlCl + Cl2 in MeCl at -45 °C, from which a kp+A value could be obtained. The reactions and procedures can be summarised as follows ... [Pg.562]

Calorimetry involves the use of a laboratory instrument called a calorimeter. Two types of calorimeters are commonly used, a simple coffee-cup calorimeter and a more sophisticated bomb calorimeter. In both, we carry out a reaction with known amounts of reactants and the change in temperature is measured. Check your textbook for pictures of one or both of these. [Pg.99]

The coffee-cup calorimeter can be used to measure the heat changes in reactions that are open to the atmosphere, qp, constant pressure reactions. We use this type of calorimeter to measure the specific heats of solids. We heat a known mass of a substance to a certain temperature and then add it to the calorimeter containing a known mass of water at a known temperature. The final temperature is then measured. We know that the heat lost by the added substance (the system) is equal to the heat gained by the surroundings (the water and calorimeter, although for simple coffee-cup calorimetry the heat gained by the calorimeter is small and often ignored) ... [Pg.100]


See other pages where Measured calorimeter types is mentioned: [Pg.227]    [Pg.35]    [Pg.241]    [Pg.155]    [Pg.224]    [Pg.310]    [Pg.374]    [Pg.288]    [Pg.594]    [Pg.1902]    [Pg.1905]    [Pg.1908]    [Pg.1910]    [Pg.1916]    [Pg.1918]    [Pg.466]    [Pg.245]    [Pg.780]    [Pg.309]    [Pg.71]    [Pg.228]    [Pg.340]    [Pg.348]    [Pg.35]    [Pg.515]    [Pg.311]    [Pg.318]    [Pg.62]    [Pg.155]    [Pg.95]    [Pg.306]   
See also in sourсe #XX -- [ Pg.86 ]




SEARCH



Calorimeters

Measurements types

© 2024 chempedia.info