Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Limit of quantitation, LOQ

Determine the limit of detection LOD and limit of quantitation LOQ according to the interpolation at level y = a + CL of the regression line and its lower CL this is sensitive to the calibration-point pattern ... [Pg.352]

Positive identification of low-ppb (pg/L) levels of endosulfan in human blood has been achieved by GC equipped with a microcoulometric detector (GC/MC) (Griffith and Blanke 1974). Although GC/MC is specific and nearly as sensitive as GC/ECD for detecting endosulfan in blood, GC/MC is more difficult to operate. Both isomers of endosulfan can be measured in blood using a method described by Guardino et al. (1996). According to the authors, endosulfan can be recovered and measured with an approximate limit of quantitation (LOQ) of 0.2 pg/L (sub-ppb). [Pg.249]

The limit of determination [or limit of quantitation (LOQ)] is defined in Directive 96/46/EC as the lowest concentration tested at which an acceptable mean recovery (normally 70-110%) and acceptable relative standard deviation (normally <20%) are obtained. The specific requirements for LOQ in crops, food, feed, soil, drinking and surface water, air, body fluids, and tissues are described in Section 4. Because the abbreviation LOD usually means limit of detection rather than limit of determination, the authors prefer not to use this abbreviation here in order to avoid confusion, and LOQ is used throughout. According to Directive 96/46/EC no data with regard to the limit of detection must be given. [Pg.24]

Table 5 Relation between the maximum residue Umit (MRL) and the limit of quantitation (LOQ)... Table 5 Relation between the maximum residue Umit (MRL) and the limit of quantitation (LOQ)...
The required limit of quantitation (LOQ) and limit of detection (LOD) have been extended to the parts per billion range as the European Community (EC) baby food -related guideline and the US consumer basket requirements became effective. [Pg.59]

Adequate sensitivity should be demonstrated and estimates of the limit of detection (LOD) and the limit of quantitation (LOQ) should be provided. The slope of the calibration line may indicate the ability of the method to distinguish the tme analyte concentration. The LOD of a method is the lowest analyte concentration that produces a reproducible response detectable above the noise level of the system. The LOQ is the lowest level of analyte that can be accurately and precisely measured. For a regulatory method, quantitation is limited by the lowest calibration standard. The techniques for these estimations should be described. [Pg.85]

Each commodity required a specifically customized workbook, containing a worksheet for each analyte determined in the commodity. Each laboratory received electronic copies of either three or four workbooks, which served as templates for the three or four commodities assigned to the laboratory. Each set of up to 10 commodity samples scheduled for colleetion and analysis required the creation of a copy of the appropriate template. Each workbook template contained one primary worksheet for each analyte, in which analytical data were recorded and residue levels were calculated, as described below. Eor example, the template for green beans contained 17 primary worksheets, one for each of the 17 analytes determined in each green bean sample. Additional worksheets were inserted into copies of the template as needed, to describe results of further analyses, such as confirmation of analytes present above the limit of quantitation (LOQ) or dilutions to bring the concentration of the analyte into the calibration range. [Pg.244]

Sensitivity is a measure of the smallest concentration that can be either measured [limit of detection (LOD)] or accurately quantitated [limit of quantitation (LOQ)]. In the USA, the method for measuring LOD or LOQ is left up to the method developer. European requirements for determining LOD and LOQ are very specific the LOD is based on the mean plus three standard deviations for 20 control blank samples, and the LOQ is defined as the lowest concentration giving an acceptable CV. [Pg.319]

The limit of detection (LOD) is an important criterion of the efficiency of an analytical method. It is characterized by the smallest value of the concentration of a compound in the analytical sample. The detectable amount of anilide compounds is in the range 0.01-0.5 ng by GC and 0.1 ng by HPLC. The limit of quantitation (LOQ) ranges from 0.005 to 0.01 mg kg for vegetables, fruits and crops. The recoveries from untreated plant matrices with fortification levels between 10 and 50 times the LOD and the LOQ are 70-120%. The relative standard deviation (RSD) at 10-50 times the level of the LOD and LOQ are <10 % and <20%, respectively. [Pg.335]

The limit of detection (LOD) and limit of quantitation (LOQ) were statistically determined in alfalfa and clover raw agricultural commodities (rotational crops). The method LOD and LOQ for the EMA-producing metabolite were 0.004 and 0.012 mg kg respectively, and the LOD and LOQ for the HEMA-producing metabolite were 0.004 and 0.014mgkg respectively, for the alfalfa and clover... [Pg.360]

At least four chromatographic standards prepared at concentrations equivalent to 50-70% of the limit of quantitation (LOQ) up to the maximum levels of analytes expected in the samples should be prepared and analyzed concurrently with the samples. In LC/MS/MS analysis, the first injection should be that of a standard or reagent blank and should be discarded. Then, the lowest standard should be injected, followed by two to four blanks, control samples, fortifications or investigation samples, followed by another chromatographic standard. This sequence is then repeated until all the samples have been injected. The last injection should be that of a standard. In order to permit unattended analysis of a normal analysis set, we recommend that samples and standards be made up in aqueous solutions of ammonium acetate (ca 5 mM) with up to 25% of an organic modifier such as acetonitrile or methanol if needed. In addition, use of a chilled autosampler maintained at 4 °C provides additional prevention of degradation during analysis. [Pg.404]

The MDCs are estimated from an S/N of the diphenyl ether peaks of at least of 3 in the recovery test. With fortification levels between 0.2 and 0.5mgkg recoveries of bifenox from brown rice matrices ranged from 85 to 102% with the limit of detection (LOD) and limit of quantitation (LOQ) being 0.010 mg kg according to the analytical method of the Notification of the Ministry of the Environment, Japan. By the residue analysis method described in Section 2.2.2(3), recoveries of chlornitrofen and CNP-NH2 from brown rice and vegetables with fortification levels of 0.04-0.10 mg kg ranged from 82 to 98%. The LOD for each sample was 0.005 mg kg for chlornitrofen and CNP-NH2. ... [Pg.458]

This enforcement method has been validated on the (raw agricultural commodities) (RAC) and processed parts of various crops. The method limit of quantitation (LOQ) was validated at 0.05 mg kg and the method limit of detection (LOD) was set at 0.01 mgkg for all of the crop matrices. The method flow chart is presented in Figure 1. [Pg.477]

Wheat samples are extracted with dilute ammonia on the ASE200. The extracts are amended with isotopically labeled internal standards. The extracts are purified by sequential octadecyl reversed-phase solid-phase extraction (Cig SPE) and ethylenediamine-iV-propyl anion exchange (PSA) SPE. The samples are analyzed by LC/MS/MS. This method determines crop residues of flucarbazone-sodium and A-desmethyl flucarbazone with a limit of quantitation (LOQ) of 0.01 mgkg for each analyte. [Pg.490]

For plant samples, the average recovery of flumioxazin from untreated confiol samples fortified within the range 0.1-0.01 mgkg ranged from 75 to 106%. The limit of quantitation (LOQ) is 0.01 mg kg and the limit of detection (LOD) is 0.005 mg kg. ... [Pg.507]

Dilute the extract with 1.5% acetic acid in water and mix completely. Suggested final dilution volumes are 2.5 mL for samples containing expected residues near the limit of quantitation (LOQ) level of lOngL The extracts are placed in vials for LC/MS/MS analysis. [Pg.512]

Another consideration when planning field fortification levels for the matrices is the lowest level for fortification. The low-level fortification samples should be set high enough above the limit of quantitation (LOQ) of the analyte so as to ensure that inadvertent field contamination does not add to and does not drive up the field recovery of the low-fortification samples. Setting the low field fortification level too low will lead to unacceptably high levels of the analyte in low field spike matrix samples if inadvertent aerial drift or pesticide transport occurs in and around where the field fortification samples are located. Such inadvertent aerial drift or transport is extremely hard to avoid since wind shifts and temperature inversions commonly occur during mixer-loader/re-entry exposure studies. [Pg.1009]


See other pages where Limit of quantitation, LOQ is mentioned: [Pg.260]    [Pg.285]    [Pg.26]    [Pg.42]    [Pg.99]    [Pg.140]    [Pg.394]    [Pg.501]    [Pg.530]    [Pg.597]    [Pg.607]    [Pg.637]    [Pg.689]    [Pg.721]    [Pg.763]    [Pg.775]    [Pg.833]    [Pg.850]    [Pg.954]    [Pg.1135]    [Pg.1152]    [Pg.1306]    [Pg.1325]    [Pg.218]   
See also in sourсe #XX -- [ Pg.268 ]




SEARCH



LOQ

Limit of quantitation

Quantitative limitations

© 2024 chempedia.info