Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Isolation blanks

To prevent vapor leaks, flashbacks, and other hazards, workers should completely isolate the space. To completely isolate a confined space, the closing of valves is not sufficient. All pipes must be physically disconnected or isolation blanks bolted in place. Other special precautions must be taken in cases where flammable liquids or vapors may recontaminate the confined space. The pipes blanked or disconnected should be inspected and tested for leakage. Other areas of concern are steam valves, pressure lines, and chemical transfer pipes. [Pg.55]

Only authorised, experienced and competent persons should carry out pressure testing. Work must be controlled, preferably by a permit-to-work system. The manager or supervisor must carry out an inspection of the system before starting the test, to ensure that isolations, blanks and other specified safety devices are in place before any testing begins. [Pg.217]

Maintenance error omission of an isolating blank/spade between the MIC tank and the connected product line under cleaning process using water... [Pg.124]

Some vessels are provided with two full-size relief valves so that one can be changed with the plant on line. On the plant side of the relief valves, isolation valves are usually provided below each relief valve, interlocked so that one relief valve is always open to the plant (Figure 10-2). If the relief valves discharge into a flare system, it is not usual to provide such valves on the flare side. Instead the relief valve is simply removed and a blank fitted quickly over the end of the flare header before enough air is sucked in to cause an explosion. Later the blank is removed and the relief valve replaced. [Pg.210]

When a blank reaction was run by purging the solution of pyrazinone (Scheme 22, pyrazinone b) in o-DCB with ethylene gas and irradiating it at 190 °C for 100 min, only a mere 53% conversion of the starting material was observed. Microwave-enhanced hydrolysis of the sensitive imidoyl chloride moiety of the cycloadduct using aqueous NaOH resulted in a yield of only 12%. However, the situation changed dramatically when the vial was pre-pressurized with ethylene gas at 5 bar. The reaction was completed after 30 min of microwave irradiation at 190 °C, and the hydrolyzed product was isolated in 87% yield. The reaction could be completed in a mere 10 min when carried out at 220 °C at an increased ethylene pressure of 10 bar, or in 20 min at 190 °C at 10 bar ethylene pressure. [Pg.284]

An extract blank (i.e., distilled water only) was also included throughout the extraction procedure to ensure that all isolated compounds were of biological origin. [Pg.250]

An ideal method for the preconcentration of trace metals from natural waters should have the following characteristics it should simultaneously allow isolation of the analyte from the matrix and yield an appropriate enrichment factor it should be a simple process, requiring the introduction of few reagents in order to minimise contamination, hence producing a low sample blank and a correspondingly lower detection limit and it should produce a final solution that is readily matrix-matched with solutions of the analytical calibration method. [Pg.303]

Flanged Joints. A flanged joint at which a blank is inserted to isolate other equipment during a test need not be tested. [Pg.130]

Soine PJ, Blanke RV, Chinchilli VM, et al. 1984a. High-density lipoproteins decrease the biliary concentration of chlordecone in isolated perfused pig liver. J Toxicol Environ Health 14(2-3) 319-335. [Pg.285]

The exit stream was immediately hydrolyzed with water, and fluo-rene was isolated by filtration and recrystallization from ethanol. Blank experiments indicated that no further exchange occurred after hydrolysis. In one experiment using a flow rate of 4.35 ml. per minute the exchanged fluorene analyzed (m/e, relative intensity) 166, 2.25 167, 32.68 168, 187.93. This corresponded to 84.45% d2, 14.6% dl9 and 1.0% do-fluorene. In another experiment the flow rate was 2.18 ml. per minute, and the product was analyzed (m/e, relative intensity) 166, 4.45 167, 36.0 168, 167.92. This corresponds to 80.7% d2, 17.3% d, and 2.0% d0-Auorene. [Pg.211]

It is not possible to prescribe specific pretreatment procedures here because these can only be decided upon when the system and the purpose of the experiments has been properly defined. However, a wealth of information exist in various biochemical reference books on how to isolate various biological compounds. The recommended techniques and methods could be used as part of the trace element speciation protocol often after slight modification, taking into consideration the following points First, the trace element blank levels have to be low, less than 10% of the total concentration in the sample. Second, the regents used should not interfere with subsequent analytical determinations. Third, the experimental conditions should not deviate markedly from those found in vivo, especially the pH and ionic strength of the medium. [Pg.149]

MS sensitivity depends on both the type of instrumentation and the nature of the analytes, but, typically, a minimum sample size of 5-10 ng is in most cases sufficient. Limited sensitivity in a certain application is often not due to the inherent sensitivity of the MS but rather the level of background impurities that are in the isolate. It is not always appreciated that very slowly eluting LC solutes from previous separations can create substantial MS background peaks that obscure analyte identification. Thus, it is important to use a blank sample to ensure that the background of the trapped fraction is adequately free of possible interferences to the desired identification. [Pg.715]

FTIR Microspectroscopy.3 A microscope accessory coupled to a liquid-nitrogen-cooled mercury-cadmium-telluride (MCT) detector can be used to obtain an IR spectrum. This is possible in both the transmission and reflectance modes. Several beads are spread on an IR-transparent window (NaCl, KBr, diamond) and possibly flattened via a hand-press or a compression cell. The IR beam is focused on a single bead using the view mode of the microscope. The blank area surrounding the bead is isolated using an adjustable aperture, and a spectrum is recorded using 32 scans (<1 min). A nearby blank area of the same size on the IR transparent window is recorded as the background. [Pg.221]

Sometimes the analyte is in such low concentration that it is impossible to isolate. It can be noted from equations (17.1) and (17.3) that it is not necessary to know Ax and As individually if their ratio can be determined. To achieve this, a reproducible reaction can be conducted on the labelled standard (analytical blank) and, in an identical fashion, on the sample in order to obtain the same quantity of derivatised compound. Thus the sub-stoichiometric method is similar to the immunochemical method for trace analysis. [Pg.334]


See other pages where Isolation blanks is mentioned: [Pg.695]    [Pg.695]    [Pg.388]    [Pg.965]    [Pg.235]    [Pg.5]    [Pg.88]    [Pg.79]    [Pg.142]    [Pg.410]    [Pg.702]    [Pg.280]    [Pg.660]    [Pg.27]    [Pg.95]    [Pg.217]    [Pg.263]    [Pg.323]    [Pg.109]    [Pg.210]    [Pg.444]    [Pg.140]    [Pg.64]    [Pg.93]    [Pg.342]    [Pg.247]    [Pg.95]    [Pg.128]    [Pg.82]    [Pg.291]    [Pg.471]    [Pg.552]    [Pg.623]    [Pg.273]    [Pg.273]    [Pg.110]   
See also in sourсe #XX -- [ Pg.55 ]




SEARCH



Blank

Blank, blanking

Blanking

© 2024 chempedia.info