Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Instruments used

The instrument uses a sinusoidal driver. The spectrum is very clean as we use a 14 bits signal generator. The probe signal is modulated in amplitude and phase by a defect signal. The demodulation is intended to extract the cartesian values X and Y of this modulation. [Pg.280]

Fig.3 Corrosion calibration standard for eddy curent instrument using the gradient method. Fig.3 Corrosion calibration standard for eddy curent instrument using the gradient method.
The first of them to determine the LMA quantitatively and the second - the LF qualitatively Of course, limit of sensitivity of the LF channel depends on the rope type and on its state very close because the LF are detected by signal pulses exceeding over a noise level. The level is less for new ropes (especially for the locked coil ropes) than for multi-strand ropes used (especially for the ropes corroded). Even if a skilled and experienced operator interprets a record, this cannot exclude possible errors completely because of the evaluation subjectivity. Moreover it takes a lot of time for the interpretation. Some of flaw detector producers understand the problem and are intended to develop new instruments using data processing by a computer [6]. [Pg.335]

It should be noted that these results are only preliminary and have to be considered as a proof of concept. As is clear from eq. (2) the phase contrast can be improved drastically by improving the global resolution and sensitivity of the instrument. Currently, a high resolution desktop system is under construction [5] in which the resolution is much better than that of the instrument used in this work, and in which the phase contrast is expected to be stronger by one order of magnitude. [Pg.577]

Another instrument used in physical chemistry research that employs quadnipole mass filters is the guided ion beam mass spectrometer [31]. A schematic diagram of an example of this type of instrument is shown in figure B 1.7.13. A... [Pg.1345]

The other type of x-ray source is an electron syncluotron, which produces an extremely intense, highly polarized and, in the direction perpendicular to the plane of polarization, highly collimated beam. The energy spectrum is continuous up to a maximum that depends on the energy of the accelerated electrons, so that x-rays for diffraction experiments must either be reflected from a monochromator crystal or used in the Laue mode. Whereas diffraction instruments using vacuum tubes as the source are available in many institutions worldwide, there are syncluotron x-ray facilities only in a few major research institutions. There are syncluotron facilities in the United States, the United Kingdom, France, Genuany and Japan. [Pg.1378]

Detection of cantilever displacement is another important issue in force microscope design. The first AFM instrument used an STM to monitor the movement of the cantilever—an extremely sensitive method. STM detection suffers from the disadvantage, however, that tip or cantilever contamination can affect the instrument s sensitivity, and that the topography of the cantilever may be incorporated into the data. The most coimnon methods in use today are optical, and are based either on the deflection of a laser beam [80], which has been bounced off the rear of the cantilever onto a position-sensitive detector (figme B 1.19.18), or on an interferometric principle [81]. [Pg.1693]

The isotope boron-10 is used as a control for nuclear reactors, as a shield for nuclear radiation, and in instruments used for detecting neutrons. Boron nitride has remarkable properties and can be used to make a material as hard as diamond. The nitride also behaves like an electrical insulator but conducts heat like a metal. [Pg.14]

Polarimeter (Section 7 4) An instrument used to measure op tical activity... [Pg.1291]

Measurements are made using appropriate equipment or instruments. The array of equipment and instrumentation used in analytical chemistry is impressive, ranging from the simple and inexpensive, to the complex and costly. With two exceptions, we will postpone the discussion of equipment and instrumentation to those chapters where they are used. The instrumentation used to measure mass and much of the equipment used to measure volume are important to all analytical techniques and are therefore discussed in this section. [Pg.25]

Techniques, such as spectroscopy (Chapter 10), potentiometry (Chapter 11), and voltammetry (Chapter 11), in which the signal is proportional to the relative amount of analyte in a sample are called concentration techniques. Since most concentration techniques rely on measuring an optical or electrical signal, they also are known as instrumental techniques. For a concentration technique, the relationship between the signal and the analyte is a theoretical function that depends on experimental conditions and the instrumentation used to measure the signal. For this reason the value of k in equation 3.2 must be determined experimentally. [Pg.38]

Analytical chemists make a distinction between calibration and standardization. Calibration ensures that the equipment or instrument used to measure the signal is operating correctly by using a standard known to produce an exact signal. Balances, for example, are calibrated using a standard weight whose mass can be traced to the internationally accepted platinum-iridium prototype kilogram. [Pg.47]

An error due to limitations in the equipment and instruments used to make measurements. [Pg.58]

To ensure that S eas is determined accurately, we calibrate the equipment or instrument used to obtain the signal. Balances are calibrated using standard weights. When necessary, we can also correct for the buoyancy of air. Volumetric glassware can be calibrated by measuring the mass of water contained or delivered and using the density of water to calculate the true volume. Most instruments have calibration standards suggested by the manufacturer. [Pg.130]

Frequently an analyst must select, from several instruments of different design, the one instrument best suited for a particular analysis. In this section we examine some of the different types of instruments used for molecular absorption spectroscopy, emphasizing their advantages and limitations. Methods of sample introduction are also covered in this section. [Pg.388]

Infrared instruments using a monochromator for wavelength selection are constructed using double-beam optics similar to that shown in Figure 10.26. Doublebeam optics are preferred over single-beam optics because the sources and detectors for infrared radiation are less stable than that for UV/Vis radiation. In addition, it is easier to correct for the absorption of infrared radiation by atmospheric CO2 and 1420 vapor when using double-beam optics. Resolutions of 1-3 cm are typical for most instruments. [Pg.393]

An analysis of variance can be extended to systems involving more than a single variable. For example, a two-way ANOVA can be used in a collaborative study to determine the importance to an analytical method of both the analyst and the instrumentation used. The treatment of multivariable ANOVA is beyond the scope of this text, but is covered in several of the texts listed as suggested readings at the end of the chapter. [Pg.697]

Other types of mass spectrometer may use point, array, or both types of collector. The time-of-flight (TOF) instrument uses a special multichannel plate collector an ion trap can record ion arrivals either sequentially in time or all at once a Fourier-transform ion cyclotron resonance (FTICR) instrument can record ion arrivals in either time or frequency domains which are interconvertible (by the Fourier-transform technique). [Pg.201]

Another form of array is called a microchannel plate detector. A time-of-flight (TOP) mass spectrometer collects ions sequentially in time and can use a point detector, but increasingly, the TOP instrument uses a microchannel plate, most particularly in an orthogonal TOP mode. Because the arrays and microchannel plates are both essentially arrays or assemblies of small electron multipliers, there may be confusion over their roles. This chapter illustrates the differences between the two arrays. [Pg.213]

Instruments Used to Measure Accurate Isotope Ratios... [Pg.365]

As normally used in the process industries, the sensitivity and percentage of span accuracy of these thermometers are generally the equal of those of other temperature-measuring instruments. Sensitivity and absolute accuracy are not the equal of those of short-span electrical instruments used in connection with resistance-thermometer bulbs. Also, the maximum temperature is somewhat limited. [Pg.760]

The principal classes of flow-measuring instruments used in the process industries are variable-head, variaBle-area, positive-displacement, and turbine instruments, mass flowmeters, vortex-shedding and iiltrasonic flowmeters, magnetic flowmeters, and more recently, Coriohs mass flowmeters. Head meters are covered in more detail in Sec. 5. [Pg.762]

Several centrifugal cuvet photocentrifuges are commercially available. These instruments use the same theoiy as the disc photocentrifuges but are hmited in operation to the homogeneous mode of operation. [Pg.1826]

Instrumentation used in different processes during its lifecycle. Surplus instrumentation or existing instrumentation reinstalled for different use. Possibility of instrumentation being used outside its design limits. [Pg.116]

The indicating instruments used for electrical measurements must conform to lEC 6005 I and have the following aectiracies ... [Pg.251]

After the heat run, the compressor continues to run on air and the highest pressure practical is imposed, while the speed is set to the normal operating speed. The capacity and power should be noted as well as bearing temperature and the other instrumentation used during the test. If oil buffered seals are used and the test run is expected to exceed 250°F, the test procedure may have to be modified to avoid the possibility of an explosion hazard. [Pg.413]


See other pages where Instruments used is mentioned: [Pg.9]    [Pg.50]    [Pg.251]    [Pg.306]    [Pg.318]    [Pg.1122]    [Pg.1244]    [Pg.445]    [Pg.374]    [Pg.389]    [Pg.583]    [Pg.775]    [Pg.195]    [Pg.96]    [Pg.165]    [Pg.65]    [Pg.344]    [Pg.411]    [Pg.8]   
See also in sourсe #XX -- [ Pg.388 , Pg.389 , Pg.390 , Pg.391 , Pg.392 , Pg.393 ]




SEARCH



Detection Using Instrumentation

Direct Optical Evaluation Using Instruments

Instrument Considerations when Using Ultra-High Pressures

Instrumentation Used for Pyrolysis

Instruments Used for Source Apportionment

Instruments Used in Safety Laboratories

Instruments and their use

Iron instruments used

Polarimeter An instrument used to measure

Pricing Derivative Instruments Using the Black-Scholes Model

Reasons for Using IR Instruments

Supercritical fluid chromatography instrumentation used

Telephone instruments, plastics used

Ultrasonic characterization instrument used

Use in electrochemical instrumentation

Use of light measuring instruments

What Tax Instruments Should Governments Use to Support Safety Nets

© 2024 chempedia.info