Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Sources instrumentation

That proportion of a measurement which arises from sources other than the analyte itself. Individual contributions from instrumental sources, added reagents and the matrix can, if desired, be evaluated separately. [Pg.617]

Pump and compressor seal areas are by far the most common areas where vapor releases may occur. This is followed by instrumentation sources, valve seals, gaskets and sample points and the most rare but usually catastrophic erosion and corrosion failures of process piping. [Pg.186]

There are some potential instrumental sources of error that could occur using this quantitation technique. It is critical to have minimal variability between each independent injection, as the quantitation is based on the comparison of the sample and standard areas. However, the current autosamplers are able to minimize this variability to less than 0.5% relative standard deviation (RSD). [Pg.13]

The primary difference between optical and electron microscopy is that the latter uses an electron beam as the probe. Since 10- to 500-keV electron beams have much lower wavelengths than light, the resolution is greater. At the same time, the electron beam requires completely different instrumentation (source, collimator, detector, magnification control, etc.). Moreover, electrons are very readily absorbed by matter. Therefore, the entire path of the beam, from source to specimen to detector, has to be in vacuum. From the sample preparation point of view, this is of major significance. For specimens that may change in vacuum, biological tissues, for instance, this can be a major concern, and newly developed accessories such as environmental cells [8] need to be added to the microscope. [Pg.380]

When alerted to a control problem, the first step should be to carefully inspect the analytical method, equipment, reagents, and specimens. Does everything look, feel, smell, and sound correct An inspection may seem to be a very qualitative and subjective technique, but it can be exceedingly useful when performed with checklists developed for specific analytical methods. This inspection should include a review of records documenting changes that occur with the instrument and reagents. Brief instrument function checks are often performed to verify proper system performance and to separate chemical and instrumental sources of errors. An experienced analyst can often spot the problem by making this kind of inspection, whereas inexperienced analysts will be aided by formal checklists. [Pg.519]

A possible instrumental source of error in the determination of organic analytes by LC methods or in FLA setups is adsorption on the tubing and ducts of the instrument. [Pg.956]

The interpretation of SSMS data falls into two distinct areas — element Identification and estimates of quantity. The criteria used in our laboratory for positive elemental identification are the presence of the doubly ionized species and the Identification of the Isotopic pattern when possible. Quantitation will be discussed later. Last, the Instrument source must be cleaned regularly to avoid memory problems. Our approach to the memory problem is to have a complete set of source parts for each matrix. A set of parts are dedicated for silicon analyses, another for gallium arsenide, etc. These parts and the source Itself are cleaned on a regular and frequent basis. When these factors are under control, SSMS has proved to be a reliable, reproducible technique for the bulk analysis of trace impurities. [Pg.309]

Other chemical and instrumental sources of nonlinearity in absorbance measurements may include hydrogen bonding, interaction with the solvent, nonlinear detector, nonlinear electronics, noncollimated radiation, and high signal levels (saturation). [Pg.505]

One source of noise of ihls type is the slow drift in the radiant output of the source. This type of noise can be called source flicker noise (Section 5B 2). The effects of nuctuations in the intensity of a source can be minimized by the use of a constani-voUagc power supply or a feedback system in which the source intensity is maintained at a constant level. Modern double-beam spectrophotometers (Sections I3D-2and I3D-3) can also help cancel the effect of flicker noise. With many instruments, source flicker noise does not limit performance. [Pg.346]

Instrumental Factors. Unsatisfactory performance of an instrument may be caused by fluctuations in the power-supply voltage, an unstable light-source, or a nonlinear response of the detector-amplifier system. A double-beam system helps to minimize deviations due to these factors. In addition, the following instrumental sources of possible deviations should be understood ... [Pg.170]

Figure 36 Assembly of Driver Control Instruments. Source KUKA Roboter GmbH)... Figure 36 Assembly of Driver Control Instruments. Source KUKA Roboter GmbH)...
Use of in situ testing equipment to obtain strength data encounters many actual or potential problems. Most of these problems cause a reduction in strength. A summary of instrumental sources of strength degradation for CPT and vane tests is listed in Table 4.4. [Pg.116]

Instrumental sources of errors must also be taken into account [14]. [Pg.414]

Figure 8.16 SLB test setup with instrumentation. Source [64] Reproduced with permission from Elsevier... Figure 8.16 SLB test setup with instrumentation. Source [64] Reproduced with permission from Elsevier...
Autozero the instrument while aspirating a deionized water blank. Monitor the variation in the baseline absorbance reading (baseline noise) for a few minutes to insure that the instrument, source lamp and associated equipment are in good operating condition. [Pg.1011]

Errors from instrumental sources, c/uenching. and luminescence can be kept very low (ca. [Pg.134]

Many spectrophotometers incorporate one or more moving parts that are subject to wear. A consequential slow loss of accuracy can go unnoticed unless the previous procedure is followed routinely. Another cause of slow loss of accuracy is instrument source aging. Variations from consecutive runs of a specimen or standard indicate that the source might be approaching the end of its useful life and should be changed. If this is the case, a new bulb will enable the instrument to output stable results. [Pg.484]


See other pages where Sources instrumentation is mentioned: [Pg.319]    [Pg.131]    [Pg.168]    [Pg.16]    [Pg.334]    [Pg.87]    [Pg.87]    [Pg.29]    [Pg.33]    [Pg.17]    [Pg.20]    [Pg.81]    [Pg.408]    [Pg.318]    [Pg.146]    [Pg.5]    [Pg.12]    [Pg.30]    [Pg.2964]    [Pg.2964]    [Pg.648]    [Pg.2]    [Pg.499]    [Pg.173]    [Pg.556]    [Pg.567]    [Pg.363]    [Pg.472]    [Pg.10]   
See also in sourсe #XX -- [ Pg.294 ]




SEARCH



Fluorescent detection, instrument excitation source

Instrumentation excitation sources

Instrumentation name-current source

Instrumentation radiation sources

Instrumentation radioactive source

Instrumentation/control systems ignition sources

Instruments Used for Source Apportionment

Luminescence instrumentation sources

Post-source decay instruments

Post-source decay reflectron instruments

SOURCE RANGE NUCLEAR INSTRUMENTATION

© 2024 chempedia.info