Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Instrumental settings

In general, work with the widest dynamic range available since transients and melting peaks can be very large, and set data point collection rates to be as fast as possible. For most work in this chapter, data were collected at a rate of 20 data points per second. [Pg.85]


The method covers the full dynamic range of linear velocities from a few centimetres/second to over 100 meters/seconds with one and the same instrumental set-up. Only the amount of tracer used per injection is varied. [Pg.1055]

The basic instrumental set-up for dynamic SIMS is the same as for SSIMS (Sect. 3.1.2). Depending on the intensity, beam diameter, and ion species needed, dif ferent ion sources are used. Several mass analyzers with different characteristics enable a broad field of applications. [Pg.108]

In addition, a further mechanism of chattering may be introduced in some liquid PR valve installations if the response characteristics of a control valve in the same system are such that hunting between the two occurs. Generally, this can be eliminated by adjustment of instrument settings or by installation of two valves with staggered set points... [Pg.168]

Have proper revisions been made to die process control logic, instrumentation set points and alarm points, especially for computer control systems to properly respond to the modification ... [Pg.25]

Melvin Mooney devised an instrument to measure the stiffness of uncured compounds, also known as the compound s viscosity. The unit of measurement is expressed in Mooney units. Figure 28.1 shows an instrument set up for measuring viscosity, the Mooney viscometer, in which a knurled knob (rotor) rotates (at 2 revolutions per minute) in a closed heated cavity filled with rubber [2]. [Pg.777]

In the preceding section, we presented principles of spectroscopy over the entire electromagnetic spectrum. The most important spectroscopic methods are those in the visible spectral region where food colorants can be perceived by the human eye. Human perception and the physical analysis of food colorants operate differently. The human perception with which we shall deal in Section 1.5 is difficult to normalize. However, the intention to standardize human color perception based on the abilities of most individuals led to a variety of protocols that regulate in detail how, with physical methods, human color perception can be simulated. In any case, a sophisticated instrumental set up is required. We present certain details related to optical spectroscopy here. For practical purposes, one must discriminate between measurements in the absorbance mode and those in the reflection mode. The latter mode is more important for direct measurement of colorants in food samples. To characterize pure or extracted food colorants the absorption mode should be used. [Pg.14]

A table of correlations between the variables from the instrumental set and variables from the sensory set may reveal some strong one-to-one relations. However, with a battery of sensory attributes on the one hand and a set of instrumental variables on the other hand it is better to adopt a multivariate approach, i.e. to look at many variables at the same time taking their intercorrelations into account. An intermediate approach is to develop separate multiple regression models for each sensory attribute as a linear function of the physical/chemical predictor variables. [Pg.438]

The instrumentation is also as varied as the variety of parameters to be measured. However, the choice is severely limited if the instrumentation needs to be of a turnkey variety. An obvious example is MRI instruments for clinical medicine that are technologically well suited for particle density imaging. MRI technicians in clinical MRI facilities are well trained to make standard images but, with rare exceptions, would not be able to make the more complex measurements needed, for example, for velocity imaging. Unfortunately, clinical MRI instruments are very expensive and also require expensive physical facilities to house them. Furthermore, medical facilities are usually not willing to divert the use of their expensive MRI instrument for purposes other than their needs, especially if instrument settings need to be changed. [Pg.506]

Principles and Characteristics Analytical multistage mass spectrometry (MSn) relies on the ability to activate and dissociate ions generated in the ion source in order to identify or obtain structural information about an unknown compound and to analyse mixtures by exploiting two or more mass-separating steps. A basic instrument for the currently most used form, tandem mass spectrometry (MS/MS), consists of a combination of two mass analysers with a reaction region between them. While a variety of instrument set-ups can be used in MS/MS, there is a single basic concept involved the measurement of the m/z of ions before and after a reaction in the mass spectrometer the reaction involves a change in mass and can be represented as ... [Pg.398]

Coulometric acid-base titration. The method below, described by Dahmen and Bos167, was applied to non-aqueous media (see part B), where precautions should be even more strictly adhered to than in aqueous media. The instrumental set-up, as depicted in Figs. 3.87-3.89, was originally developed by their former colleague Dubbeling together with Van den Enk-van Twillert). [Pg.239]

Figure 11.6 Positive ion electrospray mass spectra of an equimolar mixture of five standard proteins, under different instrumental settings, showing cases where prominent signals for the different charge states of (A) insulin, (B) ubiquitin, (C) cytochrome c, (D) lysozyme, and (E) myoglobin were preferentially observed, and (F) where signals for all the proteins were more uniformly detected. Figure 11.6 Positive ion electrospray mass spectra of an equimolar mixture of five standard proteins, under different instrumental settings, showing cases where prominent signals for the different charge states of (A) insulin, (B) ubiquitin, (C) cytochrome c, (D) lysozyme, and (E) myoglobin were preferentially observed, and (F) where signals for all the proteins were more uniformly detected.
The Inspect Disc File function is sometimes useful in diagnosing problems due to instrument malfunctions or inappropriate instrument settings. All functions that write files to the disc Include software write protection to guard against accidental erasure. [Pg.133]

There are several variations on the theme of instrument set up, which have been used in an attempt to overcome the shortcomings inherent in the concept. For example, as an alternative to the stop-flow method, the various fractions can be collected into sample loops (small loops of capillary tubing) which can then be flushed into the flow cell and studied at leisure. After spectroscopic examination, each sample can then be returned to its loop and the next pumped in. Fractions suffer dilution in this way but this approach would seem to offer an advantage over stop-flow in that at least the chromatography is not compromised by diffusion on the column. [Pg.144]

In the z-direction (depth-direction) the resolution is determined by the confocal instrument settings. While it is essential to be aware of the limitations of confocal measurements, as mentioned above, it is possible to create three-dimensional maps. That means probing a sample in the x, y and z-directions. In Figure 2 the data cube of a two-dimensional map is shown. One element (row) of the cube has been picked out and its content enlarged on the right side of the figure. It is evident that each row therefore contains the information of a whole spectrum. [Pg.531]

The diversity of markers that are reported in the literature for the three proteinaceous binders, even when the same derivatising agent is used, highlights how several aspects come together to determine the final pyrolysis products. Animal glue for example, when pyrolysed at 600°C in the presence of HMDS, gives rise to different markers depending on the pyrolyser and instrumental set-up used ... [Pg.307]

For these reasons, when the pyrolysis of samples from works of art is undertaken for the first time, as well as when a new device or a new instrumental set-up is introduced in a laboratory, it is very important to build one s own libraries of chromatographic profiles, based on the analysis of reference materials. [Pg.308]

Moreover, the diversity of results obtained for the aged linseed oil-lead white paint samples shown in Figure 11.1 and Figure 11.3a, highlights the important role played by reagent concentration and instrumental set-up. [Pg.312]

Typical instrument set-ups for online sample preparation, for example, solid phase extraction and sample pre-concentration, require the control of a six-port valve with an additional special column and a second pump. Ideally, this system will be fully controlled by the data acquisition software (Figure 3.12 (A)). [Pg.112]

Figure 6.1. Mass spectra of synthetic peptide, FLFQPQRF-NH2. Both spectra were recorded on an ESI ion trap mass spectrometer at different instrument settings (modifications of ion optics and ion-trap parameters). Figure 6.1. Mass spectra of synthetic peptide, FLFQPQRF-NH2. Both spectra were recorded on an ESI ion trap mass spectrometer at different instrument settings (modifications of ion optics and ion-trap parameters).
Automatic activation from process instrumentation set points. [Pg.117]

If you get to play with (in a nonpejorative sense) the instrument settings, you ll have to wade through the entire description. [Pg.228]

Open the valve on the pressure regulator of the carrier gas bottle and ensure that there is flow through the system. Turn on the instrument. Set the temperatures of the column, injector, and detector, and also set up the temperature program indicated in the introduction above. Allow time for all components to come to the set temperatures. [Pg.358]

A run is therefore regarded as being carried out under repeatability conditions, i.e. the random measurement errors are of a magnitude that would be encountered in a short period of time. In practice the analysis of a run may occupy sufficient time for small systematic changes to occur. For example, reagents may degrade, instruments may drift, minor adjustments to instrumental settings may be called for, or the laboratory temperature may rise. However, these systematic effects are, for the purposes of IQC, subsumed into the... [Pg.86]


See other pages where Instrumental settings is mentioned: [Pg.299]    [Pg.205]    [Pg.523]    [Pg.138]    [Pg.277]    [Pg.278]    [Pg.408]    [Pg.159]    [Pg.415]    [Pg.482]    [Pg.661]    [Pg.166]    [Pg.245]    [Pg.247]    [Pg.306]    [Pg.310]    [Pg.524]    [Pg.525]    [Pg.183]    [Pg.280]    [Pg.180]    [Pg.22]    [Pg.739]    [Pg.268]    [Pg.101]    [Pg.86]    [Pg.75]   


SEARCH



© 2024 chempedia.info