Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Pressure, measurement instrumentation limitation

A very small pressure difference is obtained for low rates of flow of gases, and the lower limit of velocity that can be measured is usually set by the minimum difference in pressure that can be measured. This limitation is serious, and various methods have been adopted for increasing the reading of the instrument although they involve the need for calibration. Correct alignment of the instrument with respect to the direction of flow is important this is attained when the differential reading is a maximum. [Pg.245]

Pressure drop measurements. For the majority of experiments the instrumentation was relatively similar. Due to limitations associated with the small size of the channels, pressures were not measured directly inside the micro-channels. To obtain the channel entrance and exit pressures, measurements were taken in a plenum or supply line prior to entering the channel. It is insufficient to assume that the friction factor for laminar compressible flow can be determined by means of analytical predictions for incompressible flow. [Pg.90]

You are to specify an orifice meter for measuring the flow rate of a 35° API distillate (SG = 0.85) flowing in a 2in. sch 160 pipe at 70°F. The maximum flow rate expected is 2000 gal/hr and the available instrumentation for the differential pressure measurement has a limit of 2 psi. What size orifice should be installed ... [Pg.336]

This chapter will be limited to the description of CSIL therapy to ex vivo studies in adult mammalian hearts. Due to page limitations, cell culture, gene delivery and in vivo studies will not be included. Therapeutic efficacy of CSIL in preservation of myocardial viability as well as function (by left ventricular developed pressure measurements) as assessed in globally ischemic Langendorff instrumented hearts is both dose and time dependent. This approach of cell membrane lesion repair and sealing may have broader applications in other cell systems. [Pg.305]

Details on points 1,2 and 3 are given in Sects. 2.3,2.4, and 2.5, respectively. The given lower limit of the dynamic range of the vapor pressure measurement is determined by the sensitivity of the instrument (upper limit see Sect. 2.4). Recent developments to improve the sensitivity are mentioned in Sect. 2.6.1. [Pg.101]

In Figure 10, the differential pressure measured between the inside and outside is plotted against time for toluene-filled containers at RT. The untreated containers quickly developed and maintained a vacuum at the limit of instrumentation (-30 mm Hg). [Pg.289]

The use of cone-and-plate rheometers for polymer melts is limited to relatively low shear rates by the onset of flow instabilities, typically occurring not far beyond the onset of shear-rate dependence for t](y) and the a iVi crossing point. A capillary rheometer is sketched in Fig. 3.22. Stable operation at much higher shear rates is possible, but usually t]o cannot be determined because of instrumental limitations at low shear rates. The steady-state viscosity, however, can be obtained from measurements of the volumetric flow rate, Q, and the pressure drop, AP = P — Po, Po being the ambient pressure. For long tubes (L/D 1), the following equation applies for Newtonian liquids ... [Pg.175]

For reasons of experimental or instrumental limitations, phase equilibria measurements are usually possible only over restricted ranges of pressure and temperature. If the equilibrium has been achieved at one point P-iT-i, it can be extrapolated to another point P2T2 by evaluation of a relation of the type -... [Pg.152]

For some hood types, measurements usually seen as indirect method, are used to measure the hood s performance to determine regulatory compliance. For example, regulations specify minimum and maximum face velocities for laboratory fume hoods and static pressure (negative) inside enclosed hoods. Continuously monitoring instruments can be connected to alarms that sound when the measurement is outside the specified limits. [Pg.1012]

Practically, polymers with molar masses between 2 x 104 and 2 x 106 g/mol can be characterized by membrane osmometry, but measurements of Mn <104 g/mol have also been reported with fast instruments and suitable membranes [16]. The lower limit is set by insufficient retention of short polymer chains. Above M 2 x 106 g/mol, the osmotic pressure, which is proportional to Mr1, is too low for a reasonable signal-to-noise ratio. An advantage of the low molar mass cut-off is that impurities with a very low molar mass can permeate through the membrane and, hence, do not contribute to the measured osmotic pressure. Their equilibration time may, however, be different from that of the solute, leading to complex time-dependent signals. [Pg.215]


See other pages where Pressure, measurement instrumentation limitation is mentioned: [Pg.36]    [Pg.335]    [Pg.76]    [Pg.181]    [Pg.76]    [Pg.181]    [Pg.278]    [Pg.278]    [Pg.160]    [Pg.582]    [Pg.571]    [Pg.229]    [Pg.232]    [Pg.3]    [Pg.31]    [Pg.588]    [Pg.282]    [Pg.366]    [Pg.36]    [Pg.528]    [Pg.57]    [Pg.386]    [Pg.283]    [Pg.159]    [Pg.382]    [Pg.325]    [Pg.327]    [Pg.104]    [Pg.43]    [Pg.117]    [Pg.131]    [Pg.150]    [Pg.267]    [Pg.64]    [Pg.187]    [Pg.255]    [Pg.21]    [Pg.81]    [Pg.187]   
See also in sourсe #XX -- [ Pg.119 ]




SEARCH



Instrumental limitations

Instrumentation limitations

Instrumentation measurement

Instruments measurements

Instruments, measuring measures

Measurement limitations

Measuring pressure

Pressure limit

Pressure limiting

Pressure measurement instruments

Pressure measurements

© 2024 chempedia.info