Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Instrument noise levels

Just as police officers need to maintain their weapons and routinely put in time on a shooting range, analytical chemists must always ensure that their instruments are calibrated and functioning properly. In police work, even if the butler confesses, without the proper police procedure, the most damning evidence can be thrown out. In the same manner, analytical chemists must be able to defend their results by reporting parameters such as confidence limits instrument noise levels, and significant figures. [Pg.316]

A detector s limit of detection (LOD) is the lowest concentration level that it can identify with a certain degree of confidence. There are many definitions of LOD, such as the concentration at which the response signal generated is three times the instrument noise level. Here LOD is referred to as the minimum detection level (MDL) of concentration that will consistently cause the detector to alarm. It is affected by background noise and blank signals. The LOD of a detector may vary widely for different chemicals. Environmental and operational conditions could drastically affect the LOD. Manufacturers normally provide LOD information obtained under optimum conditions. [Pg.106]

Time series, power spectra, and phase portraits are shown in Fig. 1 for the BZ reaction for three different flow rates [21]. The power spectra for the periodic states in Figs. 1(a) and (c) contain an instrumentally sharp fundamental frequency and its harmonics, while the spectrum in (b) consists of broadband noise that is well above the instrumental noise level. This spectral noise could arise from either stochastic or deterministic processes. However, at least in principle it should be possible to distinguish stochastic and deterministic processes from the behavior of the power spectrum in the high frequency limit [34] for stochastic differential equations of order n, P(a)) 0) ", while for nonperiodic behavior given by deterministic differential equations, P( o)) exp(-ra)). To our knowledge this... [Pg.125]

A Gaussian, process noise level of 0.1 % RMS noise-to-signal ratio is assumed, while the observation noise corresponds to approximately 10 % RMS. No process noise is added for the time-invariant parameter components. The addition of some minor noise could improve parameter estimation for the PF however, it might also lead to instabilities and non-converging behavior. The observation noise is chosen so as to reflect a realistic instrumentation noise level, whereas the process noise is kept to a low level indicating confidence that the observed system can be described by this type of a model formulation. [Pg.1687]

Berger et al. 2004) this is especially true for long-period seismic data (>100 s period) on very broadband instruments (e.g., Streckeisen STS-1 seismometer). The interpretation of such data only makes sense if the instrumental noise level is known. Also, research on noise levels in seismic recordings, the effect of noise reduction by the installation technique, and the nature and contribution of different noise sources to the recordings require knowledge of instrumental self-noise to rule out bias from the instrumentation self-noise. [Pg.3221]

The first of them to determine the LMA quantitatively and the second - the LF qualitatively Of course, limit of sensitivity of the LF channel depends on the rope type and on its state very close because the LF are detected by signal pulses exceeding over a noise level. The level is less for new ropes (especially for the locked coil ropes) than for multi-strand ropes used (especially for the ropes corroded). Even if a skilled and experienced operator interprets a record, this cannot exclude possible errors completely because of the evaluation subjectivity. Moreover it takes a lot of time for the interpretation. Some of flaw detector producers understand the problem and are intended to develop new instruments using data processing by a computer [6]. [Pg.335]

In the pneumatic pumping system, the pressure (and not the flow rate) is maintained constant as variations in chromatographic conditions occur. Thus, a change in mobile phase viscosity (e.g. gradient elution) or column back pressure will result in a change in flow rate for these types of pumps. The gas displacement pump in which a solvent is delivered to the column by gas pressure is an example of such a pneumatic pump. The gas displacement system is among the least expensive pumps available and is found in several low cost instruments. While the pump is nonpulsating and hence, produces low noise levels with the detectors in current use, its flow stability and reproducibility are only adequate. In addition, its upper pressure limit is only 2000 psi which may be too low in certain applications. [Pg.232]

In the single wavelength detector, 254 nm and 280 nm are strong lines from the mercury source and hence, are selected in most comnercial instruments. Of the three types of U.V. detectors, this one gives the lowest noise level (down to 0.00002 O.D.), but, of course, some flexibility is lost in not being able to work at other wavelengths. Nevertheless, when one purchases an instrument, this is the first detector selected. [Pg.235]

FTIR instrumentation is mature. A typical routine mid-IR spectrometer has KBr optics, best resolution of around 1cm-1, and a room temperature DTGS detector. Noise levels below 0.1 % T peak-to-peak can be achieved in a few seconds. The sample compartment will accommodate a variety of sampling accessories such as those for ATR (attenuated total reflection) and diffuse reflection. At present, IR spectra can be obtained with fast and very fast FTIR interferometers with microscopes, in reflection and microreflection, in diffusion, at very low or very high temperatures, in dilute solutions, etc. Hyphenated IR techniques such as PyFTIR, TG-FTIR, GC-FTIR, HPLC-FTIR and SEC-FTIR (Chapter 7) can simplify many problems and streamline the selection process by doing multiple analyses with one sampling. Solvent absorbance limits flow-through IR spectroscopy cells so as to make them impractical for polymer analysis. Advanced FTIR... [Pg.316]

Some tanks are installed with permanent leak identification sensors, which can check for leaked fuel vapor or liquid as it comes into contact with the sensors.21 However, these, as well as all the environmental sign tests (visual or instrumental) may be triggered by a spill instead of a leak. The success of external systems depends on the sensitivity of the sensor, the ability of the sensor to distinguish the stored chemical from other chemicals, the ambient background noise level of the stored chemical, the migration properties of the chemical, and the sampling network. [Pg.693]

Refractive index detectors are not as sensitive as uv absorbance detectors. The best noise levels obtainable are about 1CT7 riu (refractive index units), which corresponds to a noise equivalent concentration of about 10-6 g cmT3 for most solutes. The linear range of most ri detectors is about 104. If you want to operate them at their highest sensitivity you have to have very good control of the temperature of the instrument and of the composition of the mobile phase. Because of their sensitivity to mobile phase composition it is very difficult to do gradient elution work, and they are generally held to be unsuitable for this purpose. [Pg.73]

The size, speed, noise levels, precision, accuracy, and cost vary among the instruments. Higher cost does not necessarily mean better performance. Unlike typical spectroscopy, where the sample is reduced and, often, diluted in a non-interfering matrix, samples in NIR are read as is and one size instrument does NOT fit all. The application will determine the instrument (an idea that escapes many instrument salespeople). [Pg.172]

Limit of detection The method you choose must be able to detect the analyte at a concentration relevant to the problem. If the Co level of interest to the Bulging Drums was between 1 and 10 parts per trillion, would flame atomic absorption spectroscopy be the best method to use As you consider methods and published detection limits (LOD), remember that the LOD definition is the analyte concentration producing a signal that is three times the noise level of the blank, i.e., a S/N of 3. For real-world analysis, you will need to be at a level well above the LOD. Keep in mind that the LOD for the overall analytical method is often very different than the LOD for the instrumental analysis. [Pg.816]

This instrument features five detectors (Table 1.7). In the flame ionization detector, the high-speed electrometer, which ensures a very low noise level, is best suited to trace analysis and fast analysis using a capillary column. [Pg.65]

As an alternative to observing the structure of the residuals, statistical information about their magnitude is also readily available. In essence, after removing the correct number of eigenvalues, the standard deviation of the residuals should be the same as the noise level of the instrument, or in our case, the level of noise added. [Pg.223]

Of course, the problem in decay time measurements is improved noticeably if the noise level can be reduced. This is possible, but expensive. SPC instruments with peak counts of 500,000 have been developed. 59 ... [Pg.98]

Many of the problems encountered with the application of the Acoustic Emission (A.E) technique to corrosion problems are due to the fact that the signal level was of a similar order of magnitude to the background noise level of the instrumentation. [Pg.116]

The detection method can also be a source of potential variations from instrument to instrument. For example, UV detectors usually specify the wavelength accuracy to within +/-1-3 nm, and a change of this value could prove significant. Another, often ignored, factor is the time constant of the detector. Too low a setting of this factor can show significant noise levels and too high a value can distort the peak shape. [Pg.200]

Noise in the UV-Vis measurement originates primarily from the light source and electronic components. Noise in the measurement affects the accuracy at both ends of the absorbance scale. Photon noise from the light source affects the accuracy of the measurements at low absorbance. Electronic noise from the electronic components affects the accuracy of the measurements at high absorbance [8]. A high noise level affects the precision of the measurements and reduces the limit of detection, thereby rendering the instrument less sensitive. [Pg.162]


See other pages where Instrument noise levels is mentioned: [Pg.100]    [Pg.183]    [Pg.51]    [Pg.419]    [Pg.3230]    [Pg.99]    [Pg.100]    [Pg.183]    [Pg.51]    [Pg.419]    [Pg.3230]    [Pg.99]    [Pg.576]    [Pg.189]    [Pg.4]    [Pg.220]    [Pg.405]    [Pg.313]    [Pg.51]    [Pg.151]    [Pg.253]    [Pg.277]    [Pg.278]    [Pg.477]    [Pg.409]    [Pg.257]    [Pg.888]    [Pg.183]    [Pg.136]    [Pg.166]    [Pg.209]    [Pg.221]    [Pg.9]    [Pg.29]    [Pg.405]    [Pg.378]    [Pg.390]   
See also in sourсe #XX -- [ Pg.316 ]




SEARCH



Instrumental noise

Instrumentation noise

Level instruments

© 2024 chempedia.info