Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Fast HPLC chromatograms

FIGURE I Typical fast HPLC chromatograms for a set of three dissolution samples at three time points. [Pg.386]

Since the HPLC method for CU must be fast, use of DryLab or other commercially available software is recommended to develop a short run-time method. An example chromatogram in Figure 15-19 illustrates a fast LC run for a combination drug product containing three APIs. In addition, the use of HPLCs that can operate at higher pressures (up to 9000-15,000psi) could also be used. More details on fast HPLC can be found in Chapter 17. [Pg.708]

Figure 7.26. HPLC chromatogram showing the performance of a pellicular column (Zorbax Poroshell 300-SB C4, 50 x 4.6mm, 5pm) in fast protein separation. Inset shows the structure of Zorbax Poroshell 300-SB particles. Diagram courtesy of Agilent Technologies. Figure 7.26. HPLC chromatogram showing the performance of a pellicular column (Zorbax Poroshell 300-SB C4, 50 x 4.6mm, 5pm) in fast protein separation. Inset shows the structure of Zorbax Poroshell 300-SB particles. Diagram courtesy of Agilent Technologies.
Figure 24.1 HPLC chromatogram of biotin standard with post-column avidin reaction. Note the fast chromatography permitted with this sensitive and specific reaction. Detection is using fluorescence. A biotin standard, 0.04ug/mL. B supplemented milk powder extract with 12ug/100g. Figure 24.1 HPLC chromatogram of biotin standard with post-column avidin reaction. Note the fast chromatography permitted with this sensitive and specific reaction. Detection is using fluorescence. A biotin standard, 0.04ug/mL. B supplemented milk powder extract with 12ug/100g.
Over the years, in addition to developments with ELISA reagents such as labels, there have been improvements in automation. This has enabled ELISA to be utilized as a high-throughput tool. Typically, ELISAs can be performed in several hours to days. The most common practice is to precoat the microtiter plate for an overnight incubation period, with the remainder of the steps performed the following day. While ELISAs are fast when compared to other assays such as bioassays, which can take days to weeks, they might be considered slow when compared to methods like HPLC, in which the time from sample injection to chromatogram is a matter of minutes. [Pg.297]

FIGURE IS Comparative chromatograms showing the deleterious effect of instrumentai bandwidth to the performance of a Fast LC separation of four paraben anti-microbiais. HPLC conditions coiumn CIS (33x4.6 mm, 3fun), mobile phase 50% acetonitriie/water, 2mL/min, detection at 254 nm. [Pg.70]

Most of the traditional HPLC detectors can be applied to LCxLC analyses the choice of the detectors used in comprehensive HPLC setup depends above all on the nature of the analyzed compounds and the LC mode used. Usually, only one detector is installed after the second-dimension column, while monitoring of the first-dimension separation can be performed during the optimization of the method. Detectors for microHPLC can be necessary if microbore columns are used. Operating the second dimension in fast mode results in narrow peaks, which require fast detectors that permit a high data acquisition rate to ensure a proper reconstruction of the second-dimension chromatograms. [Pg.109]

In parallel with recent developments in GC, multidimensional HPLC (LC x LC) is now also finding application in environmental analysis.33 The combination of two sufficiently different separation dimensions (e.g., NP-HPLC x RP-HPLC or IC x RP-HPLC), however, remains difficult because of the solvent compatibility issues discussed above. Here, too, HILIC may bring about a significant improvement, since its mobile phase requirements are much closer to RP-HPLC than those of other liquid chromatographic techniques.34 In contrast to GC x GC, LC x LC cannot be implemented with a (thermal) modulator that collects the analytes after the first separation dimension and reinjects them into the second column it is most practically realized with a double-loop interface that alternately collects and transfers the analytes from the first to the second dimension (Figure 13.7). Even though the second dimension chromatogram is also very fast, detection is not normally a problem since the peak widths in the second dimension are usually still of the order of 1-2 s. [Pg.313]

In normal high pressure liquid chromatography, typical sample volumes are 20-200 p.L this can become as little as 1 nL in capillary HPLC. Pretreatment of the sample may be necessary in order to protect the stationary phase in the column from deactivation. By employing supercritical fluids such as carbon dioxide, pretreatment can be bypassed in many instances so that whole samples from industrial and environmental matrices can be introduced directly into the column. This is due to the fact that the fluid acts as both extraction solvent and mobile phase. Post-column electrochemistry has been demonstrated. For example, fast-scan cyclic voltammo-grams have been recorded as a function of time after injection of microgram samples of ferrocene and other compounds in dichloromethane solvent and which are eluted with carbon dioxide at pressures of the order of 100 atm and temperatures of 50°C the chromatogram is constructed as a plot of peak current vs. time [18]. [Pg.577]

Before starting extensive experiments, a procedure recommended by Kaiser and Oel-rich (1981) to rule out adsorbents by fast experiments should be employed. Each elution experiment takes about 20 s. For this purpose samples are applied on a 50 X 50 mm TLC plate at 9 points, which are exactly 10 mm apart. Five microlitres of methanol are drawn into a micro-capillary with a Pt-Ir point. By applying the point of the filled capillary on one of the sample points on the plate, methanol is introduced onto the plate. A miniature radial chromatogram of ca. 7 mm diameter is produced. If the sample components remain at the point of application, the use of this adsorbent type is ruled out for HPLC usage. To make sure, the procedure is repeated with 5 pi of acetonitrile and tetrahydrofuran, respectively. If the products still remain at the point of application, the situation will not be changed by using any other mobile phase that is suitable for preparative chromatography work. [Pg.136]

HPLC is a fast separation technique. The mixture to be separated is transferred to a column with a solvent or a solvent mixture eluent mobile phase). The column is a tube, in most cases of stainless steel, filled with the stationary phase. The separation happens right there in the column. Under optimal conditions the components to be separated pass through the stationary phase at different rates and leave the colunrn after different times. The components (the solutes) are registered by a detector. This information is passed on to the data evaluation unit and the output is a chromatogram. The number of peaks is equal to the number of separated components in the sample (not necessarily of the components actually present ), and the area is proportional to the amount. [Pg.5]

Another way to speed up HPLC that does not interfere with the existing gradient separation method is by parallel operation of several HPLC columns. The development in this direction started a couple of years ago when fast gradient separation was first combined with mass spectrometry-based detection. Parallel column operation was achieved by a single pumping system and a splitter tee that transferred the gradient flow onto two HPLC columns. The effluent of the two columns was simultaneously sprayed into a modified ion spray interface of a quadrupole mass spec-trometer. From the overlay chromatogram both desired and previously known compounds were identified after their molecular ions were filtered from the total ion current (TIC). In this first system, however, it was difficult to enhance the parallelization, and the detection system created a bottleneck. ... [Pg.308]


See other pages where Fast HPLC chromatograms is mentioned: [Pg.386]    [Pg.386]    [Pg.137]    [Pg.357]    [Pg.138]    [Pg.118]    [Pg.118]    [Pg.172]    [Pg.646]    [Pg.320]    [Pg.201]    [Pg.262]    [Pg.262]    [Pg.113]    [Pg.55]    [Pg.35]    [Pg.712]    [Pg.201]    [Pg.95]    [Pg.172]    [Pg.182]    [Pg.85]    [Pg.562]    [Pg.31]    [Pg.135]    [Pg.275]    [Pg.62]    [Pg.453]    [Pg.615]    [Pg.97]    [Pg.90]    [Pg.228]    [Pg.374]   
See also in sourсe #XX -- [ Pg.386 ]




SEARCH



Fast HPLC

HPLC chromatograms

© 2024 chempedia.info