Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Faraday cage, noise

Electronic noise. For a noise amplitude of 0.1 A only atoms separated by 4A can be resolved. Thus, the electronics must be carefully shielded, with the pre-amplifier stage placed as close to the tip as possible and the STM itself placed in a Faraday cage. [Pg.79]

Contemporary potentiometers contain circuits with high internal resistances. The resistance of the electrode membranes thus creates no serious problem, but the leads to the electrodes should always be shielded and the high-resistance systems should be placed in a Faraday cage otherwise the measurement is subject to noise. [Pg.64]

The main limitation to the use of microelectrode voltammetry is the need to measure small DC currents without noise interference. This usually requires that measurements be done in a Faraday cage (a shield against electronic noise). Recent reviews give good insight into the theory and application of microelec-trode voltammetry.48-51... [Pg.79]

Avoid external electric fields External devices such as electric motors, pumps, and fluorescent lighting emanate electric fields that can contribute significantly to the apparent noise in a system. This influence, seen most easily in high-impedance systems, can be mitigated by use of a Faraday cage, described in the following section. [Pg.149]

ISEs it is common practice to use potential measuring instruments with input impedances >10 Cl to ensure that there is no error in the potential measurement. Most modern pH/mV meters constructed with field-effect transistor-type input amplifiers fulfill this requirement. However, as the electrode surface area becomes smaller, the resistance of the ISE increases dramatically. Thus, for microsized electrodes, specially designed amplifier circuits with even higher input impedances are required to obtain accurate intracellular ion values and to help eliminate noise. In many instance, the micro-type measurements must also be made within the confines of a Faraday cage to reduce noise further by shielding the electrodes finm environmental noise. In automated clinical chemistry analyzers, confinement of the electrodes within the outer metal cabinet of the instrument serves a similar purpose. [Pg.13]

Figure 8 is a block diagram of a typical two-electrode configuration for making microelectrode measurements where the function generator and recorder could, of course, be replaced by a microcomputer with appropriate interfaces. To minimize noise, the cell is mounted in a Faraday cage and cables are kept as short as possible. Using a system of this type, noise-free measurements of steady-state currents as small as 10 12 A have been made... [Pg.158]

The signal-to-noise ratio can be improved if a low-noise battery-operated potentiostat is used and it is usually not necessary to resort to a two-electrode configuration unless the photocurrents are very small or the electrode capacitance unusually large. It is often useful to reduce the sample area to match the illumination spot so as to eliminate the noise contribution from the part of the electrode that is not illuminated. Particular care is necessary to eliminate earth loops and high-frequency pick-up and a screened Faraday cage is essential. Commercial reference electrodes can be replaced by low-resistance electrodes if they cause problems or bridging capacitors can be used to bypass high-resistance liquid junctions. [Pg.366]

M KCl solution placed in another chamber, a sb spanning over the two chambers. For a two-electrode system as is usually used in the measurement, the newly cut tip of the metallic wire, coated with absorbed lipids, acts as the WE. The current through the s-BLM is measured in the auxiliary electrode during the CV. The setup is housed in a Faraday cage to minimize interference by external noise and electrical transients. In spite of shielding, external noise may still be picked up by the switch box therefore, for the critical measurements the switch box should be incorporated within the same Faraday shield as the cell. All cables used are shielded and the shields are grounded. [Pg.445]

In order to test the instrument and the data analysis procedures we made a number of experiments on dummy cells, for example the one shown in Fig.6. Here we made use of the possibility of time-domain averaging, the ac excitation was repeated 8 times and the data were averaged. The dummy cell was enclosed in a Faraday cage and the noise level was quite low, reducing the need for frequency-domain averaging, i.e., averaging of admittance data, after Fourier transformation and calculation of the admittance. [Pg.26]

Procedure. A carbon fiber beveled electrode of about 10 diameter was polished and washed thoroughly with distilled water prior to each experimental run. The electrode was inserted into the vacuole of the Chara cell immersed in APW with the aid of micromanipulator under a binocular microscope. An Ag/AgCl reference electrode and a Pt wire auxiliary electrode were placed in the APW pool as illustrated in Fig. 15. The Chara cell was then left standing for 10-60 min in APW until the protoplasmic stream was restored. The measuring system was positioned inside a Faraday cage to minimize electrical noise pick-up. Differential pulse voltammetry was applied for the measurement. [Pg.481]

R. M. Wightman When you go to very small electrodes, you indeed have very small currents to measure. It turns out that with modern electronics it is not a particularly difficult problem. Only two precautions were taken first, experiments were done in a Faraday cage, second, all our equipment is home-made so that we can build into it the noise-free level current measurements that I showed you here. It is important to have very low currents, because if we pass high currents we may stimulate the neurons themselves and perturb the system that we are trying to make a measurement on. So the fact that we use very small electrodes has two advantages we have very small currents and we cause minimum perturbation to our tissues. [Pg.203]

The complete detection system is schematically shown in Fig. 7 to avoid noise from the electrical part of the laser it is placed inside a Faraday cage. Triggering of the recording circuit is made by a part of the laser beam via a vacuum photodiode (Valvo XA 1003, jitter l ns) or externally via an opto-coupler (jitter 50 ns). ... [Pg.53]


See other pages where Faraday cage, noise is mentioned: [Pg.993]    [Pg.336]    [Pg.28]    [Pg.100]    [Pg.398]    [Pg.180]    [Pg.38]    [Pg.443]    [Pg.686]    [Pg.53]    [Pg.422]    [Pg.412]    [Pg.650]    [Pg.656]    [Pg.164]    [Pg.67]    [Pg.5]    [Pg.55]    [Pg.156]    [Pg.5]    [Pg.337]    [Pg.282]    [Pg.291]    [Pg.15]    [Pg.17]    [Pg.73]    [Pg.93]    [Pg.28]    [Pg.152]    [Pg.631]    [Pg.378]    [Pg.386]    [Pg.203]    [Pg.576]    [Pg.267]    [Pg.466]    [Pg.494]   
See also in sourсe #XX -- [ Pg.281 , Pg.335 ]




SEARCH



Faraday

Faraday cage

© 2024 chempedia.info