Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Enhancement Column Reactions

A calculation of the relative detector signals shows that the conversion of carbonic acid to potassium bicarbonate and then to potassium hydroxide in the enhancement columns is essentially quantitative. In particular, it may seem surprising that an acid as weak as carbonic acid (k] a = 4.0 X 10 ) is able to exchange its H+ for on the resin. [Pg.174]

However, two points should be kept in mind. One is that an increasing fraction of carbonic acid is ionized as the solution becomes more dilute (-6.8 % in 0.1 mM carbonic acid, for example). A second point is that the high concentration of K+ on the exchange column (-4.2 M) pushes the ion-exchange equilibrium to the right. For [Pg.174]

1 mM carbonic acid, it can be calculated that only a few theoretical plates would be needed for complete conversion of H2CO3 to K+ and HC03 . [Pg.174]

A second enhancement column, placed just after the first enhancement column, provides still more sensitive detection. This is an anion-exchange column in the hydroxide form that converts K HCOs (equivalent conductance 118) to K+OH (equivalent conductance 272). [Pg.174]

The chromatograms in Fig. 8.4 show the effect of enhancement columns. Each enhancement column increases the carbonic acid peak height, but the baseline conductance is also increased somewhat. A pre-column packed with anion-exchange resin in the OH form was then placed between the pump and loop injector to remove completely and continuously the carbon dioxide in the eluent. This arrangement resulted in a significant decrease in eluent background conductance, as shown in Fig. 8.4D. An almost linear calibration plot was obtained from 0.05 to 5.0 mM bicarbonate. The detection limit was estimated to be 1.45 pM. [Pg.174]


Continuous chromatography in the packed annular space between the walls of two concentric cylinders can be done by rotating the assembly about its longitudinal axis (1, 2, 3). Rotation transforms the temporal separation that would be obtained under fixed, pulsed operation into a spatial separation that permits continuous operation. It has recently been shown that continuous reaction chromatography can be done in similar apparatus (4, Jj). This not only provides a means of carrying out chemical reaction and separation simultaneously in one unit, but for A B + C the product separation suppresses the rate of the back reaction and provides a means of enhancing the reaction yield. Yield enhancement in pulsed column chromatography has been demonstrated (6, 8). Yields of... [Pg.297]

A protein-binding assay (BA) coupled with hplc provided a highly sensitive post-column reaction detection system for the biologically important molecule biotin and its derivative biocytin, biotin ethylenediamine, 6-(biotinoylamino) caproic acid, and 6-(biotinoylamino)caproic acid hydrazide (71). This detection system is selective for the biotin moiety and responds only to the class of compounds that contain biotin in their molecules. In this assay a conjugate of streptavidin with fluorescamine isothiocyanate (streptavidin—FITC) was employed. Upon binding of the analyte (biotin or biotin derivative) to streptavidin—FITC, an enhancement in fluorescence intensity results. This enhancement in fluorescence intensity can be directly related to the concentration of the analyte and thus serves as the analytical signal. The hplc/BA system is more sensitive and selective than either the BA or hplc alone. With the described system, the detection limits for biotin and biocytin were found to be 97 and 149 pg, respectively. [Pg.245]

An important constituent in copper pyrophosphate baths is nitrate, which enhances the maximum permissible current density [31]. Fig. 8-30 shows the respective chromatogram with the separation of nitrate and orthophosphate. The latter is the hydrolysis product of pyrophosphate that is formed during the plating process. The main component pyrophosphate may also be separated on a latexed anion exchanger. It is detected after complexation with ferric nitrate in a post-column reaction by measuring the light absorption (see Section 3.3.5.2). [Pg.369]

Detector F ex 320 em 390 following post-column reaction. The column effluent mixed with 12% triethanolamine (for fluorescence enhancement) pumped at 0.5 mL/min and flowed through aim reaction coil to the detector. [Pg.939]

In RD processes isothermal operation of the catalyst bed is not possible. At the top of the column temperatures in the range of 60 °C are common, with higher temperature towards the bottom. Higher temperatures enhance the reaction more than mass transfer. Therefore mass-transport phenomena will have more pronounced effects at higher temperatures. We tested the catalysts at temperatures of 60, 75, and 90 °C. The results are shown in Fig. 8.13 and Fig. 8.15 for polymer/carrier catalysts with different polymer content. GFP-15 is a catalyst with low polymer content GFP-12 has twice the polymer content of GFP-15. GFP-15 was chosen because it shows the classical MTBE kinetic pattern. GFP-12 was tested in an RD column. [Pg.205]

In the second, separation is used to enhance the reaction. An example of this is the Intentional biphasing of a homogeneous liquid-phase reaction by addition of a second liquid phase. The second phase can act in several ways to enhance the productivity of the system. Another example is the large-scale version of the chemist s apparatus in which a reflux column condenser is connected to a batch reactor (usually a round-bottomed flask) as shown in Figure 25.1. The product and the heat of reaction are continuously removed, and the reactant is returned to the reactor. [Pg.792]

Table 25.1 lists several combinations of reaction and separation. The sequencing of the two in the nomenclature of the different combinations clearly reveals their orientations. This chapter is primarily concerned with reactive extraction (also termed dissociation-extraction), extractive reaction, reactive distillation (or dissociation-extractive-distillation), and distillative reaction (or distillation column reactors). Crystallization is almost always used for separation and seldom for enhancing a reaction. A notable exception is when one of the reactants is a sparingly dissolving solid and the size of the crystallizing solid is less than the thickness of the film surrounding the reactant. Then the crystallizing microphase enhances the rate of dissolution and hence the rate of reaction, a situation that was considered in Chapter 23. [Pg.793]

Unfortunately, carbonic acid is a very weak acid (pfCi = 6.4), and the conductance of the carbonic acid peak is consequently very low. In order to obtain more sensitive detection, Tanaka and Fritz inserted a cation-exchange column in the lU form between the cation-exchange column and the detector [19]. Its purpose is to convert the carbonic acid to a more highly ionized form and thereby to increase the conductivity. This is called the first enhancement column. When it is in the lU form, the exchange reaction is as follows ... [Pg.222]

Two small ion-exchange enhancement columns are connected in series with the outlet of the separation column. The following reaction of hydrazine takes place in the first anion exchange column, which is in the sulfate form ... [Pg.226]

The parameter p (= 7(5 ) in gas-liquid sy.stems plays the same role as V/Aex in catalytic reactions. This parameter amounts to 10-40 for a gas and liquid in film contact, and increases to lO -lO" for gas bubbles dispersed in a liquid. If the Hatta number (see section 5.4.3) is low (below I) this indicates a slow reaction, and high values of p (e.g. bubble columns) should be chosen. For instantaneous reactions Ha > 100, enhancement factor E = 10-50) a low p should be selected with a high degree of gas-phase turbulence. The sulphonation of aromatics with gaseous SO3 is an instantaneous reaction and is controlled by gas-phase mass transfer. In commercial thin-film sulphonators, the liquid reactant flows down as a thin film (low p) in contact with a highly turbulent gas stream (high ka). A thin-film reactor was chosen instead of a liquid droplet system due to the desire to remove heat generated in the liquid phase as a result of the exothermic reaction. Similar considerations are valid for liquid-liquid systems. Sometimes, practical considerations prevail over the decisions dictated from a transport-reaction analysis. Corrosive liquids should always be in the dispersed phase to reduce contact with the reactor walls. Hazardous liquids are usually dispensed to reduce their hold-up, i.e. their inventory inside the reactor. [Pg.388]


See other pages where Enhancement Column Reactions is mentioned: [Pg.174]    [Pg.222]    [Pg.174]    [Pg.222]    [Pg.245]    [Pg.424]    [Pg.39]    [Pg.712]    [Pg.12]    [Pg.174]    [Pg.9]    [Pg.1229]    [Pg.119]    [Pg.292]    [Pg.680]    [Pg.229]    [Pg.201]    [Pg.482]    [Pg.1319]    [Pg.321]    [Pg.417]    [Pg.233]    [Pg.735]    [Pg.228]    [Pg.210]    [Pg.226]    [Pg.129]    [Pg.98]    [Pg.112]    [Pg.168]    [Pg.463]    [Pg.158]    [Pg.165]    [Pg.600]    [Pg.621]   


SEARCH



Enhancer reaction

Reaction column

© 2024 chempedia.info