Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Electrode control circuitry

The temperature sensor in the membrane center is made of polysilicon with a nominal resistance of 10 kQ. An additional reference resistor is needed for the control circuitry (Sect. 5.1). For the resistance measurement of the sensitive layer, platinum electrodes are deposited on top of the CMOS aluminum metallization in order to establish good electrical contact to the sensitive metal oxide. [Pg.31]

They also developed a system using negative DEP [37]. Field minima (unlike maxima) can be created remote from electrode surfaces and negative dielectrophoretic traps do not need balancing forces and active control circuitry. [Pg.92]

The electrodes usually are rotated at about 600 rpm. Contact to the platinum wire is made internally by filling the electrode with mercury. A stationary wire dips into the mercury pool at the top to make external contact to the potential-control circuitry. The platinum-glass seals are prone to crack, which causes erratic currents that are associated with the leaking of mercury to the electrode surface. Once cracked, the electrodes are not easily repaired and should be discarded. Table S.9 indicates the dependence of the current on the rotational speed of the electrode. [Pg.230]

Controlling the current through a cell is simpler than controlling the potential at an electrode, because only two elements of the cell, the working and counter electrodes, are involved in the control circuit. In galvanostatic experiments, one is usually interested in the potential of the working electrode with respect to a reference electrode, and circuitry is normally added to permit that measurement, but it makes no contribution to the control function. [Pg.644]

Fig. 12 Current measurement by use of a differential amplifier to measure the voltage drop across Rm connected to the auxiliary electrode. Other control circuitry is not shown. Fig. 12 Current measurement by use of a differential amplifier to measure the voltage drop across Rm connected to the auxiliary electrode. Other control circuitry is not shown.
The section of Fig. 24 to the left is marked Polari2ation Control and is involved in electrical control of the cell potential or current. Note that the instrument can operate as either a potentiostat or galvanostat and the potential or current applied to the cell is programmed by summing the internal sources, marked DC Ref and Sweep, with an external polarity input. The box marked Feedback/Bandwidth control represents actual control circuitry similar to that in Figs. 8 and 11. Relays and electronic switches, controlled by the internal microprocessor, allow switching between potentiostatic or galvanostatic mode. Tbe microprocessor also sets the control loop bandwidth, which allows the experimenter to trade bandwidth for increased control loop stability. Other inputs to the feedback control circuitry are the RE potential and IR compensation (if necessary). A power amplifier is inserted at the counter (auxiliary) electrode connection. This allows currents of up to 2 A to be applied to the cell. [Pg.48]

Amperometric with macroelectrodes is done using approved technical equipment. There are numerous commercially available devices with equal design. The electrochemical cell works with three electrodes (Chap. 2, Sect. 2.4), i.e. working, auxiliary and reference electrodes. They are driven by a potentiostat designed as an analogue electronic drcuit Digital potential control instead of analogue control circuitry did not stand the test of time, since it does... [Pg.173]

The redox shuttle [13-15, 17] is an electrolyte additive that can be reversibly oxidized/reduced at a characteristic potential and provides an intrinsic overcharge protection for lithium-ion batteries that neither increases the complexity and weight of control circuitry nor permanently disables the cell when activated. The redox shuttle molecule (S) has its defined redox potential, at which it can be oxidized on the positive electrode and form a radical cation (S ) (see Equation 1). [Pg.123]

The basic instrumentation required for controlled-potential experiments is relatively inexpensive and readily available commercially. The basic necessities include a cell (with a three-electrode system), a voltammetric analyzer (consisting of a potentiostatic circuitry and a voltage ramp generator), and an X-Y-t recorder (or plotter). Modem voltammetric analyzers are versatile enough to perform many modes of operation. Depending upon the specific experiment, other components may be required. For example, a faradaic cage is desired for work with ultramicroelectrodes. The system should be located in a room free from major electrical interferences, vibrations, and drastic fluctuations in temperature. [Pg.100]

Electrochromic materials (either as an electroactive surface film or an electroactive solute) are generally first studied at a single working electrode, under potentiostatic or galvanostatic control, using three-electrode circuitry.1 Traditional electrochemical techniques such as cyclic voltametry (CV), coulometry, and chronoamperometry, all partnered by in situ spectroscopic measurements... [Pg.582]

Aside from the selectivity criterion that is essential to all ion specific electrodes, the principal objective of applied design is to physically and chemically control the phase and interphase boundaries across the multiple layers that comprise the electrode structure. The conduction path of electrical charge across all the phases including the solid conductors and external measurement circuitry, as well as the chemical charge across polymeric and solution phases, may be represented by the schematic illustrated in Figure 4. [Pg.268]

Turn on the 02-electrode system and allow the circuitry to warm up or stabilize for 5 min. Check the operation of the recorder during this interval. Then set the recorder on pen and set the zero point adjustment control of the recorder to achieve a zero setting on the recorder during zero output from the 02-electrode system. This usually requires that you disconnect the 02-electrode system from the recorder or switch a zero output control on the 02-electrode system. [Pg.236]

The detector was claimed to be moderately linear over a dynamic range of three orders of magnitude but values for the response index are not known. It is also not clear whether the associated electronics contained signal modifying circuitry or not. The disadvantages of this detector included erosion of the electrodes due to "spluttering", contamination of the electrodes from sample decomposition and the need for a well-controlled vacuum system. [Pg.162]

Fig. 14.11 (a) The electrode assembly consisting of eight sets of Pt electrodes fitted with a row of microtiter wells, (b) Circuitry of the software-controlled 16-channel potentiostat (Reprinted with permission from Tang et al.60. Copyright (2002) American Chemical Society)... [Pg.340]

Figure 9.6.2 Schematic of speed control and disk-current processing circuitry. Only the current follower of the conventional three-electrode potentiostat that controls the disk electrode is shown. [Reprinted with permission from B. Miller, and S. Bruckenstein, Anal Chem., 46, 2026 (1974), Copyright 1974, American Chemical Society.]... Figure 9.6.2 Schematic of speed control and disk-current processing circuitry. Only the current follower of the conventional three-electrode potentiostat that controls the disk electrode is shown. [Reprinted with permission from B. Miller, and S. Bruckenstein, Anal Chem., 46, 2026 (1974), Copyright 1974, American Chemical Society.]...

See other pages where Electrode control circuitry is mentioned: [Pg.49]    [Pg.93]    [Pg.156]    [Pg.3331]    [Pg.2977]    [Pg.2082]    [Pg.141]    [Pg.313]    [Pg.230]    [Pg.270]    [Pg.348]    [Pg.434]    [Pg.2430]    [Pg.106]    [Pg.19]    [Pg.300]    [Pg.169]    [Pg.222]    [Pg.699]    [Pg.348]    [Pg.122]    [Pg.70]    [Pg.282]    [Pg.18]    [Pg.2129]    [Pg.3847]    [Pg.2695]    [Pg.823]    [Pg.64]    [Pg.269]    [Pg.643]    [Pg.2672]    [Pg.311]   
See also in sourсe #XX -- [ Pg.30 ]




SEARCH



Circuitry

Electrode control

Electrode controlling

© 2024 chempedia.info