Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Electrochemical corrosion identical metals

Electrochemical noise measurements may be performed in the potentiostatic mode (current noise is measured), the galvanostatic mode (potential noise is measured), or in the ZRA mode (zero resistance ammeter mode, whereby both current and potential noise are measured under open-circuit conditions). In the ZRA mode, two nominally identical metal samples (electrodes) are used and the ZRA effectively shorts them together while permitting the current flow between them to be measured. At the same time, the potential of the coupled electrodes is measured versus a low-noise reference electrode (or in some cases a third identical electrode). The ZRA mode is commonly used for corrosion monitoring. [Pg.451]

Identical metals in contact with different concentrations In this case, the metal immersed in a dilute solution is dissolved from the electrode and deposited on the electrode immersed in a more concentrated solution. The other type of electrochemical concentration ceU is known as a differential aeration cell. In this case, the electrode potential difference occurs when the electrode is immersed in the same electrolyte with different oxygen partial pressures. Differential aeration initiates crevice corrosion in aluminum or stainless steel when exposed to a chloride environment. [Pg.4]

Figure 29.4 shows an example, the energy diagram of a cell where n-type cadmium sulfide CdS is used as a photoanode, a metal that is corrosion resistant and catalytically active is used as the (dark) cathode, and an alkaline solution with S and S2 ions between which the redox equilibrium S + 2e 2S exists is used as the electrolyte. In this system, equilibrium is practically established, not only at the metal-solution interface but also at the semiconductor-solution interface. Hence, in the dark, the electrochemical potentials of the electrons in all three phases are identical. [Pg.568]

In another study [35], the electrochemical emission spectroscopy (electrochemical noise) was implemented at temperatures up to 390 °C. It is well known that the electrochemical systems demonstrate apparently random fluctuations in current and potential around their open-circuit values, and these current and potential noise signals contain valuable electrochemical kinetics information. The value of this technique lies in its simplicity and, therefore, it can be considered for high-temperature implementation. The approach requires no reference electrode but instead employs two identical electrodes of the metal or alloy under study. Also, in the same study electrochemical noise sensors have been shown in Ref. 35 to measure electrochemical kinetics and corrosion rates in subcritical and supercritical hydrothermal systems. Moreover, the instrument shown in Fig. 5 has been tested in flowing aqueous solutions at temperatures ranging from 150 to 390 °C and pressure of 25 M Pa. It turns out that the rate of the electrochemical reaction, in principle, can be estimated in hydrothermal systems by simultaneously measuring the coupled electrochemical noise potential and current. Although the electrochemical noise analysis has yet to be rendered quantitative, in the sense that a determination relationship between the experimentally measured noise and the rate of the electrochemical reaction has not been finally established, the results obtained thus far [35] demonstrate that this method is an effective tool for... [Pg.742]

Aqueous corrosive media are electrolyte solutions, and hence by far the greatest proportion of damage arising through corrosion is due to electrochemical processes. Since in metals these are largely identical with electrolysis processes, metal corrosion in aqueous media is best described as electrolytic corrosion. [Pg.535]

In electrochemical noise (EN) measurements fluctuations in potential or current are measured as a function of time. The measurements can be done (see Chapter 7) either without or with an externally applied signal. In the first case one monitors the open circuit corrosion potential of the test metal versus a suitable reference electrode or versus a second electrode of the seune material exposed under identical conditions. The advantage of this technique for use in MIC research is that there is no external signal to disturb the biological community on the metal surface. Alternatively, one can measure fluctuations in potential (E) at an applied current (I), or the reverse, fluctuations in I at an applied E. It has also been suggested that one could couple the metal of interest to a platinum electrode and measure the noise... [Pg.516]

Obviously, electrochemical studies are unique in the analysis of corrosion rates, as they offer direct access to the corrosion rate. If only divalent ions will result in the corrosion reaction, then a current density of 1 nA cm" is identical to a mass flux of 5.2x10 mol cm" s (equal to a metal... [Pg.294]

Figure 5.14 b Effect of anodic Tafel slope on polarization and current necessary for cathodic protection of three different metals with (hypothetically) identical orr tmd Jcorr. but different Tafel P slopes. (From Jones, D.A. (1981). Principles of Measurement and Prevention of Buried Metal Corrosion by Electrochemical Polarization. In Underground Corrosion, STP 741, ASTM. Jones, D.A. (1971). Corrosion Science. 11,439. Reproduced by the kind permission of ASTM, Philadelphia, PA, USA)... [Pg.282]


See other pages where Electrochemical corrosion identical metals is mentioned: [Pg.181]    [Pg.123]    [Pg.1046]    [Pg.347]    [Pg.1636]    [Pg.1075]    [Pg.599]    [Pg.347]    [Pg.121]    [Pg.884]    [Pg.89]    [Pg.93]    [Pg.249]    [Pg.266]   
See also in sourсe #XX -- [ Pg.32 ]




SEARCH



Corrosion electrochemical

Corrosion metals

Corrosion, metallic

Electrochemical metal

Electrochemical metallization

Identical metals

Metals electrochemical corrosion

© 2024 chempedia.info