Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Detectors timing stability

The solvent delivery system is one of the most important components of the liquid chromatograph since its performance directly affects retention time reproducibility d detector baseline stability [1,2,39,40]. As well as the pump, other components such as check valves, flow controllers, mixing... [Pg.795]

Smaller values of are obtained for interferometers operated in a double-beam mode, since the moveable mirror must be left stationary for a fraction of the cycle time to allow the detector to stabilize each time the beam is switched from the sample to the reference position. With an optical null grating spectrometer the chopper is used not only to modulate the beam but also to alternate the beam between sample and reference channels. Thus, it takes approximately the same time to measure a transmittance spectrum using a double beam optical null spectrometer as it takes to measure a single-beam spectrum with the same S/R. Hence, for this type of spectrometer may be assigned a value of 2. [Pg.23]

An additional push can be expected from new technical developments in TCSPC itself. The largest potential is probably in the development of new detectors. The introduction of direct (wide-field) imaging techniques is clearly hampered by the limited availability of position-sensitive detectors. In addition the selection of multianode PMTs is still very limited, especially for NIR-sensitive versions. Large-area detectors with 64 or more channels may result in considerable improvements in DOT techniques. Single photon APDs with improved timing stability are urgently required for single-molecule spectroscopy and time-resolved microscopy. [Pg.348]

The advent of low priced powerful microcomputers has enabled easy processing of spectrophotometric data. In addition, fast scanning detectors have allowed elimination of errors resulting from poor time stability when a dual wavelength method is used with a single wavelength spectrophotometer. Modem multiwavelength analysis utilizes the reversed matrix representation of Beer-Lambert s law. This can be combined with chemometric techniques such as principal component analysis (PCA) and partial least squares (PLS). [Pg.4506]

A beta attenuation sampler uses a 30-mCi Krypton-85 source (with energy of 0.74 MeV) and detector to determinate the attenuation caused by deposited aerosols on a moving filter. lb improve the stability over time, a refiertticc reading is period-icallv made of a foil with attenuation similar to that of the Alter and collected aerosol. [Pg.1290]

Before performing TPO, the wet catalyst was dried under N2 flowing 10 Fh at 120°C for a period of 17 h before being slowly cooled to 20°C. Once the catalyst was stabilized at 20°C the pure N2 was switched to a 4 vol.% O2 in N2 mixture flowing at 10 1/h while the catalyst s temperature was ramped to 800°C at the rate of 8°/min. The O2 content of the gas mixture leaving the catalyst bed was measured by an Oxynos 100 paramagnetic detector and plotted with respect to either time or temperature. The plot of the vol.% O2 with respect to time was used to determine the O2 consumption of the catalyst. [Pg.500]

Before any slit operation check, write down, or save the old motor positions Operation of slits can be useful to change the beam intensity (instead of operating absorbers). Imperfect thermal stabilization of mirrors and monochromators can be compensated by proper slit operation. Before such operation is undertaken, it should be made sure that the instrument is close to thermal equilibrium. In particular after opening the main beam shutter for the first time, it may be indicated to wait for several hours. Otherwise the operator will have to follow the thermal expansion continuously. This bears the risk to destroy the adjustment or even the detector. [Pg.68]

Parameters that should be tested in HPLC method development are flow rate, column temperature, batch and supplier of the column, injection volume, mobile phase composition and buffer pH, and detection wavelength [2], For GC/GLC methods, one should investigate the effects of column temperature, mobile phase flow rate, and column lots or suppliers [38], For capillary electrophoresis, changes in temperature, buffer pH, ionic strength, buffer concentrations, detector wavelength, rinse times, and capillaries lots and supplier should be studied [35, 36], Typical variation such as extraction time, and stability of the analytical solution should be also evaluated [37],... [Pg.256]

Biosensors are the analytical systems, which contain sensitive biological elements and detectors. Plant cells as a possible biosensors have natural structure that determinates their high activity and stability. Criteria in the screening of the plant cells as biosensors for allelopathy should be as under (i) Reaction is fast based on the time of response, (ii) Reaction is sensitive to small doses of analysed compounds or their mixtures and (iii) Methods of detection viz., biochemical, histochemical, biophysical (in particular, spectral changes in absorbance or fluorescence) are easy in laboratory and in the field conditions. The search of biosensors in active plant species is suitable to determine the mechanisms of action of biologically active substances or external factors of the environment (Roshchina and Roshchina, 2003 Roshchina, 2004 2005 c)). [Pg.26]

So how does the IRMS get its stability Collector slits are several times the width of the ion beams. This gives a flat-topped peak shape (Fig 6) which makes the ion current intensive to drift. The main source of drift is temperature variation which both affects the electronic components used for mass selection and caused expansion and contraction of mechanical parts. Simultaneous measurement of ion beams using a double or triple collector is more precise than sequential measurement by mass scanning with a single detector. Finally, frequent comparison of sample gas under identical conditions also contributes to stability. Ion beam stability is more important than resolution for isotopic measurements. [Pg.160]


See other pages where Detectors timing stability is mentioned: [Pg.243]    [Pg.249]    [Pg.718]    [Pg.165]    [Pg.43]    [Pg.119]    [Pg.119]    [Pg.317]    [Pg.243]    [Pg.12]    [Pg.103]    [Pg.199]    [Pg.1634]    [Pg.607]    [Pg.1426]    [Pg.389]    [Pg.240]    [Pg.53]    [Pg.3]    [Pg.181]    [Pg.435]    [Pg.640]    [Pg.656]    [Pg.357]    [Pg.30]    [Pg.194]    [Pg.211]    [Pg.245]    [Pg.389]    [Pg.345]    [Pg.347]    [Pg.292]    [Pg.282]    [Pg.179]    [Pg.202]    [Pg.1076]    [Pg.436]    [Pg.157]    [Pg.211]    [Pg.316]    [Pg.357]   
See also in sourсe #XX -- [ Pg.296 ]




SEARCH



Stabilization time

Timing stability

© 2024 chempedia.info