Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Correction for instrumental

As previously stated, GPC is the method of choice for studying polymer degradation kinetics. The GPC trace, as given by the detector output, does not provide the true MWD due to various diffusion broadening processes inside the different parts of the equipment. The first step is to correct for instrument broadening if a precise evaluation of MWD is desired. Even with the best columns available, this correction may change the MWD significantly as can be visualized... [Pg.134]

Many current multidimensional methods are based on instruments that combine measurements of several luminescence variables and present a multiparameter data set. The challenge of analyzing such complex data has stimulated the application of special mathematical methods (80-85) that are made practical only with the aid of computers. It is to be expected that future analytical strategies will rely heavily on computerized pattern recognition methods (79, 86) applied to libraries of standardized multidimensional spectra, a development that will require that published luminescence spectra be routinely corrected for instrumental artifacts. Warner et al, (84) have discussed the multiparameter nature of luminescence measurements in detail and list fourteen different parameters that can be combined in various combinations for simultaneous measurement, thereby maximizing luminescence selectivity with multidimensional measurements. Table II is adapted from their paper with the inclusion of a few additional parameters. [Pg.12]

Equation (2) is true in the case where there is no instrumental spreading, that is, where all particles present in the detector cell are of exactly the same diameter. Because of axial dispersion, and skewing caused by entrapment or adsorption, corrections for instrumental spreading are required. In this case x v) is described by ... [Pg.29]

In the simplest approach T is the full width of the peak (measured in radians) subtended by the half maximum intensity (FWHM) corrected for the instrumental broadening. The correction for instrumental broadening is very important and can be omitted only if the instrumental broadening is much less than the FWHM of the studied diffraction profile, which is always the case in presence of small nanoclusters. The integral breadth can be used in order to evaluate the crystallite size. In the case of Gaussian peak shape, it is ... [Pg.132]

Fluorescence spectra were recorded using an SLM 4800 spectrofluorimeter (Bioritech, Chamarande, France) fitted with a thermostat-controlled (30°C) front-surface accessory. The incidence angle of the excitation radiation was 60°. Coagulation kinetics were performed in a quartz cuvette 1 cm x 1cm. All spectra were corrected for instrumental distortions in excitation using a rhodamine cell in the reference channel. [Pg.283]

For precise measurement of isotopic composition by mass spectrometry, it is also common to use either a natural, known isotopic ratio to correct for instrumental mass fractionation (e g., internal normalization) or to add a tracer for this purpose. For example for natural uranium samples, one can use the natural U/ U of 137.88 to correct for fractionation. Alternatively, one can use an added double spike of ratio -unity... [Pg.27]

In contrast to thermal ionization methods, where the tracer added must be of the same element as the analyte, tracers of different elemental composition but similar ionization efficiency can be utilized for inductively coupled plasma mass spectrometry (ICPMS) analysis. Hence, for ICPMS work, uranium can be added to thorium or radium samples as a way of correcting for instrumental mass bias (e g., Luo et al. 1997 Stirling et al. 2001 Pietruszka et al. 2002). The only drawback of this approach is that small inter-element (e g., U vs. Th) biases may be present during ionization or detection that need to be considered and evaluated (e.g., Pietruszka et al. 2002). [Pg.27]

However, thorium has only two naturally occurring long-lived isotopes, and all Th measurements by TIMS are limited by the absence of a well-constrained isotope ratio that can be used for internal normalization purposes to correct for instrumental mass fractionation. In this regard, one of the most important advantages of MC-ICPMS over MC-TIMS is the ability to admix two elements with overlapping mass ranges and use the... [Pg.48]

For PyMS to be used for (1) routine identification of microorganisms and (2) in combination with ANNs for quantitative microbiological applications, new spectra must be comparable with those previously collected and held in a data base.127 Recent work within our laboratory has demonstrated that this problem may be overcome by the use of ANNs to correct for instrumental drift. By calibrating with standards common to both data sets, ANN models created using previously collected data gave accurate estimates of determi-nand concentrations, or bacterial identities, from newly acquired spectra.127 In this approach calibration samples were included in each of the two runs, and ANNs were set up in which the inputs were the 150 new calibration masses while the outputs were the 150 old calibration masses. These associative nets could then by used to transform data acquired on that one day to data acquired at an earlier data. For the first time PyMS was used to acquire spectra that were comparable with those previously collected and held in a database. In a further study this neural network transformation procedure was extended to allow comparison between spectra, previously collected on one machine, with spectra later collected on a different machine 129 thus calibration transfer by ANNs was affected. Wilkes and colleagues130 have also used this strategy to compensate for differences in culture conditions to construct robust microbial mass spectral databases. [Pg.333]

In principle, the three isotope method may be widely applied to new isotope systems such as Mg, Ca, Cr, Fe, Zn, Se, and Mo. Unlike isotopic analysis of purified oxygen, however, isotopic analysis of metals that have been separated from complex matrices commonly involves measurement of several isotopic ratios to monitor potential isobars, evaluate the internal consistency of the data through comparison with mass-dependent fractionation relations (e.g., Eqn. 8 above), or use in double-spike corrections for instrumental mass bias (Chapter 4 Albarede and Beard 2004). For experimental data that reflect partial isotopic exchange, their isotopic compositions will not lie along a mass-dependent fractionation line, but will instead lie along a line at high angle to a mass-dependent relation (Fig. 10), which will limit the use of multiple isotopic ratios for isobar corrections, data quality checks, and double-spike corrections. [Pg.17]

They are used as above, but in general they lead to more straightforward interpretations. The correction for instrumental fractionation involves an isotope pair for which the measurements are in agreement with the terrestrial ratio, whether this choice results from the measurement itself or from model considerations. In cases where all the ratios are different from the terrestrial ratios, model considerations are used to interpret the data. In the most common cases, one isotope displays wider variations than the others and constitutes a guideline for modeling the origin of the anomalies. [Pg.28]

Iron. Fe has 4 isotopes of which the heaviest Fe has a very small abimdance of about 0.3%. The precision of thermal ionization mass spectrometers is around 10 s on this isotope and there is only a hint in some normal inclusions for an excess in 5 Fe (VoUcening and Papanastassiou 1989). Recent ICPMS measurements at the 2 s precision level display normal isotopic compositions for Fe in planetary materials but no Allende inclusion was reported in this study (Kehm et al. 2003). If excesses of similar magnitude to Ca, Ti, Cr were present they would not be clearly resolved in agreement with the observations. When Fe and Fe are used to correct for instrumental mass fractionation, Fe exhibits normal abundances, suggesting all three isotopes are present in solar relative abundances. [Pg.35]

Corrections for instrumentally-produced mass fractionation that preserve natural mass dependent fractionation can be approached in one of two ways a double-spike method, which allows for rigorous calculation of instrumental mass fractionation (e.g., Dodson 1963 Compston and Oversby 1969 Eugster et al. 1969 Gale 1970 Hamelin et al. 1985 Galer 1999 see section Double-spike analysis ), or an empirical adjustment, based on comparison with isotopic analysis of standards (Dixon et al. 1993 Taylor et al. 1992 1993). The empirical approach assumes that standards and samples fractionate to the same degree during isotopic analysis, requiring carefully controlled analysis conditions. Such approaches are commonly used for Pb isotope work. However, it is important to stress that the precision and accuracy of isotope ratios determined on unknown samples may be very difficult to evaluate because each filament load in a TIMS analysis is different. [Pg.117]

Rigorous correction for instrumental mass bias is required if the precision of an isotope ratio measurement needs to be greater than l%o per mass unit. This concept is well illustrated by the definitive Ca isotope work of Russell et al. (1978), which used a double-spike approach. Prior to the Ca isotope investigation of Russell et al. (1978), natural mass-dependent Ca... [Pg.117]

To maximize analytical precision and reproducibility, Marshall and DePaolo (1982) chose n = 42, and hence report all data in terms of Ca/ Ca ratios normalized to RJa/ Ca = 0.31221. This choice allows one to use an isotope ratio spanning two mass units ( RJa/ Ca) to make a correction (for instrumental mass discrimination) to another isotope ratio spanning two mass units ( Ca/ Ca). The only other likely choice is to use Ca/ Ca (i.e., n = 44), which spans four mass units and hence would have twice as large a correction for instrumental mass discrimination. [Pg.264]

A double spike technique is essential for TIMS analyses of Se and Cr, and may also be useful in MC-ICP-MS analysis. Briefly, two spike isotopes with a known ratio are added to each sample, and the measured ratio of the spike isotopes is used to determine and correct for instrumental bias. Examples of Se and Cr double spikes currently in use are given in Table 1. The fact that small amounts of the spike isotopes are present in the samples and small amormts of nominally unspiked isotopes are found in the spikes is not a problem, as the measurements allow highly precise mathematical separation of spike from samples. Algorithms for such calculations are described by Albarede and Beard (2004) and, specifically for Se, by Johnson etal. (1999). [Pg.293]

In the first section, the mechanisms involved in size exclusion chromatography are discussed this is an area where additional understanding and clarification still are needed. Data treatment with respect to statistical reliability of the data along with corrections for instrumental broadening is still a valid concern. Instrumental advances in the automation of multiple detectors and the developm.ent of a pressure-programmed, controlled-flow supercritical fluid chromatograph are presented. [Pg.1]

Advances in TIMS-techniques and the introduction of multiple collector-ICP-MS (MC-ICP-MS) techniques have enabled the research on natural variations of a wide range of transition and heavy metal systems for the first time, which so far could not have been measured with the necessary precision. The advent of MC-ICP-MS has improved the precision on isotope measurements to about 40 ppm on elements such as Zn, Cu, Fe, Cr, Mo, and Tl. The technique combines the strength of the ICP technique (high ionization efficiency for nearly all elements) with the high precision of thermal ion source mass spectrometry equipped with an array of Faraday collectors. The uptake of elements from solution and ionization in a plasma allows correction for instrument-dependent mass fractionations by addition of external spikes or the comparison of standards with samples under identical operating conditions. All MC-ICP-MS instruments need Ar as the plasma support gas, in a similar manner to that commonly used in conventional ICP-MS. Mass interferences are thus an inherent feature of this technique, which may be circumvented by using desolvating nebulisers. [Pg.33]

Maximum extinction coefficients will not do, even for non-rigorous discussions, since band widths differ so widely. Quantitative discussions should include corrections for instrumental broadening the procedures available are described in (59). [Pg.23]

Tung (55) has shown that the normalized observed SEC chromatogram, F(v), at retention volume v is related to the normalized SEC chromatogram corrected for instrument broadening, W(y), by means of the shape function G(v,y) through the relation... [Pg.7]

M (t) and M (t) are the true or experimentally determined molecular weight averages. The "Method of Molecular Weight Averages" was included in the ASTM Standard Method, D3536-76, to correct for instrument spreading effects. [Pg.9]

Correction for Instrumental Broadening in Size Exclusion Chromatography Using a Stochastic Matrix Approach... [Pg.287]

Most methods of correction for instrumental broadening in SEC (or hydrodynamic chromatography) are based on the deterministic integral equation due to Tung ( ) ... [Pg.287]

All isotope ratio measurements have to be corrected for instrumental mass bias by normalising to an invariant isotope of the same element (internal correction) or, whenever the internal approach cannot be applied, to a well-characterised isotope standard material (external correction). However, the external correction method requires the mass discrimination of an element being identical for the sample and the standard, which is not always the case. A large benefit of the hyphenated chromatography-ICP-MS system is that all measurements of standards and real samples can be carried out with exactly the same matrix - the eluent of the HPLC system. [Pg.43]

The linewidth (corrected for instrumental effects) may also provide important chemical information of several types. For example, if the chemical environment of a resonant atom is not the same for all of the atoms in the sample, then a broadening of the observed resonance is expected. That is, the observed resonance is a sum of the contributions from each atom, the latter not all having the same Mossbauer parameters. Thus for a small catalyst particle, interesting particle size information might be contained in the linewidth due to the contribution from the surface atoms to the Mossbauer spectrum. The distribution (clustered or uniform) of resonant atoms throughout a multicomponent catalyst particle may also be reflected in the linewidth. [Pg.149]

Although in many laboratories the methods described above remain the methods of choice for determining the proximate analysis of coal, there is also a test method for the proximate analysis of coal by instrumental procedures, assuming that calibration is an integral part of the procedure (ASTM D-5142). This method covers the determination of moisture, volatile matter, and ash and the calculation of fixed carbon in the analysis of coal and coke samples prepared in accordance with standard protocols (ASTM D-2013). The results may require a correction for bias or be corrected for instrument calibration using samples of known proximate... [Pg.63]

Figure 14. Deconvolution correction for instrumental broadening—simulation. Profile b unfolded from Profile a to give Profile c. Figure 14. Deconvolution correction for instrumental broadening—simulation. Profile b unfolded from Profile a to give Profile c.

See other pages where Correction for instrumental is mentioned: [Pg.273]    [Pg.43]    [Pg.50]    [Pg.54]    [Pg.359]    [Pg.404]    [Pg.27]    [Pg.34]    [Pg.35]    [Pg.118]    [Pg.118]    [Pg.125]    [Pg.294]    [Pg.95]    [Pg.105]    [Pg.533]    [Pg.194]    [Pg.317]    [Pg.261]    [Pg.60]    [Pg.37]    [Pg.46]   


SEARCH



Correction for instrumental broadening

Instrumentation for

Instruments for

© 2024 chempedia.info