Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Calibrated system

Evaluation and calibration. A piece of tube was rotated around its own axis during four channel wall thickness mea.surements (Figure 7). The four traces are not identical A rotation apart as should be expected. The calibrations of the four equipment s from the manufacture was not the same. Especially one of the traces has less dynamic than the other three. Based on these observations a dynamic calibration system was suggested using a tube, which could be rotated around its own axis in the measuring system. The values should be verified using traditional mechanical measurement around the tube circumference. The prototype system was permanently installed in the workshop at the production hall. Experimental work was more difficult under such circumstances so our participation in the development work stopped. [Pg.901]

There are two systems used for maintaining the accuracy and integrity of measuring devices a calibration system and a verification system. The calibration system determines the accuracy of measurement and the verification system determines the integrity of the device. If accuracy is important then the device should be included in the calibration system. If accuracy is not an issue but the device s form, properties, or function is important then it should be included in the verification system. You need to decide the system in which your devices are to be placed under control and identify them accordingly. [Pg.403]

Process all measuring devices through your established calibration system. [Pg.422]

Do verify that your subcontractors have an adequate calibration system. [Pg.425]

Measured by a radioactivity calibration system before administering PO 4—10 mQ thyroid cancer 50—150 mQ... [Pg.532]

The measurement model of the time-invariant calibration system (eq. (41.5)) should now be expanded in the following way ... [Pg.598]

FIGURE I Schematic diagram showing the overall validation strategy, which includes calibration, system validation, and system suitability testing. Figure adopted with ideas from Reference I. [Pg.292]

This procedure describes conditions pertaining to an adequate and organized calibration system of measuring devices. [Pg.118]

Periodic review of the calibration system should happen once a year using an established inspection checklist. [Pg.119]

Does the calibration system identify the source and magnitude of uncertainties associated with calibration and product characteristics ... [Pg.128]

Do procedures ensure that a subcontractor employs a measurement and calibration system that complies with the company s requirements ... [Pg.129]

Current tests are generally destructive (i.e., sample is altered or destroyed) and robust estimates of measurement system variability (all aspects of the procedure including the operators) are difficult to obtain without using methods such as Gauge R R-reproducibility and repeatability (14). Suitability of current methods then is based on calibration using a calibrator system that has its own built-in variability and other assumptions (e.g., in physical testing such characteristics as size, shape, density can alter aerodynamic and/or hydrodynamic behavior of materials in a test system and contribute to systems variability). [Pg.500]

If there is any doubt about whether 100% of the analyte is presented to the measuring system or that the response of the calibrated system leads to no bias, then the assumptions must be tested during method validation and appropriate actions taken. If a series of measurements of a CRM (not used for calibration) leads to the conclusion that there is significant bias in the observed measurement result, the result should be corrected, and the measurement uncertainty should include the uncertainty of the measurement of the bias. If the bias is considered insignificant, no correction need be made, but measuring the bias and concluding that it is zero adds uncertainty (perhaps the bias was not really zero but is less than the uncertainty of its measurement). One approach to measurement uncertainty is therefore to include CRMs in the batch to be used to correct for bias, and then the uncertainty of estimation of bias, which includes the uncertainty of the quantity value of the CRM, is combined with the within-laboratory reproducibility. In some fields of analysis it is held that routine measurement and correction for bias... [Pg.180]

The [calibration system] controls, calibrates, and maintains inspection, measuring, and test equipment to demonstrate the conformance of the product to the specified requirements. Equipment is used in such a manner so as to ensure that measurement uncertainty is known and is consistent with the required measurement capability. The measurements made are identified, the accuracy is stated, and the appropriate inspection, measuring, and test equipment is used. [Pg.169]

The calibration system ensures that the measuring and test equipment is calibrated, adjusted, repaired, or replaced prior to becoming inaccurate. [Pg.231]

The American CPAC initiative NeSSI [23] developed a micro reactor sampling and calibration system intended for analytical applications in the oil industry. Industrial partners such as Swagelok and Parker/Hannifin developed the system originally designed for the gas supply in clean room facilities. This approach is well advanced with respect to valves, gauges, analytical sensors and pipe fittings. [Pg.583]

LDPE membranes were prefouled for one month in water collected from a park fountain. Membranes became heavily fouled with a thick algal and bacterial film. The fouled membranes were used for the construction of Chemcatcher samplers fitted with sorption phases previously spiked with several PRCs. Samplers fitted with either fouled or unfouled membranes were simultaneously exposed (rotation speed 40 rpm and water temperature 11°C) in a laboratory flow through a calibration system... [Pg.48]

FIGURE 3.4 Biofouling reduces the exchange kinetics of PRCs (deuterated acenaphthene and fluorene) between the nonpolar Chemcatcher sampler (fitted with either fouled or unfouled LDPE membranes) and water. The experiment was performed in a laboratory flow-through calibration system at a water temperature of 11°C with simulated water turbulence of 40 rpm. MD(t)/MD(0) is the fraction of the PRC remaining in the sampler during exposure. [Pg.49]

By using suitable calibration systems the interferences can be compensated for without being determined in detail. For this purpose calibration using model solutions or the use of the addition method have proved useful. [Pg.235]

There are various calibration systems for AAS which differ mainly in the time needed. [Pg.242]


See other pages where Calibrated system is mentioned: [Pg.56]    [Pg.481]    [Pg.73]    [Pg.33]    [Pg.34]    [Pg.74]    [Pg.171]    [Pg.141]    [Pg.163]    [Pg.298]    [Pg.438]    [Pg.26]    [Pg.16]    [Pg.127]    [Pg.127]    [Pg.127]    [Pg.251]    [Pg.150]    [Pg.76]    [Pg.481]    [Pg.249]    [Pg.328]    [Pg.103]    [Pg.59]    [Pg.361]    [Pg.122]    [Pg.322]    [Pg.342]    [Pg.368]    [Pg.140]    [Pg.236]   
See also in sourсe #XX -- [ Pg.341 ]




SEARCH



Calibration commercial systems

Calibration flow-injection analysis systems

Calibration of the HPLC system

Calibration supply systems

Calibration, 296 data system

HPLC system calibration

Raman spectroscopy system calibration

System calibration

System calibration

System calibration program

System equation of a calibration line with drift

© 2024 chempedia.info