Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Analyses Using Internal Standards

To understand the effectiveness and limitations of a VIS in eliminating uncertainties in dispensed volumes, we shall consider in some detail the example of a standard calibration curve using clean solutions of analytical standard together with a VIS, analogous to the derivation of Equation [8.64], As for any procedure incorporating a cahbration curve its characteristics must be determined, particularly the question of whether or not it encompasses the origin, and also the extent of the linear dynamic range. [Pg.437]

In the present context involving use of a VIS, a suitable calibration procedure could involve the following steps  [Pg.437]

The present treatment assumes the vahdity of a hnear response curve, possibly with a non-zero intercept (this assumption must apply to both the analyte and the VIS). Then, as before, for the analyses of the calibration solutions but for now ignoring the data obtained for the spiked-in VIS  [Pg.437]

The intercept Ayis.a pertains to the analyses of the VIS in the VIS-spiked calibration solutions and can not be determined since these solutions all contained the same quantity Qvis.a- Instead, it is assumed that this parameter is equal to Ayis determined from analysis of the series of solutions of only VIS (Equation [8.76b]). (Of course it is always preferable to ensure that intercepts of calibration curves are statistically zero, see Section 8.3.2.) The quantity [(5 . /a)/(5 vis-/vis)]/Qvis,a is determined as the slope of the least-squares regression of the values of the left side of Equation [8.77] (i.e., the ratio of signals for analyte to VIS) against Q this regression line should have a zero intercept (to within the confidence limits determined as before) if the two intercepts were measured properly. This completes the calibration procedure. [Pg.438]

Application of such a calibration to quantitative analysis of sample extracts requires that these extracts be spiked with the identical quantity Qvis.a of volumetric internal standard as used in the calibration experiments. (It is emphasized that all of these spiking procedures, i.e. calibration solutions plus sample extracts, should be done at the same time, to minimize any drift in the value of Cyis as a result of uncontrolled evaporation.) Then it is straightforward to show that  [Pg.438]


The yields were determined by GC analyses using internal standard. Also l,l-bis(4-tolyl)pent-l-ene 0.06%. Also 2-methyl-l,l-bis(4-tolyl)prop-l-ene <0.06%. Also 2-phenyl-l,l-bis(4-tolyl)ethene... [Pg.383]

Normal-phase/reversed-phase chromatography is the ideal combination for semipreparative and quantitative separations in two-dimensional HPLC. Vitamins D2 and D3 coelute during the semipreparative stage, allowing a narrow retention window to be collected for analysis using internal standardization. By this means, Johnsson et al. (201) obtained a vitamin D3 detection limit of 0.1 yug/kg for milk and milk products. [Pg.374]

Based upon glpc analysis using internal standard, BrCH2C02Et/2 = 100%. [Pg.26]

Quantitative Analysis Using Internal Standard Method [12]... [Pg.87]

Complex 8 (0.05 mmol), ester (5 mmol), and toluene (3 ml) were refluxed under argon. Conversion and yields are based on GC analysis using internal standard. [Pg.13]

Measured volume of sample is extracted. The extract is dried, concentrated, and analyzed by gas chromatographic (GC) mass spectrometric (MS) method. Qualitative compoimd identification is based on retention time and relative abundance of three characteristic masses (m/z). Quantitative analysis uses internal standard techniques with a single characteristic tn/z. [Pg.730]

The proposed PAS-FTIR procedure allows the determination of 30 samples h does not require any sample pre-treatment and avoids waste generation. So, it can conclude that the PAS-FTIR methodology is a reliable alternative to long and tedious classical method of analysis of Mancozeb and improves the sample frequency achieved by the FTIR procedure based on the measurement of KBr disks using internal standard. [Pg.93]

Qualitative HPLC methods, using area percent, are used to monitor the disappearance of starting material and the formation of byproduct. Without the inclusion of an internal standard and the calculation of response factors, it is not possible to establish with certainty whether all of the starting material can be accounted for. An internal standard must be stable in the reaction mixture, must not co-elute with any of the components, and must be stable in the mobile phase. Ideally, the internal standard has a retention time about half that of the total analysis time. Internal standardization is extremely useful for kinetic studies. Added to the reaction vessel, samples that are withdrawn at various times will contain identical concentrations of internal standard, and chromatograms can be directly compared or adjusted to identical scales to correct for variation in injection volume. [Pg.184]

It appears that purification of commercially available solvents is sometimes required for the complete elimination of impurity resonances. Occasionally, these impurities may be turned into advantage, as in the case of C2D2CI4 where the (known) C2DHCI4 content may be used as an internal standard for quantitation. Thus, removal of every impurity peak is not always essential for identification and quantitative analysis of stabilisers in PE. Determination of the concentration of additives in a polymer sample can also be accomplished by incorporation of an internal NMR standard to the dissolution prepared for analysis. The internal standard (preferably aromatic) should be stable at the temperature of the NMR experiment, and could be any high-boiling compound which does not generate conflicting NMR resonances, and for which the proton spin-lattice relaxation times are known. 1,3,5-Trichlorobenzene meets the requirements for an internal NMR standard [48]. The concentration should be comparable to that of the analytes to be determined. [Pg.698]

Holzbecker and Ryan [825] determined these elements in seawater by neutron activation analysis after coprecipitation with lead phosphate. Lead phosphate gives no intense activities on irradiation, so it is a suitable matrix for trace metal determinations by neutron activation analysis. Precipitation of lead phosphate also brings down quantitatively the insoluble phosphates of silver (I), cadmium (II), chromium (III), copper (II), manganese (II), thorium (IV), uranium (VI), and zirconium (IV). Detection limits for each of these are given, and thorium and uranium determinations are described in detail. Gamma activity from 204Pb makes a useful internal standard to correct for geometry differences between samples, which for the lowest detection limits are counted close to the detector. [Pg.282]

As stated in Basic Protocol 1, the method of HPLC used may change depending on the organic acids present in the sample. This method uses a Bio-Rad Aminex HPX-87H column for HPLC analysis. An internal standard is important for the analysis of organic acids. This provides a means of not only determining if the analysis is working but also quantitating the percent recovery of the method. [Pg.1122]

Methyl esters were identified by comparison with authentic compounds on both the Carbowax 20M column and a XE-60 (a General Electric cyanosilicone gum) column. Some of the earlier identifications were made with a Carbowax 20M + H3P04 column and confirmed on l,2,3-tris(2-cyanoethoxy) propane. Also, some of the analyses of free acids were run on a Carbowax 20M-TPA(terephthalic acid terminated) column rather than the FFAP. Quantitative analysis of free acids and methyl esters was carried out using internal standards. Acetophenone was used with the acids, and veratrole was used with the methyl esters. Formic acid in the nitrobenzene and nitric acid liquors was analyzed quantitatively by direct calibration. [Pg.197]

The extract should be then injected onto the GC column for GC or GC/MS analysis. If internal standard calibration is performed, spike three or more internal standards into the sample extract. Any organophosphorus pesticides whose analysis is not required or which is not found to be present in the screening test may be used as an internal standard. The chromatographic columns and conditions are presented below ... [Pg.217]

Quantitation. The organic species identified by the GC-MS analyses were quantitated by GC analysis using internal and external standardization methods. Pure compounds representative of the various compound classes identified by GC-MS were selected as standards and methylated. A specific amount of each standard was co-injected with each sample to confirm the GC-MS identifications. For quantitation purposes, each standard was injected onto the gas chromatograph prior to and following sample analyses. The response factor of each standard was calculated under analytical conditions identical to those of the sample analyses. [Pg.248]

As individual sample QC checks, internal standards are important in compound quantitation. They should be monitored with the same care as other QC checks. The deterioration of internal standard area counts (the area under a chromatographic peak) reflects the changes in the analytical system that may eventually degrade the quality of analysis to an unacceptable level. EPA Methods 8260 and 8270 require that the area for each internal standard be within the range of —50 to +100 percent of the area of the internal standards in the most recent CCV (EPA, 1996a). This requirement may be used as a general rule for all other methods that use internal standard calibration. [Pg.261]

Many practical electrodes are prepared from powdered active mass and some conductive additive, such as carbonaceous materials, that are bonded to a metallic current collector with a polymeric binder such as PVDF (described in the previous section). Such electrodes can be measured directly. Is it very useful to measure such electrodes in their pristine from, before any electrochemical treatment, and then as a function of their electrochemical history. For quantitative analysis (phase composition, evaluation of concentration of constituents in mixtures, etc.) it is important to use internal standards in the sample. Fortunately, several components of composite electrodes, which are, in any event, contained in the sample measured, can be used as internal standards. These include the current collector (Cu, Al), the conductive additive, such as graphite, or the binder, such as Teflon. [Pg.127]

Using internal standards in quantitative analysis is advantageous, for instance, in cases where the sample thickness cannot be determined exactly, or in gaseous samples whose total pressure is unknown. The compound which serves as an internal standard should have a simple spectrum which does not interfere with the bands of the analyte. It should be a stable substance, readily available, and non-toxic. The following substances have been recommended as internal standards by Colthup et al. (1975)... [Pg.426]

Capillary isoelectric focusing is a rapid analysis technique with typical run times of 5-30 min, fully automated with on-line detection and real-time data acquisition, and minute sample consumption (a few microliters is enough for repetitive injections). A linear dynamic range over one order of magnitude is achievable, and a detectability down to 5-10 u.g/mL. A resolution of Ap/-0.01 is possible under optimized conditions. Reproducibility of pi determination is typically <0.5% (RSD) using internal standards. [Pg.291]

For superconductor materials and precursors high-precision analysis is very important. To this aim flame atomic absorption can be applied, provided the precision is optimized, as is possible, for example, by using internal standardization and multichannel spectrometers. It should, however, be remembered that to control the stoichiometric composition, in addition the determination of non-metals such... [Pg.188]

SP-2401" and 3% SP-2250. ° Detectors used by EPA standards procedures, include photoionization (PID)," electron capture (ECD)," Eourier transform infrared spectrometry (PTIR), " and mass spectrometry detectors (MSD)." ° Method 8061 employs an ECD, so identification of the phthalate esters should be supported by al least one additional qualitative technique. This method also describes the use of an additional column (14% cyanopropyl phenyl polysiloxane) and dual ECD analysis, which fulfills the above mentioned requirement. Among MSDs, most of the procedures employ electron impact (El) ionization, but chemical ionization (CI) ° is also employed. In all MSD methods, except 1625, quantitative analysis is performed using internal standard techniques with a single characteristic m/z- Method 1625 is an isotope dilution procedure. The use of a FTIR detector (method 8410) allows the identification of specific isomers that are not differentiated using GC-MSD. [Pg.1118]

Chart LL Useful internal standards for analysis of solution-phase combinatorial libraries. 2,5-dimethylfuran (DMFu, 1) and hexamethyldisiloxane (HMDS, 2) are volatile standards that can easily be evaporated from solution-phase samples after analysis. 1,4-bis (trimethylsilyl)benzene (BTMSB, 3) is stable in DMSO for up to one month and is transparent in HPLC-UV and MS analyses. [Pg.17]

In regards to the applications of GC, different studies will be presented, among which we wiU discuss the optimization of the operational conditions used to separate the different components, the analysis of the variation in the composition of regional essential oils, the measurement, using internal standards, of oils as well as oil modification by fractional vacuum distillation, and the study of semi-synthesis reactions to obtain high added-value compounds, starting from oils or their components. [Pg.308]


See other pages where Analyses Using Internal Standards is mentioned: [Pg.383]    [Pg.436]    [Pg.130]    [Pg.811]    [Pg.30]    [Pg.275]    [Pg.212]    [Pg.131]    [Pg.162]    [Pg.648]    [Pg.726]    [Pg.136]    [Pg.117]    [Pg.407]    [Pg.84]    [Pg.166]    [Pg.131]    [Pg.28]    [Pg.1025]    [Pg.105]    [Pg.43]    [Pg.147]    [Pg.131]    [Pg.689]    [Pg.86]    [Pg.332]    [Pg.22]    [Pg.75]    [Pg.2058]   


SEARCH



Analysis of Old Lubricating Oil for Total Metal Content Using a Slurry Method with Internal Standard

Internal standards

International Standardization

International Standards

Quantitative Analysis Using Internal Standard Method

Quantitative analysis using single point internal standard

Standardization international standards

Using Internal Standardization

© 2024 chempedia.info