Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Transmission Electron Microscope TEM

A transmission electron microscope [16-18] is basically a transmission light microscope using electrons instead of light, but with a resolution some thousand times greater than a transmission light instrument enabling objects as small as a few angstroms to be studied. [Pg.460]

The instrument can be coupled with energy dispersive X-ray microanalysis, online image processing and particle size measurement. X-ray analysis ean detect all elements in the periodic table from Na through U. [Pg.462]

Leeds University [29]. They have pioneered the use of TEM studies of carbon fiber and found that for carbon fiber, the electron wavelength is equivalent to the stacking distances between the graphite plates in the crystallites, hence they can be seen. Some of their findings are  [Pg.463]

A and HT fibers have no sheath and have a homogeneous structure. [Pg.463]

The orientation of crystallites is random, but with a preferred orientation along the fiber axis. The orientation of Type A and HT fibers are about 45° and 40° respectively. [Pg.463]


Fig. Vni-3. (a) Atomic force microscope (AFM) and (b) transmission electron microscope (TEM) images of lead selenide particles grown under arachidic acid monolayers. (Pi Ref. 57.)... Fig. Vni-3. (a) Atomic force microscope (AFM) and (b) transmission electron microscope (TEM) images of lead selenide particles grown under arachidic acid monolayers. (Pi Ref. 57.)...
The history of EM (for an overview see table Bl.17,1) can be interpreted as the development of two concepts the electron beam either illuminates a large area of tire sample ( flood-beam illumination , as in the typical transmission electron microscope (TEM) imaging using a spread-out beam) or just one point, i.e. focused to the smallest spot possible, which is then scaimed across the sample (scaiming transmission electron microscopy (STEM) or scaiming electron microscopy (SEM)). In both situations the electron beam is considered as a matter wave interacting with the sample and microscopy simply studies the interaction of the scattered electrons. [Pg.1624]

To illustrate the effect of radial release interactions on the structure/ property relationships in shock-loaded materials, experiments were conducted on copper shock loaded using several shock-recovery designs that yielded differences in es but all having been subjected to a 10 GPa, 1 fis pulse duration, shock process [13]. Compression specimens were sectioned from these soft recovery samples to measure the reload yield behavior, and examined in the transmission electron microscope (TEM) to study the substructure evolution. The substructure and yield strength of the bulk shock-loaded copper samples were found to depend on the amount of e, in the shock-recovered sample at a constant peak pressure and pulse duration. In Fig. 6.8 the quasi-static reload yield strength of the 10 GPa shock-loaded copper is observed to increase with increasing residual sample strain. [Pg.197]

For the purpose of a detailed materials characterization, the optical microscope has been supplanted by two more potent instruments the Transmission Electron Microscope (TEM) and the Scanning Electron Microscope (SEM). Because of its reasonable cost and the wide range of information that it provides in a timely manner, the SEM often replaces the optical microscope as the preferred starting tool for materials studies. [Pg.70]

Transmission electron microscopes (TEM) with their variants (scanning transmission microscopes, analytical microscopes, high-resolution microscopes, high-voltage microscopes) are now crucial tools in the study of materials crystal defects of all kinds, radiation damage, ofif-stoichiometric compounds, features of atomic order, polyphase microstructures, stages in phase transformations, orientation relationships between phases, recrystallisation, local textures, compositions of phases... there is no end to the features that are today studied by TEM. Newbury and Williams (2000) have surveyed the place of the electron microscope as the materials characterisation tool of the millennium . [Pg.221]

Fig. 6. Transmission electron microscope (TEM) image of MWCNT with the open end. The cap of the tube was removed using the purification process,... Fig. 6. Transmission electron microscope (TEM) image of MWCNT with the open end. The cap of the tube was removed using the purification process,...
As mentioned above, employment of MWCNT for field emitter will be one of the most important applications of MWCNT. For this purpose, MWCNT is prepared by the chemical purification process [30,38], in which graphite debris and nanoparticles are removed by oxidation with the aid of CuCl2 intercalation [38]. Purified MWCNT is obtained in the form of black and thin "mat" (a flake with thickness of ca. a few hundreds of [im). Figure 7 shows a typical transmission electron microscope (TEM) picture of MWCNT with an open end, which reveals that a cap is etched off and the central cavity is exposed. [Pg.8]

Fig. 10. Scanning electron microscope (SEM) image of the probe with an MWCNT attached to a silicon cantilever [35]. Protruding of one individual MWCNT has been confirmed by transmission electron microscope (TEM) measurement (not shown here). Fig. 10. Scanning electron microscope (SEM) image of the probe with an MWCNT attached to a silicon cantilever [35]. Protruding of one individual MWCNT has been confirmed by transmission electron microscope (TEM) measurement (not shown here).
As an indication of the changes in deformation modes that can be produced in ionomers by increase of ion content, consider poly(styrene-co-sodium methacrylate). In ionomers of low ion content, the only observed deformation mode in strained thin films cast from tetra hydrofuran (THF), a nonpolar solvent, is localized crazing. But for ion contents near to or above the critical value of about 6 mol%, both crazing and shear deformation bands have been observed. This is demonstrated in the transmission electron microscope (TEM) scan of Fig. 3 for an ionomer of 8.2 mol% ion content. Somewhat similar deformation patterns have also been observed in a Na-SPS ionomer having an ion content of 7.5 mol%. Clearly, in both of these ionomers, the presence of a... [Pg.146]

The majority of PH As biosynthesis is performed by various microorganisms, especially bacteria. They can produce PHAs from a number of substrates and accumulate in their cells as carbon source and energy reserve under imbalanced growth conditions such as nutrient limitation. Fig.7 shows PHA accumulated in their cells that are characterized by transmission electron microscopic (TEM) technique. [Pg.50]

Fig. 7. (A and B) Images of PHAs accumulated in bacteria Halomonas TDOl based on transmission electron microscope (TEM) (Tan et al., 2011)... Fig. 7. (A and B) Images of PHAs accumulated in bacteria Halomonas TDOl based on transmission electron microscope (TEM) (Tan et al., 2011)...
FIGURE 3.3 (a) Transmission electron microscopic (TEM) image of acrylic rubber (ACM)-siUca hybrid nanocomposite synthesized from 10 wt% of tetraethoxysilane (TEOS). (From Bandyopadhyay, A., Bhowmick, A.K., and De Sarkar, M., J. Appl. Polym. Sci., 93, 2579, 2004. Courtesy of Wiley Interscience.) Transmission electron microscopic (TEM) photographs of acrylic rubber (ACM)-silica hybrid nanocomposites prepared from (b) 30 wt% and (c) 50 wt% tetraethoxysilane (TEOS) concentrations. (From Bandyopadhyay, A., Bhowmick, A.K., and De Sarkar, M., J. Appl. Polym. Sci., 93, 2579, 2004. Courtesy of Wiley InterScience.)... [Pg.63]

FIGURE 3.16 Morphology and visual appearance of acrylic rubber (ACM)-silica and epoxidized natural rubber (ENR)-silica hybrid composites prepared from different pH ranges (a) transmission electron microscopic (TEM) picture of ACM-siUca in pH 1.0-2.0, (b) scanning electron microscopic (SEM) picture of ACM-siUca in pH 5.0-6.0, (c) SEM image of ACM-siUca in pH 9.0-10.0, (d) TEM picture of ENR-silica in pH... [Pg.74]

FIGURE 4.2 Transmission electron microscopic (TEM) image of ethylene-vinyl acetate (EVA)-expanded graphite (EG) (4 wt%) nanocomposites. (From George, J.J. and Bhowmick, A.K., J. Mater. Sci., 43, 702, 2008. Courtesy of Springer.)... [Pg.91]

Fourier Transform (FT) Ranun spectroscopy (Model RFS 100/S, BRUKER Co.) using ND YAG laser was used to analyze the products on their structure electronic and vibration properties. The morphology of CNTs was observed by scanning dartron microscopy (SEM, Model S-4200, Hitach Co.) and transmission electron microscope (TEM, Modd JEOL 2000FX-ASID/EDS, Philips Co.). [Pg.750]

Analysis of individual catalyst particles less than IMm in size requires an analytical tool that focuses electrons to a small probe on the specimen. Analytical electron microscopy is usually performed with either a dedicated scanning transmission electron microscope (STEM) or a conventional transmission electron microscope (TEM) with a STEM attachment. These instruments produce 1 to 50nm diameter electron probes that can be scanned across a thin specimen to form an image or stopped on an image feature to perform an analysis. In most cases, an electron beam current of about 1 nanoampere is required to produce an analytical signal in a reasonable time. [Pg.362]

Nitrogen adsorption isotherms were measured with a sorbtometer Micromeretics Asap 2010 after water desorption at 130°C. The distribution of pore radius was obtained from the adsorption isotherms by the density functional theory. Electron microscopy study was carried out with a scanning electron microscope (SEM) HitachiS800, to image the texture of the fibers and with a transmission electron microscope (TEM) JEOL 2010 to detect and measure metal particle size. The distribution of particles inside the carbon fibers was determined from TEM views taken through ultramicrotome sections across the carbon fiber. [Pg.56]

The droplet sizes of the nanoemulsions characterized by dynamic light scattering at O/S ratios between 50 0 and 70 30 and a constant water content of 90 wt% were between 200 and 220 nm, displaying a slight increase with increasing O/S ratio. Figure 6.3 shows atypical cryo-transmission electron microscope (TEM) image of an... [Pg.168]

III. Transmission electron microscopy of radish seeds Transmission electron microscopy (TEM) of radish seeds was done as listed below For TEM preparations, the specimens after fixation and dehydration, were embedded in Epon 812 resin (Luft, 1961). Thick sections (ca. 1mm each) were stained with 0.1% toluidine blue and observed with a Zeiss light photomicroscope. Thin sections, obtained with a diamond knife on a Supernova microtome, were sequentially stained at room temperature with 2% uranyle acetate (aqueous) for 5 min and by lead citrate for 10 min (Reynolds, 1963). Ultrastructural studies were made using a Philips CM12 transmission electrone microscope (TEM) operated at 80 KV. [Pg.79]


See other pages where Transmission Electron Microscope TEM is mentioned: [Pg.331]    [Pg.191]    [Pg.121]    [Pg.135]    [Pg.161]    [Pg.50]    [Pg.65]    [Pg.230]    [Pg.49]    [Pg.29]    [Pg.32]    [Pg.145]    [Pg.62]    [Pg.559]    [Pg.216]    [Pg.47]    [Pg.49]    [Pg.322]    [Pg.777]    [Pg.786]    [Pg.577]    [Pg.554]    [Pg.522]    [Pg.547]    [Pg.234]   
See also in sourсe #XX -- [ Pg.147 ]




SEARCH



Electron microscop

Electron microscope

Electron microscopic

Microscopes electron microscope

TEM

TEM (transmission electron

Transmission electron microscope TEM) images

© 2024 chempedia.info