Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Study sample, defining

Acceptance Criteria. Daily acceptance criteria are the performance standards against which applications of method are judged. Two approaches are used to define the acceptance criteria (1) with respect to standard deviation of data obtained during validation, and (2) with reference to externally imposed requirements. These criteria should be established during method development or immediately following validation, prior to application of the method to study samples [1,3,5,11,14,17]. Aspects of the assay that could or should be included in acceptance criteria include ... [Pg.125]

Because the catabolic and metabolic pathways of biotech drugs are often poorly defined and sufficiently sensitive comparator assays are lacking, additional matrix effect tests by parallelism should be conducted with actual study samples. These are often performed on subject samples with aberrant PK profiles. A pool from several time points with sufficient analyte concentration of that subject is serially diluted. The observed concentration times the dilution factors should be within... [Pg.159]

Over recent years, oncology has been a key focus of proteomic technologies as applied in both expression and functional studies. Expression proteomic studies are screening for differences in protein patterns between tumor and control tissues. Functional proteomic studies are defined as the combination of readouts with proteomics data from the same sample at different points of time (under different conditions), with the aim of detecting and prioritizing certain proteins or polypeptides of functional relevance. [Pg.114]

To study the relaxation of Ht to the high field seeking hyperfine states we plot our data as V tl/N2 versus time as shown in Fig.4. W(t) is the total number of atoms in the trap at a given time t, obtained by integrating the observed flux from t to . Vy is the effective volume of the sample defined by V 0/V2e,... [Pg.924]

The final report submitted to the FDA must discuss both safety and efficacy. In the efficacy section, a precise definition of which specific patients are included in the efficacy analysis is necessary. The Guidelines suggest doing an intent-to-treat analysis, including all patients randomized in the study. This is done in addition to the study sample that is defined as clinically relevant. [Pg.307]

ANOVA can be extended to situations where the experimental units (in our context, study participants) are classified on a number of factors. When they are classified on the basis of one factor, it is referred to as a one-way ANOVA. The result of partitioning the total variance into its components, in this case among and within samples defined by one factor, is displayed in Table 11.2. [Pg.154]

Sometimes a clinical site may deviate from the study protocol-defined sample-collection procedures. For example, instead of collecting blood into heparinized tubes, the sample may be collected into a tube containing another anticoagulant, such as citrate or EDTA. Under these conditions, it is necessary to cross-validate each of the plasma/serum matrices QC samples in the deviant matrix must be prepared and processed along with the deviant samples to ensure that the method is validated in the new matrix. [Pg.272]

Defining the scope of the method Support for PK/PD/other studies Desired sensitivity Matrix of study samples Sample collection and storage conditions Assay format/technology platform... [Pg.46]

As defined in ICH-Q2A [ 1 ], linearity of an analytical procedure is its ability, within a defined range, to obtain results directly proportional to the concentrations of the analyte in the sample. This is sometimes misinterpreted to imply that the assay response should be a linear function of the analyte concentration. The linearity criteria should be applied only to the back-calculated (calibrated) results, that is, the back-calculated results should be directly proportional to the analyte concentrations of the sample within the range in which the study samples are expected to fall. [Pg.117]

A decision about the validity/suitability of an analytical method for routine testing of study samples is taken, based on the estimated measurement error profile. Such a decision is possible only when no more than one section of the concentration range has its measurement error profile within the acceptance limits. For example, it is not reasonable to consider a method to be valid in a certain section of low concentrations and a different section of higher concentrations. In such cases, the measurement error profile is probably not precise enough to define a unique section of the range where the method is valid. In such cases, it is recommended to add a few more runs in the experimental design to obtain a more reliable estimate of the measurement error profile. [Pg.122]

Absorption spectra are obtained for each compound to determine which optical density (OD) to use for the growth experiment, and at what pH to prepare the solutions and the growth media. These studies also defined what pH value to use for the reference sample (control). The solubility of the oxometalates were first determined in distilled water and then in growth media. With a few exceptions, all the compounds used in this study were soluble and stable at 5.0 mM in distilled water. However, the solubility was significantly less for compounds PW,2 (0.50 mM) and PV(V)Wn (1.3 mM) in the growth media. The stability of the complexes were determined in the stock solutions prepared in distilled water and then in media. The polyoxoanion solution in media is prepared without buffer to increase the solubility and to assure that the pH in media does not overpower the pH resulting from the oxometalate. [Pg.183]

Because of the risk of analytical errors, it is crucial that every laboratory runs and reports a quality assurance program (cf. Friberg, 1988). This should include both internal and external quality control. The internal quality control procedure should include, in each analytical series, a sufficient number of samples with the same matrix as the study samples, and with a relevant mercury concentration. The results obtained are compared with pre-defined levels of acceptability. The external quality control may be assessed by inter-laboratory exchange of samples with (to the iaboratory) unknown concentrations, and by analysis of reference samples with well-defined levels. [Pg.412]


See other pages where Study sample, defining is mentioned: [Pg.193]    [Pg.154]    [Pg.42]    [Pg.62]    [Pg.145]    [Pg.715]    [Pg.715]    [Pg.239]    [Pg.389]    [Pg.771]    [Pg.294]    [Pg.246]    [Pg.117]    [Pg.31]    [Pg.100]    [Pg.221]    [Pg.290]    [Pg.291]    [Pg.293]    [Pg.484]    [Pg.306]    [Pg.74]    [Pg.408]    [Pg.77]    [Pg.240]    [Pg.158]    [Pg.100]    [Pg.2]    [Pg.84]    [Pg.100]    [Pg.143]    [Pg.205]    [Pg.210]    [Pg.40]    [Pg.749]    [Pg.387]    [Pg.63]   
See also in sourсe #XX -- [ Pg.715 ]




SEARCH



Sample define

Sample defined

Sampling Defined

© 2024 chempedia.info